Bao Shu1, Fangjun Bao1,2, Giacomo Savini3, Weicong Lu1,2, Ruixue Tu1,2, Haisi Chen1, Benhao Song1, Qinmei Wang1,2, Jinhai Huang1,2.
Abstract
BACKGROUND: To evaluate the effect of orthokeratology on precision of measurements in children using a new swept-source optical coherence tomography (SS-OCT) optical biometer (OA-2000), and agreement between its measurements and those provided by the commonly used IOLMaster based on partial coherence interferometry (PCI).Entities:
Keywords: Agreement; Children; Orthokeratology; Precision; Swept source optical coherence tomography
Year: 2020 PMID: 32161772 PMCID: PMC7053116 DOI: 10.1186/s40662-020-00177-4
Source DB: PubMed Journal: Eye Vis (Lond) ISSN: 2326-0254
Intra-observer repeatability of ocular biological parameters measured using the new swept-source optical coherence tomography biometer in children before wearing orthokeratology lens
| Parameter | Observer | Mean ± SD | Sw | R | CoV (%) | ICC |
|---|---|---|---|---|---|---|
| AL (mm) | 1st | 25.15 ± 0.99 | 0.02 | 0.06 | 0.08 | 1.000 |
| 2nd | 25.15 ± 0.99 | 0.01 | 0.03 | 0.04 | 1.000 | |
| CCT (μm) | 1st | 537.59 ± 31.76 | 4.38 | 12.13 | 0.81 | 0.981 |
| 2nd | 537.46 ± 31.49 | 4.01 | 11.10 | 0.75 | 0.984 | |
| ACD (mm) | 1st | 3.72 ± 0.19 | 0.02 | 0.06 | 0.61 | 0.986 |
| 2nd | 3.73 ± 0.19 | 0.02 | 0.06 | 0.63 | 0.984 | |
| LT (mm) | 1st | 3.43 ± 0.14 | 0.03 | 0.09 | 0.91 | 0.953 |
| 2nd | 3.43 ± 0.14 | 0.03 | 0.08 | 0.83 | 0.960 | |
| Km (Φ = 2.5) (D) | 1st | 43.15 ± 1.45 | 0.10 | 0.28 | 0.23 | 0.995 |
| 2nd | 43.18 ± 1.44 | 0.08 | 0.22 | 0.18 | 0.997 | |
| Km (Φ = 3.0) (D) | 1st | 43.13 ± 1.46 | 0.09 | 0.26 | 0.21 | 0.996 |
| 2nd | 43.15 ± 1.46 | 0.07 | 0.19 | 0.16 | 0.998 | |
| CD (mm) | 1st | 11.96 ± 0.43 | 0.06 | 0.17 | 0.51 | 0.980 |
| 2nd | 11.96 ± 0.41 | 0.06 | 0.17 | 0.51 | 0.978 |
AL= Axial length; CCT= central corneal thickness; ACD= anterior chamber depth (corneal epithelium to lens); LT= lens thickness; Km= mean keratometry; CD= corneal diameter; SD= standard deviation; Sw= within-subject standard deviation; R= reproducibility limit (2.77 Sw); CoV= within-subject coefficient of variation; ICC= intraclass correlation coefficient
Intra-observer repeatability of ocular biological parameters measured using the new swept-source optical coherence tomography biometer in children after wearing orthokeratology lens
| Parameter | Observer | Mean ± SD | Sw | R | CoV (%) | ICC |
|---|---|---|---|---|---|---|
| AL (mm) | 1st | 25.15 ± 0.99 | 0.01 | 0.03 | 0.04 | 1.000 |
| 2nd | 25.14 ± 0.99 | 0.01 | 0.04 | 0.06 | 1.000 | |
| CCT (μm) | 1st | 529.21 ± 31.95 | 4.01 | 11.10 | 0.76 | 0.983 |
| 2nd | 529.22 ± 32.64 | 4.27 | 11.83 | 0.81 | 0.983 | |
| ACD (mm) | 1st | 3.65 ± 0.22 | 0.02 | 0.06 | 0.64 | 0.983 |
| 2nd | 3.65 ± 0.23 | 0.05 | 0.13 | 1.23 | 0.962 | |
| LT (mm) | 1st | 3.49 ± 0.20 | 0.03 | 0.08 | 0.82 | 0.967 |
| 2nd | 3.48 ± 0.20 | 0.04 | 0.11 | 1.18 | 0.96 | |
| Km (Φ = 2.5) (D) | 1st | 40.81 ± 1.55 | 0.08 | 0.22 | 0.19 | 0.993 |
| 2nd | 40.80 ± 1.57 | 0.13 | 0.36 | 0.32 | 0.993 | |
| Km (Φ = 3.0) (D) | 1st | 41.17 ± 1.56 | 0.07 | 0.19 | 0.16 | 0.994 |
| 2nd | 41.16 ± 1.58 | 0.11 | 0.30 | 0.26 | 0.995 | |
| CD (mm) | 1st | 11.94 ± 0.39 | 0.06 | 0.17 | 0.51 | 0.974 |
| 2nd | 11.95 ± 0.38 | 0.08 | 0.22 | 0.66 | 0.958 |
AL= Axial length; CCT= central corneal thickness; ACD= anterior chamber depth (corneal epithelium to lens); LT= lens thickness; Km= mean keratometry; CD= corneal diameter; SD= standard deviation; Sw= within-subject standard deviation; R= reproducibility limit (2.77 Sw); CoV= within-subject coefficient of variation; ICC= intraclass correlation coefficient
Inter-observer reproducibility of ocular biological parameters measured using the new swept-source optical coherence tomography biometer in children before and after wearing orthokeratology lens
| Parameter | Status | Sw | R | CoV (%) | ICC |
|---|---|---|---|---|---|
| AL (mm) | Before | 0.01 | 0.02 | 0.03 | 1.000 |
| After | 0.01 | 0.02 | 0.03 | 1.000 | |
| CCT (μm) | Before | 2.36 | 6.53 | 0.44 | 0.994 |
| After | 2.53 | 7.02 | 0.48 | 0.994 | |
| ACD (mm) | Before | 0.02 | 0.06 | 0.54 | 0.988 |
| After | 0.02 | 0.06 | 0.64 | 0.989 | |
| LT (mm) | Before | 0.03 | 0.07 | 0.75 | 0.967 |
| After | 0.02 | 0.07 | 0.71 | 0.985 | |
| Km (Φ = 2.5) (D) | Before | 0.06 | 0.17 | 0.14 | 0.998 |
| After | 0.08 | 0.21 | 0.19 | 0.998 | |
| Km (Φ = 3.0) (D) | Before | 0.06 | 0.17 | 0.14 | 0.998 |
| After | 0.06 | 0.18 | 0.15 | 0.998 | |
| CD (mm) | Before | 0.06 | 0.17 | 0.50 | 0.980 |
| After | 0.06 | 0.17 | 0.50 | 0.976 |
AL= Axial length; CCT= central corneal thickness; ACD= anterior chamber depth; LT= lens thickness; Km= mean keratometry; CD= corneal diameter; SD= standard deviation; Sw= within-subject standard deviation; R= reproducibility limit (2.77 Sw); CoV= within-subject coefficient of variation; ICC= intraclass correlation coefficient
Comparison of ocular biological parameters measured using the new swept-source optical coherence tomography biometer and the partial coherence interferometry biometer in children before and after wearing orthokeratology lens
| Device Pairings | Status | Mean ± SD | P Value | 95% LoA |
|---|---|---|---|---|
| AL (mm) | Before | 0.01 ± 0.03 | 0.004 | −0.04 to 0.06 |
| After | 0.01 ± 0.02 | 0.000 | −0.02 to 0.05 | |
| ACD (mm) | Before | 0.08 ± 0.08 | 0.000 | − 0.07 to 0.24 |
| After | 0.06 ± 0.10 | 0.000 | −0.13 to 0.25 | |
| Km (Φ = 2.5) (D) | Before | −0.04 ± 0.14 | 0.070 | − 0.32 to 0.24 |
| After | 0.05 ± 0.24 | 0.144 | −0.42 to 0.52 | |
| Km (Φ = 3.0) (D) | Before | −0.06 ± 0.15 | 0.013 | − 0.36 to 0.25 |
| After | 0.41 ± 0.33 | 0.000 | −0.24 to 1.05 | |
| CD (mm) | Before | −0.15 ± 0.39 | 0.008 | − 0.92 to 0.61 |
| After | −0.25 ± 0.34 | 0.000 | − 0.92 to 0.42 |
AL= Axial length; ACD= anterior chamber depth; Km= mean keratometry; CD corneal diameter; SD= Standard deviation; LoA= limit of agreement
Fig. 1Bland-Altman plots show agreement between OA-2000 and IOLMaster for axial length (a), chamber depth (b), mean keratometry at 2.5 mm diameter (c), mean keratometry at 3.0 mm diameter (d), corneal diameter (e) measurement pre-orthokeratology. The solid line indicates the mean difference (bias), and the dotted lines represent the 95% limits of agreement
Fig. 2Bland-Altman plots show agreement between OA-2000 and IOLMaster for axial length (a), chamber depth (b), mean keratometry at 2.5 mm diameter (c), mean keratometry at 3.0 mm diameter (d), corneal diameter (e) measurement post-orthokeratology. The solid line indicates the mean difference (bias), and the dotted lines represent the 95% limits of agreement