Abstract
BACKGROUND: Cataract surgery is the most common surgical procedure in ophthalmology. Biometry data and accurate intraocular lens (IOL) calculations are very important in achieving the desired refractive outcomes. The aim of this study was to compare measurements using a new optical low coherence reflectometry (OLCR) biometer (OA-2000) and the gold standard partial coherence interferometry (PCI) optical biometer (IOLMaster 500).Entities:
Keywords: Cataract surgery; Intraocular lens power calculation; Optical biometer; Optical low coherence reflectometer; Partial coherence interferometry
Year: 2016 PMID: 27833928 PMCID: PMC5066293 DOI: 10.1186/s40662-016-0059-1
Source DB: PubMed Journal: Eye Vis (Lond) ISSN: 2326-0254
The differences in parameters between the new OCLR and the standard PCI biometers
| Parameter | New biometer | PCI biometer | Difference |
|
| ICC |
|---|---|---|---|---|---|---|
| Axial length (mm) | ||||||
| Mean ± (SD) | 23.12 ± 1.34 | 23.18 ± 1.08 | -0.05 ± 0.05 | <0.001 | 0.999 | 0.996 |
| Range | 20.79, 26.5 | 20.93, 26.60 | -0.16, 0.05a | |||
| Keratometry (D) | ||||||
| Mean ± (SD) | 44.50 ± 1.8 | 44.60 ± 1.81 | -0.11 ± 0.37 | <0.001 | 0.980 | 0.980 |
| Range | 40.44, 47.38 | 40.57, 47.37 | -0.84, 0.63 | |||
| ACD (mm) | ||||||
| Mean ± (SD) | 3.06 ± 0.39 | 3.15 ± 0.39 | -0.09 ± 0.24 | <0.001 | 0.824 | 0.824 |
| Range | 2.27, 3.10 | 2.29, 3.16 | -0.58, 0.39 | |||
| WTW (mm) | ||||||
| Mean ± (SD) | 11.64 ± 0.77 | 11.86 ± 0.48 | -0.21 ± 0.82 | 0.018 | 0.259 | 0.231 |
| Range | 8.74, 13.83 | 10.84, 12.80 | -1.85, 1.42 | |||
| IOL power (D) | ||||||
| SRK/T | ||||||
| Mean ± (SD) | 20.79 ± 2.81 | 20.44 ± 2.63 | 0.32 ± 0.60 | <0.001 | 0.989 | 0.988 |
| Range | 12.5, 27.5 | 12.0, 27.0 | -1.53, 0.88 | |||
OCLR= optical low coherence reflectometry, PCI= partial coherence interferometry, SD= standard deviation, D= diopters, ACD= anterior chamber depth, WTW= white-to-white corneal diameter, IOL= intraocular lens, r value= Pearson’s correlation coefficient, ICC= intraclass correlation coefficient
a95 % limits of agreement
Fig. 1A Bland-Altman plot comparing parameter measurements between the new optical low coherence reflectometry biometer and the partial coherence interferometry biometer. a axial length; b keratometry; c anterior chamber depth; and d white-to-white corneal diameter
Fig. 2The distribution of the intraocular lens (IOL) power differences: new optical low coherence reflectometry biometer - the partial coherence interferometry biometer = the IOL power difference