Zirconium implant abutments: fracture strength and influence of cyclic loading on retaining-screw loosening.

Peter Gehrke1, Günter Dhom, Jochen Brunner, Dietrich Wolf, Marco Degidi, Adriano Piattelli.   

Abstract

OBJECTIVE: The purpose of this study was to determine the fracture strength of zirconium implant abutments and the torque required to unfasten the retaining screw before and after applying cyclic loading to the implant-abutment assembly. The dynamic behavior and stress distribution pattern of zirconium abutments were also evaluated. METHODS AND MATERIALS: Static and cyclic loading of 7 XiVE implants with straight Cercon zirconium abutments were simulated under worst-case conditions. Cyclic loading tests were performed via a servohydraulic dynamic testing machine at loads between 100 and 450 N, for up to 5 million loading cycles, at 15 Hz. The dynamic behavior of the zirconium abutments was analyzed by finite element modeling and Pro/Mechanica software, comparing van-der-Mises and maximum stress levels.
RESULTS: Cercon zirconium-ceramic abutments exhibited a maximum fracture strength of 672 N during static loading and 269 N at 800,000 to 5 million cycles runout point, and 403 N at 10,000 cycles runout point during cyclic loading. The mean torque value required to unfasten the abutment retaining screws after (initial) tightening was 21 Ncm +/- 1 and 20 Ncm +/- 1 (measurement accuracy +/- 2 Ncm) after loading with up to 5 million cycles respectively. Torque values decreased minimally and screw loosening did not occur. Within the limited number of test specimens (7), the difference was statistically significant (P = .015). FEM analysis displayed higher stress peaks up to 800 MPa at the cervical aspect of the zirconium abutment and at the apical third of its retaining screw at an external load of 250 N.
CONCLUSION: Within the limitations of this study, zirconium implant abutments exceeded the established values for maximum incisal bite forces reported in the literature and tightly fit into the titanium implant after several millions of loading cycles.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16429699

Source DB:  PubMed          Journal:  Quintessence Int        ISSN: 0033-6572            Impact factor:   1.677


  9 in total

Review 1.  Mechanical resistance of zirconium implant abutments: a review of the literature.

Authors:  R Velázquez-Cayón; C Vaquero-Aguilar; D Torres-Lagares; M Jiménez-Melendo; J-L Gutiérrez-Pérez
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2012-03-01

2.  Fracture behavior of straight or angulated zirconia implant abutments supporting anterior single crowns.

Authors:  Frank P Nothdurft; Klaus E Doppler; Kurt J Erdelt; Andreas W Knauber; Peter R Pospiech
Journal:  Clin Oral Investig       Date:  2010-01-19       Impact factor: 3.573

3.  Effect of Surface Modification on Viability of L929 Cells on Zirconia Nanocomposite Substrat.

Authors:  Moluk Aivazi; Mohammadhossein Fathi; Farahnaz Nejatidanesh; Vajihesadat Mortazavi; Batoul Hashemi Beni; Jukka Pekka Matinlinna
Journal:  J Lasers Med Sci       Date:  2017-03-20

4.  Mechanical behavior of single-layer ceramized zirconia abutments for dental implant prosthetic rehabilitation.

Authors:  Manuel Jiménez-Melendo; Oriol Llena-Blasco; August Bruguera; Jaime Llena-Blasco; Rosa-María Yáñez-Vico; Manuel García-Calderón; Cristina Vaquero-Aguilar; Rocío Velázquez-Cayón; José-Luis Gutiérrez-Pérez; Daniel Torres-Lagares
Journal:  J Clin Exp Dent       Date:  2014-12-01

5.  Fracture behaviour of implant-implant- and implant-tooth-supported all-ceramic fixed dental prostheses utilising zirconium dioxide implant abutments.

Authors:  Frank Philipp Nothdurft; Sabine Merker; Peter Reinhard Pospiech
Journal:  Clin Oral Investig       Date:  2010-01-05       Impact factor: 3.573

6.  Influence of albumin on the electrochemical behaviour of Zr in phosphate buffered saline solutions.

Authors:  Lu-Ning Wang; Xian-Qiu Huang; Alyssa Shinbine; Jing-Li Luo
Journal:  J Mater Sci Mater Med       Date:  2012-11-21       Impact factor: 3.896

7.  Resistance of three implant-abutment interfaces to fatigue testing.

Authors:  Cleide Gisele Ribeiro; Maria Luiza Cabral Maia; Susanne S Scherrer; Antonio Carlos Cardoso; H W Anselm Wiskott
Journal:  J Appl Oral Sci       Date:  2011-07-01       Impact factor: 2.698

8.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

9.  Comparing the Fracture Resistance and Modes of Failure in Different Types of CAD/CAM Zirconia Abutments with Internal Hexagonal Implants: An In Vitro Study.

Authors:  Yu-Tsen Chang; Yu-Ling Wu; Hung-Shyong Chen; Ming-Hsu Tsai; Chia-Chen Chang; Aaron Yu-Jen Wu
Journal:  Materials (Basel)       Date:  2022-04-04       Impact factor: 3.623

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.