Anterior segment biometry with 2 imaging technologies: very-high-frequency ultrasound scanning versus optical coherence tomography.

David P Piñero1, Ana Belén Plaza, Jorge L Alió.   

Abstract

PURPOSE: To determine the interchangeability of 2 anterior segment imaging systems: a very-high-frequency (VHF) ultrasound scanning system (Artemis 2, Ultralink LLC) and an optical coherence tomography (OCT) system (Visante, Zeiss).
SETTING: Vissum Instituto Oftalmologico de Alicante, Alicante, Spain.
METHODS: This study comprised 20 eyes without pathology or previous surgery. The anterior chamber depth (ACD), central corneal thickness (CCT), angle-to-angle distance (ATA), and the iridocorneal angle size (IAS) at the 0-degree and 180-degree positions were measured with 2 imaging techniques: VHF ultrasound scanning and OCT. Analysis of agreement and interchangeability was performed by the Bland and Altman method. In addition, each measurement was performed 3 times consecutively to determine intrasession repeatability by means of the coefficient of variation (CV) and the intraclass correlation coefficient (ICC).
RESULTS: No statistically significant differences were found between imaging techniques in ACD, CCT, or ATA (P>.40). The ranges of agreement were 0.20 mm, 16.11 mum, and 0.80 mm for ACD, CCT, and ATA, respectively. Regarding IAS, no statistically significant differences were found in the nasal (P = .78) or temporal (P = .63) measurements between devices. However, the range of agreement for nasal (14.3 degrees) and temporal (14.90 degrees) values was relevant, indicating the 2 techniques cannot be used interchangeably for IAS measurement. Excellent intrasession repeatability scores were obtained (CV and ICC).
CONCLUSION: The Artemis 2 and the Visante OCT systems provide equivalent and repeatable measurements of the ACD, CCT, and ATA and can be used interchangeably for these purposes.

Entities:  

Mesh:

Year:  2008        PMID: 18165088     DOI: 10.1016/j.jcrs.2007.08.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  19 in total

1.  Comparison of anterior section parameters using anterior segment optical coherence tomography and ultrasound biomicroscopy in myopic patients after ICL implantation.

Authors:  Jing Zhang; Hui-Hui Luo; Jing Zhuang; Ke-Ming Yu
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

2.  Evaluation of anterior chamber parameters with spectral-domain optical coherence tomography.

Authors:  Isabel Pinilla Lozano; Carmen López de la Fuente; Francisco Segura; Elvira Orduna Hospital; Ana Sánchez-Cano
Journal:  Jpn J Ophthalmol       Date:  2017-11-29       Impact factor: 2.447

3.  Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.

Authors:  Noelia Martínez-Albert; Jose J Esteve-Taboada; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-02       Impact factor: 3.117

4.  Semiautomatic extraction algorithm for images of the ciliary muscle.

Authors:  Chiu-Yen Kao; Kathryn Richdale; Loraine T Sinnott; Lauren E Grillott; Melissa D Bailey
Journal:  Optom Vis Sci       Date:  2011-02       Impact factor: 1.973

5.  Validation of optical coherence tomography-based crystalline lens thickness measurements in children.

Authors:  Bret M Lehman; David A Berntsen; Melissa D Bailey; Karla Zadnik
Journal:  Optom Vis Sci       Date:  2009-03       Impact factor: 1.973

6.  Optical coherence tomography and ultrasound biomicroscopy imaging of opaque corneas.

Authors:  Shi-you Zhou; Chun-xiao Wang; Xiao-yu Cai; David Huang; Yi-zhi Liu
Journal:  Cornea       Date:  2013-04       Impact factor: 2.651

7.  Anterior segment optical coherence tomography in eye injuries.

Authors:  Edward Wylegala; Dariusz Dobrowolski; Anna Nowińska; Dorota Tarnawska
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-09-03       Impact factor: 3.117

8.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

9.  Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles.

Authors:  Roxana Ursea; Matthew Feng; Raksha Urs; Arindam RoyChoudhury; Ronald H Silverman
Journal:  J Refract Surg       Date:  2012-11-30       Impact factor: 3.573

10.  Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

Authors:  Muriël Doors; Lars P J Cruysberg; Tos T J M Berendschot; John de Brabander; Frenne Verbakel; Carroll A B Webers; Rudy M M A Nuijts
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-04-30       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.