Anterior chamber depth in relation to refractive status measured with the Orbscan II Topography System.

Tanja M Rabsilber1, Klio A Becker, Inez B Frisch, Gerd U Auffarth.   

Abstract

PURPOSE: To evaluate the anterior chamber depth (ACD) according to refractive status, assess the reliability of repeated ACD measurements using the Orbscan II Topography System (Bausch and Lomb), compare Orbscan II and IOLMaster (Carl Zeiss Meditec AG) ACD measurements, and investigate the correlation between refraction, axial length (AL), and ACD.
SETTING: Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany.
METHODS: In this clinical study, 60 patients with a mean age of 43.8 years +/- 18.74 (SD) were assigned to 1 of 3 groups of 20 patients each according to refraction: emmetropia group; hyperopia group (mean +4.84 +/- 1.60 diopters [D]); myopia group (mean -9.64 +/- 3.79 D). Using the Orbscan II system, 3 consecutive ACD measurements (apex and 3.0 mm zone) were performed. The IOLMaster was used to measure ACD and AL.
RESULTS: The mean ACD (from epithelium) with the Orbscan II and IOLMaster, respectively, was 3.61 +/- 0.24 mm and 3.61 +/- 0.24 mm in the emmetropia group, 3.03 +/- 0.21 mm and 3.06 +/- 0.24 mm in the hyperopia group, and 3.72 +/- 0.26 mm and 3.73 +/- 0.23 mm in the myopia group. The standard deviation of the repeated Orbscan II measurements increased from 13 to 15 microm from the apex to the 3.0 mm zone. The difference between the apex and 3.0 mm zone of the cornea in all groups ranged from 0.1 to 0.12 mm. The mean AL was 23.52 +/- 0.82 mm in the emmetropia group, 22.14 +/- 0.64 mm in the hyperopia group, and 27.44 +/- 1.67 mm in the myopia group. There was a significant correlation between the spherical equivalent and AL (r = 0.94).
CONCLUSIONS: Significantly lower ACD values were found in the hyperopia group than in the other 2 groups. There was no difference in ACD between the emmetropia and myopia groups even though the AL in the myopia group was 4.0 mm longer. No statistical difference in ACD measurements was found between the Orbscan II and IOLMaster.

Entities:  

Mesh:

Year:  2003        PMID: 14670419     DOI: 10.1016/s0886-3350(03)00409-7

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  10 in total

1.  White-to-white corneal diameter, pupil diameter, central corneal thickness and thinnest corneal thickness values of emmetropic subjects.

Authors:  Juan A Sanchis-Gimeno; Daniel Sanchez-Zuriaga; Francisco Martinez-Soriano
Journal:  Surg Radiol Anat       Date:  2011-10-22       Impact factor: 1.246

2.  Anterior chamber width measurement by high-speed optical coherence tomography.

Authors:  Jason A Goldsmith; Yan Li; Maria Regina Chalita; Volker Westphal; Chetan A Patil; Andrew M Rollins; Joseph A Izatt; David Huang
Journal:  Ophthalmology       Date:  2005-02       Impact factor: 12.079

3.  Heritability and shared environment estimates for myopia and associated ocular biometric traits: the Genes in Myopia (GEM) family study.

Authors:  Christine Yi-Chin Chen; Katrina Jacqueline Scurrah; Jim Stankovich; Pam Garoufalis; Mohamed Dirani; Kelly Kathleen Pertile; Andrea Jane Richardson; Paul Mitchell; Paul Nigel Baird
Journal:  Hum Genet       Date:  2007-01-05       Impact factor: 4.132

4.  Comparison of techniques for measuring anterior chamber depth: Orbscan imaging, Smith's technique, and van Herick's method.

Authors:  Frank Eperjesi; Claire Holden
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-10       Impact factor: 3.117

5.  A longitudinal study of the effect of ocular biometrics measures on myopia onset.

Authors:  Zhengyang Tao; Hongwei Deng; Huahong Zhong; Yang Yu; Jun Zhao; Shuiqiang Chen; Shan Li; Minjuan Zhu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2020-11-17       Impact factor: 3.117

6.  Anterior segment study with the Pentacam Scheimpflug camera in refractive surgery candidates.

Authors:  Masih Hashemi; Khalil Ghasemi Falavarjani; Gholam Hosseyn Aghai; Kaveh Abri Aghdam; Arzhang Gordiz
Journal:  Middle East Afr J Ophthalmol       Date:  2013 Jul-Sep

7.  Agreement between Orbscan II, VuMAX UBM and Artemis-2 very-high frequency ultrasound scanner for measurement of anterior chamber depth.

Authors:  Haya Matuoq Al Farhan
Journal:  BMC Ophthalmol       Date:  2014-02-25       Impact factor: 2.209

8.  Effect of anterior chamber depth on the choice of intraocular lens calculation formula in patients with normal axial length.

Authors:  Mohammad Miraftab; Hassan Hashemi; Akbar Fotouhi; Mehdi Khabazkhoob; Farhad Rezvan; Soheila Asgari
Journal:  Middle East Afr J Ophthalmol       Date:  2014 Oct-Dec

9.  The distribution of orbscan indices in young population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Negareh Yazdani; Hadi Ostadimoghaddam; Payam Nabovati; Raheleh Moravej; AbbasAli Yekta
Journal:  J Curr Ophthalmol       Date:  2016-08-31

10.  Anterior Segment Changes during Accommodation in Accommodative Esotropia.

Authors:  Mustafa Koç; Hakan Halit Yaşar; Mehmet Murat Uzel; Salih Çolak; Irfan Durukan; Pelin Yılmazbaş
Journal:  Korean J Ophthalmol       Date:  2018-01-25
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.