Resistance of internal-connection implant connectors under rotational fatigue loading.

H W Wiskott1, Robin Jaquet, Susanne S Scherrer, Urs C Belser.   

Abstract

PURPOSE: To aid in developing mechanically optimized implant-abutment connectors, the fatigue resistance of 5 connector configurations of the Replace Select system (Easy abutment, Easy abutment without antirotational mechanism, Multi-unit abutment, Esthetic Alumina abutment, Esthetic Zirconia abutment) was investigated. Other purposes of the study were to determine whether the connector's antirotational mechanism participates in fatigue resistance and to compare the results with previous data on Straumann connectors.
MATERIALS AND METHODS: The repetitive, alternating, and multivectorial intraoral force pattern was reproduced by subjecting the test specimens to the rotating cantilever beam test. To this end, the samples were spun around their long axis while clamped into a revolving collet on one end and loaded normal to their long axis on the other end. The aim was to determine the load level at which 50% of the specimens survived and 50% fractured before 106 cycles. Means were determined using the staircase procedure. They were fitted with 95% confidence intervals for intergroup comparisons.
RESULTS: In the chosen testing configuration, 2 statistical groups emerged. The Easy abutments with and without antirotational mechanism were statistically similar, with mean failure loads in the 70 to 72 N range. Both ceramic and the Multi-unit abutments belonged to the second group, with mean failure loads in the 53 to 58 N range.
CONCLUSIONS: (1) The fatigue resistance of ceramic and the Multi-unit abutments was approximately 20% less than that of the Easy Abutments. (2) The antirotational mechanism did not participate in mechanical resistance. (3) The fatigue strength of the Easy abutment connectors was approximately 20% greater than the equivalent abutments in the Straumann system.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17465350

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  4 in total

1.  Mechanical analysis of prosthetic bars and dental implants in 3 and 4 implant-supported overdenture protocols using finite element analysis.

Authors:  Luiz Bassi-Junior; Rafael Oliveira de Souza Silva; Victor Hugo Dias Dos Santos; Abner da Rocha Lourenço; Paulo Vinicius Trevizoli; Hugo Gaêta-Araujo; Polyane Mazucatto Queiroz; Vilmar Divanir Gottardo
Journal:  J Oral Biol Craniofac Res       Date:  2021-05-15

2.  [Clinical classification and treatment decision of implant fracture].

Authors:  Y Li; H J Yu; L X Qiu
Journal:  Beijing Da Xue Xue Bao Yi Xue Ban       Date:  2022-02-18

3.  Resistance of three implant-abutment interfaces to fatigue testing.

Authors:  Cleide Gisele Ribeiro; Maria Luiza Cabral Maia; Susanne S Scherrer; Antonio Carlos Cardoso; H W Anselm Wiskott
Journal:  J Appl Oral Sci       Date:  2011-07-01       Impact factor: 2.698

4.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.