Evaluation of the ITI Morse taper implant/abutment design with an internal modification.

Thomas A Ding1, Ronald D Woody, Frank L Higginbottom, Barbara H Miller.   

Abstract

PURPOSE: The purpose of this study was to compare internal Morse taper connections in 2 separate modes: repeated torque/reverse-torque values and compressive bending at a 30-degree off-axis angle.
MATERIALS AND METHODS: Three sample groups (n = 12 in each group)--a solid-screw implant paired with a 5.5-mm solid abutment (SSI), a synOcta implant with a 5.5-mm solid abutment (SOI), and a synOcta implant with a synOcta 5.5-mm solid abutment (SOSA)--were torqued to 35 Ncm, and the reverse torque to remove the abutment was recorded. This was repeated for 3 trials. Additionally, the sample groups were loaded 30 degrees off-axis, and the ultimate compressive values were recorded.
RESULTS: There was a significant difference in the initial reverse-torque values. The SOSA setup showed significantly lower torque than the SOI and SSI setups (P < .05). In addition, the compressive bending test showed that the SOSA setup was significantly different (P < .05) from the SSI and SOI setups. Radiographic survey of the test groups following compressive bending revealed no implant fractures, but bending of the implant-abutment complex occurred. DISCUSSION: The alteration within the Morse taper did not reduce the strength of the implant-abutment connection, ie, the reduction in surface area did not significantly reduce the torque properties or tensile properties. The new 2-piece synOcta 5.5-mm solid abutment was shown to have a stronger implant-abutment connection when torqued down a second time.
CONCLUSIONS: In this in vitro study, alteration of the Morse taper with an internal octagon indexing did not significantly reduce the strength of the implant connection. Sufficient strength was exhibited, which would indicate this implant-abutment design for anterior as well as posterior edentulous sites.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14696662

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  7 in total

Review 1.  Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review.

Authors:  Marcelo Coelho Goiato; Eduardo Piza Pellizzer; Emily Vivianne Freitas da Silva; Liliane da Rocha Bonatto; Daniela Micheline dos Santos
Journal:  Oral Maxillofac Surg       Date:  2015-04-25

2.  Resistance of three implant-abutment interfaces to fatigue testing.

Authors:  Cleide Gisele Ribeiro; Maria Luiza Cabral Maia; Susanne S Scherrer; Antonio Carlos Cardoso; H W Anselm Wiskott
Journal:  J Appl Oral Sci       Date:  2011-07-01       Impact factor: 2.698

Review 3.  Morse taper dental implants and platform switching: The new paradigm in oral implantology.

Authors:  José Paulo Macedo; Jorge Pereira; Brendan R Vahey; Bruno Henriques; Cesar A M Benfatti; Ricardo S Magini; José López-López; Júlio C M Souza
Journal:  Eur J Dent       Date:  2016 Jan-Mar

4.  Influence of abutment design on retention of metal copings cemented to implants.

Authors:  Albano Porto da Cunha; Glauco Pereira Moysés; Ana Christina Claro Neves; Rafael Pino Vitti; Flávia Cardoso da Rosa Goulart; Laís Regiane da Siva-Concílio
Journal:  Acta Biomater Odontol Scand       Date:  2016-01-26

5.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

6.  Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections.

Authors:  Saied Nokar; Habib Hajimiragha; Leyla Sadighpour; Azam Sadat Mostafavi
Journal:  Dent Res J (Isfahan)       Date:  2020-12-10

7.  Effect of the prosthetic index on stress distribution in Morse taper connection implant system and peri-implant bone: a 3D finite element analysis.

Authors:  Wen-Tao Zhang; Kang-Jie Cheng; Fan Yang; Lin-Hong Wang; Yun-Feng Liu; Russell Wang; Yun-Fang Chen; Yu-de Ding
Journal:  BMC Oral Health       Date:  2022-09-30       Impact factor: 3.747

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.