UNLABELLED: Although the use of zirconia abutments for implant-supported restorations has gained momentum with the increasing demand for esthetics, little informed design rationale has been developed to characterize their fatigue behavior under different clinical scenarios. However, to prevent the zirconia from fracturing, the use of a titanium connection in bi-component aesthetic abutments has been suggested. OBJECTIVE: Mechanical testing of customized thin-walled titanium-zirconia abutments at the connection with the implant was performed in order to characterize the fatigue behavior and the failure modes for straight and angled abutments. MATERIAL AND METHODS: Twenty custom-made bi-component abutments were tested according to ISO 14801:2007 either at a straight or a 25° angle inclination (n=10 each group). Fatigue was conducted at 15 Hz for 5 million cycles in dry conditions at 20°C±5°C. Mean values and standard deviations were calculated for each group. All comparisons were performed by t-tests assuming unequal variances. The level of statistical significance was set at p≤0.05. Failed samples were inspected in a polarized-light and then in a scanning electron microscope. RESULTS: Straight and angled abutments mean maximum load was 296.7 N and 1,145 N, the dynamic loading mean Fmax was 237.4 N and 240.7 N, respectively. No significant differences resulted between the straight and angled bi-component abutments in both static (p=0.253) and dynamic testing (p=0.135). A significant difference in the bending moment required for fracture was detected between the groups (p=0.01). Fractures in the angled group occurred mainly at the point of load application, whereas in the straight abutments, fractures were located coronally and close to the thinly designed areas at the cervical region. CONCLUSION: Angled or straight thin-walled zirconia abutments presented similar Fmax under fatigue testing despite the different bending moments required for fracture. The main implication is that although zirconia angled or straight abutments presented similar mechanical behavior, the failure mode tended to be more catastrophic in straight (fracture at the cervical region) compared to angled abutments.
UNLABELLED: Although the use of zirconia abutments for implant-supported restorations has gained momentum with the increasing demand for esthetics, little informed design rationale has been developed to characterize their fatigue behavior under different clinical scenarios. However, to prevent the zirconia from fracturing, the use of a titanium connection in bi-component aesthetic abutments has been suggested. OBJECTIVE: Mechanical testing of customized thin-walled titanium-zirconia abutments at the connection with the implant was performed in order to characterize the fatigue behavior and the failure modes for straight and angled abutments. MATERIAL AND METHODS: Twenty custom-made bi-component abutments were tested according to ISO 14801:2007 either at a straight or a 25° angle inclination (n=10 each group). Fatigue was conducted at 15 Hz for 5 million cycles in dry conditions at 20°C±5°C. Mean values and standard deviations were calculated for each group. All comparisons were performed by t-tests assuming unequal variances. The level of statistical significance was set at p≤0.05. Failed samples were inspected in a polarized-light and then in a scanning electron microscope. RESULTS: Straight and angled abutments mean maximum load was 296.7 N and 1,145 N, the dynamic loading mean Fmax was 237.4 N and 240.7 N, respectively. No significant differences resulted between the straight and angled bi-component abutments in both static (p=0.253) and dynamic testing (p=0.135). A significant difference in the bending moment required for fracture was detected between the groups (p=0.01). Fractures in the angled group occurred mainly at the point of load application, whereas in the straight abutments, fractures were located coronally and close to the thinly designed areas at the cervical region. CONCLUSION: Angled or straight thin-walled zirconia abutments presented similar Fmax under fatigue testing despite the different bending moments required for fracture. The main implication is that although zirconia angled or straight abutments presented similar mechanical behavior, the failure mode tended to be more catastrophic in straight (fracture at the cervical region) compared to angled abutments.
Since osseointegration became a safe treatment modality in dentistry, several designs of
implant-abutment systems have been available for clinical use, with a plethora of data
being produced on titanium as the main abutment material. Historically, the external
hexagon connection was designed to provide an engagement method for implant placement
and anti-rotational feature for single-unit prosthesis, and is likely the functioning
system with the longest clinical follow-up[15,20,21]. Despite the high survival rates of titanium abutments,
the esthetic outcome should also be considered as a success factor, especially in the
anterior esthetic zone. Key elements to be considered include: height of the smile line,
gingival biotype, color of the neighboring teeth and the esthetic expectations of the
patient[19].In essence, implant abutments can be made of ceramic (yttrium tetragona zirconia
polycristals, i.e. Y-TZP) or titanium (commercially pure or alloyed) and can also be
prefabricated or customized. The standardization of the prefabricated abutments may
present a limitation to the establishment of an appropriate emergence profile,
especially in cases where a discrepancy exists between the implant and the crown
diameter. In such cases, the compensation will be made by the final contour of the
crown, which could result in unfavorable core/porcelain thickness ratios. A way to
overcome this issue is the use of a customized abutment that will allow for an
individual emergence profile forthrightly by the abutment instead of the crown[13].Customized-zirconia abutments have shown comparable survival rates relative to titanium
abutments[17,22].However, the internal connection between a zirconia customized abutment and the implant
continues to be a mechanical challenge. Thus, to prevent the zirconia from fracturing at
the connection, the use of a titanium connection in bi-component aesthetic abutments
have been suggested[2].Among the concerns regarding the use of CAD/CAM customized-zirconia abutments is their
unknown mechanical performance in clinical situations where the implant positioning and
occlusal restraints result in thin abutment walls that may negatively influence the
long-term prosthetic functional outcome[3]. Although the buccolingual positioning and angulation of the implant
is crucial for the final prosthetic outcome, oftentimes there is a need to compensate
non-ideal implant positioning with angled customized abutments. Whereas, the mechanical
property of customized titanium abutments may be less affected by the resulting wall
thinning, due to its ductile nature, the same assumption may not hold true for brittle
ceramic abutments. Therefore, this study evaluated the mechanical performance of thinned
zirconia abutments at the connection, in an attempt to simulate 2 implantation scenarios
in terms of inclination. In fact, in the anterior aesthetic zone, according to the
different available bone conditions, implant insertion may require straight (mostly in
canine and bicuspid zones) or inclined (incisor zones) abutments.The tested null hypothesis was that a straight bi-component abutment does not present a
different fatigue life compared to an angled one.
MATERIAL AND METHODS
Twenty custom-made zirconia abutments (Sweden & Martina, Padua, Italy) were tested
according to Norm UNI eN ISO 14801:2007 (Dynamic fatigue test of endosseous dental
implants) either with a straight inclination (n=10) or a 25º angle relative to the
implant long axis (n=10) (Figures 1A and B). In
order to evaluate the abutments under a worst case scenario, an industrially imposed
minimum thickness of 0.3 mm at the connection area with the implant was chosen.
Figure 1
Testing configuration for the A) straight zirconia (Zr) abutment showing the
distances between the abutment assembly and the potting surface and B) the 30º
loading orientation. The area circled in B is magnified in C) where the component
parts are described and the circumferential 0.3 mm thicknesses of the zirconia
abutment at the connection is depicted. The same patterns regarding the D)
distance of the abutment assembly and E) loading orientation are observed in the
25º angled abutment groups. Components and thicknesses circled in E are described
in magnified figure F
Testing configuration for the A) straight zirconia (Zr) abutment showing the
distances between the abutment assembly and the potting surface and B) the 30º
loading orientation. The area circled in B is magnified in C) where the component
parts are described and the circumferential 0.3 mm thicknesses of the zirconia
abutment at the connection is depicted. The same patterns regarding the D)
distance of the abutment assembly and E) loading orientation are observed in the
25º angled abutment groups. Components and thicknesses circled in E are described
in magnified figure FThe implant fixture was inserted in a prefabricated rigid Plexiglas® clamping
device. Individualized zirconia abutments were cemented on a titanium connection and
fixed to the implant using a titanium screw according to the manufacturer's
instructions: after cleaning with alcohol, the zirconia abutment was cemented onto the
titanium connection (All Stone cement, Sweden & Martina, Padua, Italy). The mixed
cement was applied onto the contact part of the titanium connection. The zirconia
abutment was then fit onto the titanium connection. The remaining cement was removed
immediately with steam and after polymerization, the excess cement was removed.A PEEK (polyetherketone) ring was used for stress distribution at the zirconia/titanium
interface interposed between the zirconia abutment and retaining screw (Figure 2).
Figure 2
Components of the custom-made abutment composed by an individualized zirconia
abutment cemented on a titanium connection linked to the implant using a titanium
screw. A peek ring was used for stress distribution at the zirconia/titanium
interface interposed between the zirconia abutment and retaining screw
Components of the custom-made abutment composed by an individualized zirconia
abutment cemented on a titanium connection linked to the implant using a titanium
screw. A peek ring was used for stress distribution at the zirconia/titanium
interface interposed between the zirconia abutment and retaining screwFor mechanical testing, single load to fracture (SLF) was performed (6027, Instron, USA)
(n=10 each), and the mean values were used to calculate the maximum load
(Fmax). The loading geometry was fixed by UNI EN ISO 14801:2007 both for
the SLF and dynamic loading tests. The implant/abutment system was positioned with its
long axis inclined at a 30º±1º angle relative to the loading direction. The dimensions
of the indenter resulted in a distance of l=11.0 mmº±0.1 mm from the
center of the hemisphere to the clamping plane (Figure
3a).
Figure 3
(A) Single load to fracture (SLF) and (B) dynamic loading testing machines
(A) Single load to fracture (SLF) and (B) dynamic loading testing machinesA hemispherical indenter was used for the load application. Eighty percent of the
Fmax was used as a starting load value for the dynamic testing, which used
an uni-axial testing equipment (FPF, Italsigma, Italy), with a 2.000 N load cell (TSTM,
AeP, Italy) and an extensometer (2620-601, Instrom, USA) (Figure 3b). Fatigue was conducted at 15 Hz for 5 million cycles in dry
conditions at 20ºC±5ºC. The bending moment was calculated as follows:where y is the moment arm (defined by l x sin 30º), and F is the force expressed in
newtons.Failed samples were inspected in a polarized-light microscope (MZ-APO, Carl Zeiss Micro
Imaging, Thornwood, NY, USA) and then in a scanning electron microscope at an
acceleration voltage of 10 Kv (XVP®, evo 50, Carls Zeiss AG, Oberkochen,
Germany).
Statistical analysis
Mean values and standard deviations were calculated for each outcome variable (static
and dynamic tests) as a function of the straight and inclined abutment groups.All comparisons were performed by t-test assuming unequal variances [SigmaStat 3.0
(SPSS Inc., Chicago, IL, USA)]. The level of statistical significance was set at
p≤0.05.
RESULTS
The mean Fmax value for SLF, dynamic loading are presented in Table 1, as well as the bending moments required
for fracture.
Table 1
Mean values for mechanical testing of the straight and 25 angled zirconia abutment
groups
STATIC Mean (N) (SD)
DYNAMIC Fmax (N)
BENDING MOMENT Mean (Nmm) (SD)
STRAIGHT
296.7 (34.42)
237.4
690.6 (60)
25º ANGLED
1145 (30.69)
240.7
410.5 (5.45)
SD: Standard Deviation
Mean values for mechanical testing of the straight and 25 angled zirconia abutment
groupsSD: Standard DeviationNo significant differences in load to failure were observed between the straight and
angled bi-component abutments in both the static (p=0.253) and dynamic tests (p=0.135).
A significant difference in the bending moment required for fracture was detected
between the groups (p=0.01).Fractures in the angled group occurred chiefly at the point of load application, whereas
in the straight abutments fractures were located coronally and close to the thinly
designed areas at the cervical region (Figure
4).
Figure 4
Representative polarized light and scanning electron microscope micrographs of
abutment samples failed during fatigue. A) Straight abutment group in a frontal
view depicting that fractures occurred coronally at B) thinly designed areas
(pointer) mostly close to the base of the abutment, as observed in this occlusal
view. C) In the angled group, fracture occurred at the point of load application
involving D) thinly designed areas (pointer), as observed in this occlusal
view
Representative polarized light and scanning electron microscope micrographs of
abutment samples failed during fatigue. A) Straight abutment group in a frontal
view depicting that fractures occurred coronally at B) thinly designed areas
(pointer) mostly close to the base of the abutment, as observed in this occlusal
view. C) In the angled group, fracture occurred at the point of load application
involving D) thinly designed areas (pointer), as observed in this occlusal
view
DISCUSSION
The present study simulated clinical scenarios where customized zirconia abutments
presented the thinnest possible section at the connection area, resulting in significant
implications in its mechanical response, according to our SLF testing. However, when
evaluated under a more clinically realistic loading scenario, (fatigue) Fmax
were not significantly different between the groups, but bending moments for fracture
were higher for the straight relative to the angled abutments. Such a result was not
expected considering the differences in the lever scenario created for both groups, but
may be accounted by the difference in the moment of inertia between the two designs.The resulting failure modes where the straight abutments showed fractures at the
cervical region compared to fractures at the point of load application in the angled
abutments likely reflected the resulting bending moments for both groups. From a
clinical standpoint, although both scenarios may fail at similar time points, it can be
speculated that in a restoration supported by a thinned angled zirconia abutment, the
crown is more likely to shield the functional area receiving the load compared to a
thinned straight abutment. Such an assumption warrants further investigation. The
bending moments of both groups were significantly higher compared to values observed in
the literature[18]. This could be
related to the double zirconia/titanium anchorage system.As to the testing methods available for the evaluation of different implant-abutment
system configurations, several have been described, such as the single load to
fracture[1], the use of fatigue
followed by the application of a static load until fracture[5,11], the staircase
method[16], fatigue limit (ISO
14801:2007), step-stress accelerated life testing[6], and others. While the ISO 14801 was created in 2003 and revised
in 2007, with the aim of standardizing the testing procedures and data presentation in
the fatigue of dental implants, it has been shown that the results produced by such a
method should be interpreted with caution. The wide range of testing parameters allowed
in the ISO 14801:2007 regarding the testing frequency (2 to 15 Hz), environment (water
or dry when testing above 15 Hz) and amount of cycles (2 or 5 million, depending on the
chosen frequency) have shown that a very different failure probability distribution may
result[10], as well as failure
modes[12]. Whereas, the present
study utilized the highest end of the speed scale allowed by the norm, especially to
optimize the machine testing time, potential differences in the fatigue limit and
failure modes when testing in 2 Hz in water may exist and warrant future investigation.
It may be suggested that, while attending industry requirements for implants quality
assurance and control, the ISO 14801:2007 testing methodology may likely continue to be
developed as observed from 2003 to the currently advocated version[10,12].As generally reported and also in agreement with our testing results, the discrepancy in
values arising from the static relative to the dynamic testing and the clinical
relevance of each, should be acknowledged. If the selection of a system was based on the
static test results, which do not simulate physiological loading scenarios where
repetitive, lower load cycling is the chief mechanism leading to failure, most systems
would be suitable irrespective of the area of application (molar or incisor). With
static loading, materials will commonly fail at loads higher than those in fatigue. The
mechanisms of subcritical crack growth that should be simulated in testing include those
that operate in use-related failures which are stress corrosion and cyclic
fatigue[14]. In this study, static
testing was only performed to provide load values for subsequent fatigue testing.
Fatigue can provide more effective ways of simulating failures observed
clinically[4].Considering that the reported maximal bite force in the incisor area may vary from 108
N[8] to 190 N[7,9],
the Fmax fatigue loads observed for both thin-walled straight or angled
zirconia abutments are above the physiologic range. Therefore, in esthetically demanding
areas, zirconia abutments may be indicated especially when thin peri-implant soft
tissues are present[19]. Although no
direct comparison was performed with titanium abutments, short-term clinical studies and
systematic reviews point toward similar success rates for both materials and no
influence in the failure behavior of the overlying restorative system[9]. Future long-term clinical studies are
warranted.
CONCLUSION
The postulated null hypothesis that a straight bi-component abutment would not present
different Fmax than an angled abutment under fatigue was accepted. However,
failure modes differed between the groups.