Resistance of three implant-abutment interfaces to fatigue testing.

Cleide Gisele Ribeiro1, Maria Luiza Cabral Maia, Susanne S Scherrer, Antonio Carlos Cardoso, H W Anselm Wiskott.   

Abstract

UNLABELLED: The design and retentive properties of implant-abutment connectors affect the mechanical resistance of implants. A number of studies have been carried out to compare the efficacy of connecting mechanisms between abutment and fixture.
OBJECTIVES: The aims of this study were: 1) to compare 3 implant-abutment interfaces (external hexagon, internal hexagon and cone-in-cone) regarding the fatigue resistance of the prosthetic screw, 2) to evaluate the corresponding mode of failure, and 3) to compare the results of this study with data obtained in previous studies on Nobel Biocare and Straumann connectors.
MATERIALS AND METHODS: In order to duplicate the alternating and multivectorial intraoral loading pattern, the specimens were submitted to the rotating cantilever beam test. The implants, abutments and restoration analogs were spun around their longitudinal axes while a perpendicular force was applied to the external end. The objective was to determine the force level at which 50% of the specimens survived 10(6) load cycles. The mean force levels at which 50% failed and the corresponding 95% confidence intervals were determined using the staircase procedure.
RESULTS: The external hexagon interface presented better than the cone-in-cone and internal hexagon interfaces. There was no significant difference between the cone-in-cone and internal hex interfaces.
CONCLUSION: Although internal connections present a more favorable design, this study did not show any advantage in terms of strength. The external hexagon connector used in this study yielded similar results to those obtained in a previous study with Nobel Biocare and Straumann systems. However, the internal connections (cone-in-cone and internal hexagon) were mechanically inferior compared to previous results.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21710094      PMCID: PMC4223795          DOI: 10.1590/s1678-77572011005000018

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


INTRODUCTION

Dental implantology has revolutionized the treatment for edentulous and partially edentulous patients, and successful implant integration has been well documented for patients with those clinical conditions. With the high rate of implant success for edentulous individuals, the concept of osseointegration and implant therapy has flourished as a predictable treatment modality[10]. Clinical observations have indicated that the major causes of implant failure are (a) deficient osseointegration, (b) complications of the neighboring soft tissues (peri-mucositis and periimplantitis) and (c) mechanical complications[13]. Among the biomechanical problems, screw loosening, abutment rotation, and abutment fracture are the major issues[15,28]. In a prospective multicenter investigation, Henry, et al.[12] (1996) evaluated 92 patients with 107 implants and found that the problems most frequently experienced during the first year were related to lose screws. The two mechanisms involved in screw loosening are: excessive bending (plastic deformation that takes place when a load larger than the yield strength of the screw is applied) and settling (when external loads applied to the screw interface create micromotion between both surfaces). As the mating surfaces wear, they "settle" closer together[15]. The factors that contribute to screw instability are: misfit of the prosthesis, insufficient tightening force, screw settling, mechanical overload, and mismatch in screw material and design[2]. A number of studies have been conducted to compare the efficacy of different connecting mechanisms securing the abutment to the implant head[16]. The design of the implant-to-abutment mating surface and the retentive properties of the screw joints affect the mechanical resistance of the implant-abutment complex[3,20]. The implant-abutment connection is also influenced by factors such as component fit, machining accuracy, saliva contamination and screw preload[4]. Current designs are derived from two basic designs: the "butt-joint", consisting of 2 parallel flat contacting surfaces[3], and the internal "cone-in-cone" design. The latter has been introduced in the ITI implant system (Institute Straumann AG, Waldenburg, Basel, Switzerland) and offered a sound, stable, and self-locking interface[29]. Recent studies have indicated a potential mechanical advantage of conical connectors over butt-joint designs[20]. Indeed, the mechanics of the ITI cone-in-cone[20] resulted in lower incidences of mechanical complications, specifically abutment screw loosening and fracture, in comparison with those reported for butt-joint implants[23,25]. With few exceptions, most of the long-term clinical data on implant performance involve external hexagons. This is primarily the result of their extensive use, the broad number of prescribed clinical applications, the level of complications reported, and the resulting efforts to find solutions (specific torque application to abutment screws)[3]. Industry surveys have shown that external hex implants still dominate the European market[18]. Fatigue is a progressive, localized and permanent structural damage that occurs in a material subjected to repeated or fluctuating strains. Experimentally, three modes of loading may be used to duplicate fatigue failures: direct axial loading (the specimen is submitted to a uniform stress through its cross-section), plane-bending (the majority of the specimen is subjected to a uniform bending stress) and rotating-beam loading (the specimen is rotation-symmetric and is subjected to dead-weight loading while swivel bearings permit rotation)[24]. In order to duplicate the multivectorial force pattern of the mouth, a laboratory test has been developed by Wiskott, et al.[34] (2004) using the rotating beam principle. The test consists in spinning a specimen while holding it at one end and loading it at the protruding end. The samples are thus subjected to a 360-degree field of transverse tensile and compressive force vectors. Actuator-driven fatigue testing systems are unable to reproduce the complex force patterns that are active clinically. Hence data obtained using rotational fatigue testing have a superior pertinence relative to single-axis testing designs[32-33]. To overcome some of the inherent design limitations of the external hex connector, a variety of alternative connections have been developed. The goals of this study were: (1) to evaluate the fatigue resistance of 3 implant-abutment connectors (external hexagon, internal hexagon and cone-in-cone) analyzing the prosthetic screw; 2) to determine their failure modes; and (3) to compare the obtained results with previous data generated from Nobel Biocare-Replace and Straumann-ITI connectors.

MATERIAL AND METHODS

Three geometries of implant-abutment interfaces were evaluated. Thirty implants (4.0 mm diameter and 13 mm long) of each connector type were connected to Micro-unit abutments (Conexão Sistemas de Prótese, Arujá, SP, Brazil) and torqued to 30 Ncm using a calibrated torque controller. The Micro-unit abutments are industrially machined prosthetic components that are intended for use in fixed partial and complete implant-supported dentures at all sulcus depths and for all platforms. They were designed to provide versatility while optimizing the esthetics of multiple unit, screw retained, restorations. Therefore, the sole differences between groups were the variations in the interface geometry between the implant head and the Micro-unit abutment. The cone-in-cone Micro-unit abutment used in this study presents an internal modification (internal hexagon), which is located at the bottom of the cone to allow the angular repositioning of the abutments. The groups were set up as follows: group A (external hex implant+micro-unit abutment+restoration analog); group B (cone-in-cone implant+micro-unit abutment+restoration analog); group C (internal hex implant+micro-unit abutment+restoration analog). In order to duplicate the mouth's multivectorial force pattern, the specimens (implant, abutment and restoration analog) were configured as rotating cantilever beams (Figure 1). The rotating beam principle demands that a concentric arrangement be established between all the components. One end of the test specimen is clamped into a collet and rotated, while a perpendicular force is applied to the other end via a ball bearing. This perpendicular force submits the specimens to alternating sinusoidal tension-compression stresses which, depending on the magnitude of the load applied, cause breakage of the components within a predetermined number of cycles. The fatigue resistance of the connectors is expressed as the force level at which 50% of the specimens survive 106 load cycles without breakage and 50% fail.
Figure 1

Schematic drawing of the sample (internal hexagon interface- group C)

Schematic drawing of the sample (internal hexagon interface- group C)

Restoration analog

The Micro-unit abutments used were 1-mm-high collar platforms. To allow valid comparisons with previous data, the restoration analog was 20 mm in length. This provided a 11.3 mm distance between the midplane of the ball bearing and the emergence of the implant from the collet. The torque recommended by the manufacturer was 20 Ncm for the abutment screw and 20 Ncm for the prosthetic screw. At this torque, however, both screws loosened during the course of the experiment. It was therefore decided to torque the abutment screw to 30 Ncm and the prosthetic screw to 25 Ncm to induce failure by screw breakage and not by screw loosening.

Experimental procedure and data analysis

The three implant-abutment interfaces were evaluated regarding their fatigue resistance at 106 cycles (an arbitrarily set number whose theoretical and practical basis has been previously explained)[33]. The experimental procedure required that a number of specimens be tested in sequence. To this effect, the specimens were loaded via the ball bearing and spun at 1.000 rpm (16.7 Hz). After 106 cycles, the experimenter checked whether the specimen was intact or whether it had broken. If it was intact, the next specimen was loaded at the previous magnitude plus 5 N. The same force (5 N) was subtracted from the former load magnitude if the previous specimen had failed. This leads to the characteristic up-and-down pattern of run-outs and failures that characterizes the staircase procedure. After suitable arrangement of the data, the mean F50 (at which 50% of the samples failed and 50% ran out) and the standard deviation were calculated (Table 1). When applying the staircase procedure, the examiner must set an appropriate force increment (or decrement) (F) - 5 N in the present test series. If it is too large, the test looses its discriminating potential. In this experiment, F was taken from previous studies[32,34].
Table 1

Example of data arrangement for staircase analyses (external hexagon interface)

Applied forces in newtonsForce level (i)# of failures (nI)i nIi2 nI
6551525
60441664
55361854
502124
451111
400000
  n=13A=42B=148

n=Σ nI , A=Σ inI, B=Σ i2nI

Example of data arrangement for staircase analyses (external hexagon interface) n=Σ nI , A=Σ inI, B=Σ i2nI During testing, the results were graphically charted as in Figure 2. After all tests had been completed, they were arranged as shown in Table 1. Taking A and B from Table 1, F50 was calculated as
Figure 2

Staircase data for the implant abutment interfaces analyzed in the study

Staircase data for the implant abutment interfaces analyzed in the study with: + if the test is based on run-outs, - if the test is based on failures Whenever the number of run-outs and failures differed, data analysis was based on the least frequent event. The corresponding standard deviation was taken as: and Where F50 was the mean force level at which 50% of specimens ran-out and 50% failed; F0 was the lowest load level at which failure occurred; Fincr was the chosen force increments or decrement, that is, 5 N; n=Σ ni (ni : the number of failures of each load level) (see table 1); A=Σ ini (i being the load level) and B=Σ i2ni. To assess whether the F50's of each group were significantly different, the means were fitted with 95% confidence intervals using a method described by Collins[9] (1993). Means with overlapping intervals were considered equivalent.

Stereomicroscope examination and scanning electron microscopy (SEM)

Ten prosthetic screws of each interface were randomly selected and evaluated using a stereomicroscope (Wild M3Z, Heerbrugg, Switzerland) to inspect thread wear, defects and the fractured surfaces at low magnification. Stereomicroscopy is often used to conduct preliminary observations of fractured components. After this evaluation, 3 samples of fractured screws of each interface were gold-sputtered and examined with a scanning electron microscope (Philips XL Series - XL 20; Philips, Eindhoven, The Netherlands).

RESULTS

The fatigue resistance for each connector expressed as the mean force level at which 50% of the samples survived 106 cycles and 50% failed (F50) is shown in Table 2, which also details each mean's 95% CI. Statistically, the external hexagon interface presented superior result compared to the conical and internal hexagon interfaces. There was no significant difference between the conical and internal hex interfaces.
Table 2

Fatigue resistance of the connectors subjected to the rotating-bending test

 F50SDUpperLower
External hex interface53.57.8049.557.5
Cone-in-cone interface44.02.4942.345.7
Internal hex interface45.03.4043.146.9

F50=mean failure level (force level at which 50% of the samples survive and 50% fail before 106 cycles). When the interfaces with overlapping CIs were combined, 1 group was identified: group A - external hex interface. SD= standard deviation

Fatigue resistance of the connectors subjected to the rotating-bending test F50=mean failure level (force level at which 50% of the samples survive and 50% fail before 106 cycles). When the interfaces with overlapping CIs were combined, 1 group was identified: group A - external hex interface. SD= standard deviation Analysis of the fractured screws stereomicroscopy and SEM revealed that the mode and the region of fracture were the same for 24 of the 30 screws evaluated (fracture of the threaded part - Figures 3, 4). Six screws presented damages in the last two threads but no fracture (Figure 5). Fatigue striations were seen on the SEM micrographs (Figure 6). Such striations are an absolute indication of fatigue failure. "Overload" or fast fracture zone, that is, the portions of the components where final catastrophic failure occurred were also seen. The surface structure of this zone was similar for all groups (Figures 7A-C).
Figure 3

Stereomicroscope image of the prosthetic screw at the fractured surface

Figure 4

Scanning electron microscopy (SEM) micrograph of the screw abutment at the fractured surface

Figure 5

Scanning electron microscopy (SEM) micrograph demonstrating damages in the screw threads without fracture

Figure 6

Scanning electron microscopy (SEM) micrograph showing fatigue striations

Figure 7

Scanning electron microscopy (SEM) micrograph images of prosthetic screws showing the same mode of fracture to all the types of implant-abutment interfaces (acone- in-cone; b- external hexagon and c- internal hexagon)

Stereomicroscope image of the prosthetic screw at the fractured surface Scanning electron microscopy (SEM) micrograph of the screw abutment at the fractured surface Scanning electron microscopy (SEM) micrograph demonstrating damages in the screw threads without fracture Scanning electron microscopy (SEM) micrograph showing fatigue striations Scanning electron microscopy (SEM) micrograph images of prosthetic screws showing the same mode of fracture to all the types of implant-abutment interfaces (acone- in-cone; b- external hexagon and c- internal hexagon)

DISCUSSION

The effect of connector design on the mechanical resistance of a dental implant screw joint is still fraught with uncertainties. This is demonstrated by the numerous configurations available in today's market. Several systems are in clinical use, most notably the external hexagon, the internal hexagon and the tapered joints. According to Binon[3] (2000), contemporary implant systems are configured with about 20 different implant/abutment interface geometries. Each implant-abutment interface has its pros and cons. According to Maeda, et al.[16] (2006), the external hex interface has advantages such as suitability for the two stage method, provision of an anti-rotation mechanism, retrievability and compatibility among different systems[6]. The external hex interface provides more versatility for the laboratory technician in solving problems related to emergence profile and esthetics, since the technician is able to bring the porcelain of a porcelain- fused-to-gold crown closer to the implant interface[6]. However, increased screw loosening, component fracture, and difficulty in seating abutments in deep subgingival tissues are problems commonly experienced with external hexagon connectors[29]. Regarding the internal hex system, according to Maeda, et al.[16] (2006), its advantages are: ease in abutment connection, suitability for one stage implant installation, higher stability and suitability for single-tooth restoration, higher resistance to lateral loads due to the lower centre of rotation and better force distribution[6]. A systematic review conducted by Theoharidou, et al.[31] (2008) demonstrated stable abutment screw connections for internal-connection implants as well for external-connection implants with improved screw materials (altering the screw alloys and their surfaces) and preload. Tapered joint connections with a conical interface have advantages of better sealing capacity in closing the micro-gap on top of those in an internal hex system. Most in vitro studies have demonstrated that internal connections are more stable mechanically than external flat connections[16,20]. The general focus is clearly on deep internal connections, in which the screw takes little or no load and provides intimate contact with the implants walls to resist micromovement[3]. The present data are in agreement with those of Piermatti, et al.[22] (2006), who reported inferior results for the internal hex connections when compared to external connections. Steinebrunner, et al.[26] (2008) evaluated the influence of long-term dynamic loading on the fracture strength of different implant-abutment connectors. External hex connections yielded better results compared to internal hex connections. Also, the internal tube-and-tube connections with a cam indexing system obtained the superior results with regard to longevity and fracture strength. Former research was able to show that there is a direct correlation between the amount of misfit of the components and screw loosening[4]. Binon and McHugh[4] (1996) pointed towards manufacturing tolerances as a reason for the screw loosening of the prefabricated parts and requested manufacturers to improve the fit of their implant components. Preload protects the screw from breakage during cyclic loading. If the joint is compressed, preload will be lost, the screw and the interface are subjected to plastic deformation and the joint may separate[35]. The optimum preload force recommended for an implant screw is 60-80% of the yield strength of the material from which the screw is machined. At stresses at or beyond yield, the screw will function in its plastic deformation zone with resulting loss of preload and inefficient function. Conversely, stresses within the elastic region of the material are the most appropriate to resist the separation forces induced during occlusal loading[21]. Thus, the greater the clamping force (preload), the tighter the clamped joint. However, preload values should not be too high and should be within the elastic domain, else retaining screws may yield or break under repeated functional bite forces[1]. The torque used in this study was 30 Ncm in the abutment screw and 25 Ncm in the prosthetic screw. SEM analysis of screws tightened to 25 Ncm and to 40 Ncm demonstrated no damages in the screw morphology thereby indicating that the torque applied was below the elastic limit of the material. The yield strength and the breakage strength of screws are not commonly reported by manufacturers. The literature provides an abundance of studies that analyzed the fatigue resistance of dental implants and prosthetic components. However, there was no standardization of the applied forces 300 N[1,35]; 100-150 N[7]; 10-250 N[17]; 20-200 N[8]; 100-450 N[11]; 50 N[27]; 120 N[26] and in the mode of loading (angle of load application) and simple (fatigue only) or combined (fatigue plus monotonic load). The loading frequencies were different also. The present study was carried out using comparable implant diameters, identical abutments and levers therefore rendering intergroup comparisons possible. The results demonstrated that the fatigue strength of the external hex interface used in this study was of the same magnitude as the Nobel Biocare Replace Select, multi-unit abutment and the Straumann ITI, standard abutment. The internal connections though (conical and internal hexagon) revealed inferior results compared to the previous data[32,34]. The prosthetic screw that connects the fixed dental prosthesis to the abutment is intended as the weak link, that is, in case of occlusal overload, it is designed to break first and thus protect the implant and the bone from damage due to excessive stresses[5,23]. This is supported by the finding that the incidence of abutment screw and implant fracture is much lower than that of prosthetic screw loosening or fracturing[5,19]. Conversely, according to Sutter, et al.[29] (1993), in the two-stage system it is the abutment screw that most frequently fractures. This apparent incrongruity of the more massive abutment screw failing before the smaller occlusal screw might be explained by simple mechanics. In a two-stage system, the abutment screw secures the abutment to the implant. This interface is subjected to a higher level of stress because it is located near the alveolar crest, that is, where the applied lever is greatest. The abutment screw therefore, is subjected to much greater forces than the occlusal screw when the force vectors are nonaxial in nature. It is thus more susceptible to fatigue failure, although it is a more massive structure[30]. The present study confirms that the prosthetic screw fails more frequently than the abutment screw and failure varies according to the type of interface analyzed.

CONCLUSIONS

Within the limitation of this study, the following conclusions can be drawn: 1. This study demonstrated the superior fatigue resistance of external hex interface. There was no significant difference between the conical and internal hex interfaces. Probably, the quality of the surface machining of the flat-to-flat mating surfaces (mainly, the machining accuracy of the screw and thread) determined the superior resistance of the connector; 2; The mode and region of fracture in prosthetic screws observed in this study suggested that failure of these screws occurred by fatigue (presence of fatigue striations) and involved the threaded part; 3. The present tests demonstrated that the fatigue strength of the external hex interface used in this study was of comparable strength as that determined in a previous study on Nobel Biocare and Straumann implants when similar abutments and level torque were used. It is important to emphasize that the prosthetic screw used in this study was designed specifically to support a higher torque (25 Ncm) than the conventional torque used for this screw (10 Ncm). In addition, the micro-unit abutment received a higher torque than the one recommended by the manufacturer; 4. The internal connections (cone-in-cone and internal hexagon) had inferior results compared to those found in previous results. Internal connections require accurate machining and tolerances and the reason for the present results may be a lack of precision of the components that allowed micromovement at the connector interface.
  31 in total

Review 1.  Implants and components: entering the new millennium.

Authors:  P P Binon
Journal:  Int J Oral Maxillofac Implants       Date:  2000 Jan-Feb       Impact factor: 2.804

2.  An in vitro analysis of implant screw torque loss with external hex and internal connection implant systems.

Authors:  Jack Piermatti; Hoda Yousef; Allyn Luke; Richard Mahevich; Saul Weiner
Journal:  Implant Dent       Date:  2006-12       Impact factor: 2.454

3.  Influence of connection geometry on dynamic micromotion at the implant-abutment interface.

Authors:  Joe C Meng; Joshua E Everts; Fang Qian; David G Gratton
Journal:  Int J Prosthodont       Date:  2007 Nov-Dec       Impact factor: 1.681

4.  A classification system to measure the implant-abutment microgap.

Authors:  Stefania C Kano; Paul P Binon; Donald A Curtis
Journal:  Int J Oral Maxillofac Implants       Date:  2007 Nov-Dec       Impact factor: 2.804

Review 5.  Abutment screw loosening in single-implant restorations: a systematic review.

Authors:  Anna Theoharidou; Haralampos P Petridis; Konstantinos Tzannas; Pavlos Garefis
Journal:  Int J Oral Maxillofac Implants       Date:  2008 Jul-Aug       Impact factor: 2.804

6.  Zirconium implant abutments: fracture strength and influence of cyclic loading on retaining-screw loosening.

Authors:  Peter Gehrke; Günter Dhom; Jochen Brunner; Dietrich Wolf; Marco Degidi; Adriano Piattelli
Journal:  Quintessence Int       Date:  2006-01       Impact factor: 1.677

7.  Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 4: Failure analysis of 10 fractured retaining screws retrieved from three patients.

Authors:  Youssef S Al Jabbari; Raymond Fournelle; Gerald Ziebert; Jeffrey Toth; Anthony M Iacopino
Journal:  J Prosthodont       Date:  2008-01-15       Impact factor: 2.752

8.  Implant-abutment interface design affects fatigue and fracture strength of implants.

Authors:  Lars Steinebrunner; Stefan Wolfart; Klaus Ludwig; Matthias Kern
Journal:  Clin Oral Implants Res       Date:  2008-12       Impact factor: 5.977

Review 9.  Mechanical complications of dental implants.

Authors:  M S Schwarz
Journal:  Clin Oral Implants Res       Date:  2000       Impact factor: 5.977

10.  Preload and torque removal evaluation of three different abutment screws for single standing implant restorations.

Authors:  Rafael Augusto Stüker; Eduardo Rolim Teixeira; João Carlos Pinheiro Beck; Nilza Pereira da Costa
Journal:  J Appl Oral Sci       Date:  2008 Jan-Feb       Impact factor: 2.698

View more
  9 in total

1.  Comparative evaluation of the wear resistance of two different implant abutment materials after cyclic loading - An in vitro study.

Authors:  Maniamuthu Ragupathi; Vallabh Mahadevan; N S Azhagarasan; Hariharan Ramakrishnan; S Jayakrishnakumar
Journal:  Contemp Clin Dent       Date:  2020-11-26

Review 2.  Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review.

Authors:  Marcelo Coelho Goiato; Eduardo Piza Pellizzer; Emily Vivianne Freitas da Silva; Liliane da Rocha Bonatto; Daniela Micheline dos Santos
Journal:  Oral Maxillofac Surg       Date:  2015-04-25

3.  Reliability evaluation of alumina-blasted/acid-etched versus laser-sintered dental implants.

Authors:  Erika O Almeida; Amilcar C Freitas Júnior; Estevam A Bonfante; Nelson R F A Silva; Paulo G Coelho
Journal:  Lasers Med Sci       Date:  2012-07-28       Impact factor: 3.161

4.  Micro-leakage at the implant-abutment interface with different tightening torques in vitro.

Authors:  João Paulo da Silva-Neto; Marcel Santana Prudente; Thiago de Almeida Prado Naves Carneiro; Mauro Antônio de Arruda Nóbilo; Mario Paulo Amante Penatti; Flávio Domingues das Neves
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

5.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

6.  Influence of Connection Type and Platform Diameter on Titanium Dental Implants Fatigue: Non-Axial Loading Cyclic Test Analysis.

Authors:  Ana I Nicolas-Silvente; Eugenio Velasco-Ortega; Ivan Ortiz-Garcia; Alvaro Jimenez-Guerra; Loreto Monsalve-Guil; Raul Ayuso-Montero; Javier Gil; Jose Lopez-Lopez
Journal:  Int J Environ Res Public Health       Date:  2020-12-02       Impact factor: 3.390

7.  In vitro corrosion studies of stainless-steel dental substrates during Porphyromonas gingivalis biofilm growth in artificial saliva solutions: providing insights into the role of resident oral bacterium.

Authors:  Ubong Eduok; Jerzy Szpunar
Journal:  RSC Adv       Date:  2020-08-24       Impact factor: 4.036

8.  Mechanical testing of thin-walled zirconia abutments.

Authors:  Luigi Canullo; Paulo G Coelho; Estevam A Bonfante
Journal:  J Appl Oral Sci       Date:  2013 Jan-Feb       Impact factor: 2.698

9.  Effect of lateral oblique cyclic loading on microleakage and screw loosening of implants with different connections.

Authors:  Hakimeh Siadat; Hossain Najafi; Marzieh Alikhasi; Babak Falahi; Elaheh Beyabanaki; Farid Zayeri
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2018-09-18
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.