Muriël Doors1, Lars P J Cruysberg, Tos T J M Berendschot, John de Brabander, Frenne Verbakel, Carroll A B Webers, Rudy M M A Nuijts.
Abstract
BACKGROUND: The repeatability and interchangeability of imaging devices measuring central corneal thickness (CCT) and anterior chamber depth (ACD) are important in the assessment of patients considering refractive surgery. The purpose of this study was to investigate the agreement of CCT and ACD measurements using three imaging technologies in healthy eyes and in eyes after phakic intraocular lens implantation (pIOL).Entities:
Mesh:
Year: 2009 PMID: 19404662 PMCID: PMC2709228 DOI: 10.1007/s00417-009-1086-6
Source DB: PubMed Journal: Graefes Arch Clin Exp Ophthalmol ISSN: 0721-832X Impact factor: 3.117
Pairwise comparison of central corneal thickness measurements for the healthy volunteers (healthy) and phakic intraocular lens (pIOL) group using anterior segment optical coherence tomography (AS-OCT), orbscan II, and pentacam
| Pairwise comparison | Mean difference ± SD (μm) | Pearson correlation coefficient* ( | 95% limits of agreement | |||
|---|---|---|---|---|---|---|
| Healthy | pIOL | Healthy | pIOL | Healthy | pIOL | |
| Orbscan II/ Pentacam | −16.4 ± 12.4 | −2.8 ± 13.3 | 0.902 | 0.931 | 8.0 to −40.7 | 23.3 to −28.9 |
| Orbscan II/ AS-OCT | 1.8 ± 14.2 | 9.7 ± 12.8 | 0.867 | 0.938 | 29.7 to −26.1 | 34.8 to −15.5 |
| AS-OCT/ Pentacam | −19.2 ± 8.5 | −14.1 ± 11.6 | 0.947 | 0.958 | −2.6 to −35.8 | 8.7 to −36.8 |
| Automatic SS/manual SS | −13.9 ± 11.8 | NA | 0.901 | NA | 9.1 to −37.0 | NA |
| Automatic SS/manual HR | −6.21 ± 7.6 | NA | 0.954 | NA | 8.7 to −21.1 | NA |
| Manual SS/manual HR | 7.73 ± 10.5 | NA | 0.922 | NA | 28.3 to −12.8 | NA |
SD = standard deviation; SS = anterior segment single-scan; HR = high-resolution scan
*All with P < 0.001
Fig. 1Bland-Altman plots of the differences between anterior segment optical coherence tomography (AS-OCT), Orbscan II and Pentacam central corneal thickness measurements in the healthy volunteer group and the phakic intraocular lens (pIOL) group with P-value of the paired t-test. All scales in μm
Pairwise comparison of anterior chamber depth measurements for the healthy volunteers (healthy) and phakic intraocular lens (piol) group using anterior segment optical coherence tomography (AS-OCT), Orbscan II, and Pentacam
| Pairwise comparison | Mean difference ± SD (mm) | Pearson correlation coefficient* ( | 95% limits of agreement | |||
|---|---|---|---|---|---|---|
| Healthy | pIOL | Healthy | pIOL | Healthy | pIOL | |
| Orbscan II/ Pentacam | –0.08 ± 0.04 | −0.10 ± 0.04 | 0.990 | 0.983 | 0.00 to −0.17 | −0.02 to −0.19 |
| Orbscan II/ AS-OCT | −0.15 ± 0.05 | −0.21 ± 0.05 | 0.987 | 0.981 | −0.06 to −0.24 | −0.12 to −0.30 |
| AS-OCT/ Pentacam | 0.07 ± 0.04 | 0.10 ± 0.08 | 0.990 | 0.946 | 0.15 to −0.01 | 0.25 to −0.06 |
SD = standard deviation
*All with P < 0.001
Fig. 2Bland-Altman plots of the differences between anterior segment optical coherence tomography (AS-OCT), Orbscan II and Pentacam anterior chamber depth measurements in the healthy volunteer group and the phakic intraocular lens (pIOL) group with P-value of the paired t-test. All scales in mm
Intraobserver reliability from three consecutive central corneal thickness measurements of the healthy volunteer group
| Device | Overall Mean ± SD (μm) | SDwithin ± SD | CV ± SD (%) | Precision ± SD |
|---|---|---|---|---|
| AS-OCT | ||||
| Automatic segment single | 531.1 ± 25.1 | 3.8 ± 2.7 | 0.7 ± 0.5 | 7.5 ± 5.4 |
| Manual segment single | 545.3 ± 25.1 | 6.8 ± 6.0 | 1.2 ± 1.1 | 13.3 ± 11.7 |
| Manual high resolution | 538.8 ± 24.4 | 3.3 ± 4.1 | 0.6 ± 0.7 | 6.5 ± 8.0 |
| Orbscan II | 530.3 ± 40.8 | 5.4 ± 3.5 | 1.0 ± 0.7 | 10.5 ± 6.8 |
| Pentacam | 551.6 ± 25.6 | 4.0 ± 2.3 | 0.7 ± 0.4 | 7.9 ± 4.6 |
SD = standard deviation; SDwithin = mean standard deviation between three consecutive measurements; CV = coefficient of variation; AS-OCT = anterior segment optical coherence tomography