Comparison of central corneal thickness measurements by specular microscopy, ultrasound pachymetry, and ultrasound biomicroscopy.

Eric S Tam1, David S Rootman.   

Abstract

PURPOSE: To compare the reproducibility and mean values of central corneal thickness (CCT) obtained by specular microscopy, ultrasound pachymetry, and ultrasound biomicroscopy (UBM).
SETTING: Department of Ophthalmology, University of Toronto, Toronto, Ontario, Canada.
METHODS: Thirty-one healthy volunteers were recruited for a sample size of 62 eyes. All subjects had pachymetric measurements by specular microscopy, ultrasound pachymetry, and UBM. Three separate measurements meeting criteria for centrality and perpendicularity were recorded for each eye.
RESULTS: The mean CCT by specular microscopy was 572 microm (95% confidence interval (CI), 566-578 microm), which was significantly greater than 550 microm (95% CI, 545-556 microm) (P<.001) and 555 microm (95% CI, 550-560 microm) (P<.001) by ultrasound pachymetry and UBM, respectively. The mean standard deviation (SD) of repeated measurements by specular microscopy was 7.82 microm, which was significantly greater than the mean SDs of 4.14 microm (P<.001) and 3.90 microm (P<.001) by ultrasound pachymetry and UBM, respectively. There was no statistically significant difference between the mean SDs by ultrasound pachymetry and UBM (P=.156).
CONCLUSIONS: Although the CCT measurements by specular microscopy were significantly less reproducible than those by ultrasound pachymetry and UBM, the error levels were clinically acceptable. Both ultrasound pachymetry and UBM produced similar CCT measurements, which were significantly less than those generated by specular microscopy. One should be aware of the advantages and limitations of each machine and of possible differences in the CCT measurements by optical and ultrasound pachymetry.

Entities:  

Mesh:

Year:  2003        PMID: 12842687     DOI: 10.1016/s0886-3350(02)01921-1

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  20 in total

Review 1.  Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?

Authors:  Aristides Konstantopoulos; Parwez Hossain; David F Anderson
Journal:  Br J Ophthalmol       Date:  2007-04       Impact factor: 4.638

2.  Comparison of ultrasonic pachymetry with Orbscan in corneal haze.

Authors:  Rana Altan-Yaycioglu; Aysel Pelit; Yonca Aydin Akova
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-06-22       Impact factor: 3.117

3.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

4.  Explaining The Current Role Of High Frequency Ultrasound In Ophthalmic Diagnosis (Ophthalmic Ultrasound).

Authors:  D Jackson Coleman; Ronald H Silverman; Mark J Rondeau; Harriet O Lloyd; Suzanne Daly
Journal:  Expert Rev Ophthalmol       Date:  2006-10-01

5.  Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry.

Authors:  Zaina N Al-Mohtaseb; Li Wang; Mitchell P Weikert
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-02-17       Impact factor: 3.117

6.  Repeatability of central corneal thickness measurements measured with the Topcon SP2000P specular microscope.

Authors:  Kelechi C Ogbuehi; Turki M Almubrad
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-15       Impact factor: 3.117

7.  Ultrasonic pachymeter: a little different role.

Authors:  Alok Sati; Deepak Kalra; Rakesh Maggon; Rajneesh Sinha
Journal:  Med J Armed Forces India       Date:  2011-10-22

8.  Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.

Authors:  Saleh Al-Ageel; Abdulrahman M Al-Muammar
Journal:  Saudi J Ophthalmol       Date:  2009-10-24

Review 9.  High-resolution ultrasound imaging of the eye - a review.

Authors:  Ronald H Silverman
Journal:  Clin Exp Ophthalmol       Date:  2008-12-09       Impact factor: 4.207

10.  Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system.

Authors:  Tania Paul; Mira Lim; Christopher E Starr; Harriet O Lloyd; D Jackson Coleman; Ronald H Silverman
Journal:  J Cataract Refract Surg       Date:  2008-11       Impact factor: 3.351

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.