Implant design and interface force transfer. A photoelastic and strain-gauge analysis.

Murat Cehreli1, Joke Duyck, Michel De Cooman, Robert Puers, Ignace Naert.   

Abstract

OBJECTIVES: To compare stress and strain magnitudes of butt-joint and internal-cone oral implants in a bone simulant.
MATERIAL AND METHODS: Photoelastic and strain-gauged models of solitary Brånemark, ITI and Astra Tech implants were obtained. Vertical and 20 degrees oblique forces of 100 and 150 N were applied on the abutments in separate load cases. Isochromatic fringe patterns around implants were observed and photographed in the field of a polariscope. Strain-gauge measurements were performed and principal strains induced around implants were calculated.
RESULTS: Isochromatic fringe orders of all designs were similar under both loading conditions (P>0.05). Strains around Brånemark implants were lower than around Astra Tech and ITI implants, particularly under vertical loads. The latter implants had similar strain magnitudes (P>0.05), although strains around the conical Astra Tech implant were lower under vertical load.
CONCLUSIONS: Butt-joint and internal-cone oral implants have similar force distribution characteristics. The implant-abutment mating design is not a decisive factor affecting stress and strain magnitudes in a bone simulant.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15008938     DOI: 10.1111/j.1600-0501.2004.00979.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  18 in total

1.  Bone strains around apically free versus grafted implants in the posterior maxilla of human cadavers.

Authors:  Murat Cavit Cehreli; Murat Akkocaoglu; Ayhan Comert; Ibrahim Tekdemir; Kivanc Akca
Journal:  Med Biol Eng Comput       Date:  2007-03-06       Impact factor: 2.602

2.  Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional finite element method.

Authors:  João Rodrigo Sarot; Cintia Mussi Milani Contar; Ariadne Cristiane Cabral da Cruz; Ricardo de Souza Magini
Journal:  J Mater Sci Mater Med       Date:  2010-05-13       Impact factor: 3.896

3.  Stress distribution of multiple implant-supported prostheses: Photoelastic and strain gauge analyses of external hexagon and morse taper connections.

Authors:  Ana-Beatriz-Bueno-Carlini Bittencourt; Erica-de Oliveira-Paiva Rezende; Marcio Campaner; Sandro-Basso Bitencourt; Daniela-Micheline Dos Santos; Aldiéris-Alves Pesqueira; Marcelo-Coelho Goiato
Journal:  J Clin Exp Dent       Date:  2022-03-01

4.  Influence of Occlusal Bite Forces on Teeth with Altered Periodontal Support: A Three-Dimensional Finite Element Stress Analysis.

Authors:  Richa Agrawal; Sumit Narang; Hina Ahmed; Shyam Prasad; Shyamsunder Reddy; Shivaramakrishna Aila
Journal:  J Pharm Bioallied Sci       Date:  2021-06-05

5.  Photoelastic stress analysis surrounding implant-supported prosthesis and alveolar ridge on mandibular overdentures.

Authors:  Dorival Pedroso da Silva; Claudia Cazal; Fernanda Campos Sousa de Almeida; Reinaldo Brito E Dias; Rafael Yagüe Ballester
Journal:  Int J Dent       Date:  2010-05-10

6.  Stress on external hexagon and Morse taper implants submitted to immediate loading.

Authors:  Caroline H Odo; Marcele J Pimentel; Rafael L X Consani; Marcelo F Mesquita; Mauro A A Nóbilo
Journal:  J Oral Biol Craniofac Res       Date:  2015-07-29

7.  Photoelastic and finite element analyses of occlusal loads in mandibular body.

Authors:  Ana Cláudia Rossi; Alexandre Rodrigues Freire; Felippe Bevilacqua Prado; Luciana Asprino; Lourenço Correr-Sobrinho; Paulo Henrique Ferreira Caria
Journal:  Anat Res Int       Date:  2014-10-08

8.  Resistance of three implant-abutment interfaces to fatigue testing.

Authors:  Cleide Gisele Ribeiro; Maria Luiza Cabral Maia; Susanne S Scherrer; Antonio Carlos Cardoso; H W Anselm Wiskott
Journal:  J Appl Oral Sci       Date:  2011-07-01       Impact factor: 2.698

9.  Photoelastic analysis of fixed partial prosthesis crown height and implant length on distribution of stress in two dental implant systems.

Authors:  Evandro Portela Figueirêdo; Eder Alberto Sigua-Rodriguez; Marcele Jardim Pimentel; Ana Regina Oliveira Moreira; Mauro Antônio de Arruda Nóbilo; José Ricardo de Albergaria-Barbosa
Journal:  Int J Dent       Date:  2014-10-08

10.  A three-dimensional finite element study on the stress distribution pattern of two prosthetic abutments for external hexagon implants.

Authors:  Wagner Moreira; Caio Hermann; Jucélio Tomás Pereira; Jean Anacleto Balbinoti; Rodrigo Tiossi
Journal:  Eur J Dent       Date:  2013-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.