Comparison of dual rotating Scheimpflug-Placido, swept-source optical coherence tomography, and Placido-scanning-slit systems.

Yong Woo Lee1, Chul Young Choi2, Geun Young Yoon1.   

Abstract

PURPOSE: To compare measurements of corneal indices using dual rotating Scheimpflug-Placido, swept-source optical coherence tomography (OCT), and Placido-scanning-slit systems.
SETTING: Kangbuk Samsung Hospital, Seoul, South Korea.
DESIGN: Prospective evaluation of diagnostic tests.
METHODS: Corneal topography measurements were performed using dual rotating Scheimpflug-Placido (Galilei G2), swept-source OCT (Casia SS-1000), and Placido-scanning-slit (Orbscan IIz) systems. The intraclass correlation coefficients (ICCs) and Bland-Altman plots were used to evaluate the agreement between measurements.
RESULTS: Fifty post-refractive surgery eyes and 50 normal eyes were evaluated. The agreement in anterior keratometry and pachymetry between the 3 devices was high in both groups (ICC > 0.9). In both groups, the ICC values for posterior keratometry and eccentricity were high between the dual rotating Scheimpflug-Placido and swept-source OCT systems (ICC > 0.9) but not between the Placido-scanning-slit system and the other 2 systems. The Placido-scanning-slit system yielded much steeper values for posterior keratometry in both groups (P < .05). The ICC values for posterior corneal elevation were lower than 0.9 between all 3 devices. The dual rotating Scheimpflug-Placido and swept-source OCT systems showed relatively higher ICC values than the Placido-scanning-slit system in both groups. Maximum posterior elevations were highest with the Placido-scanning-slit system followed by the swept-source OCT system and then the dual rotating Scheimpflug-Placido system.
CONCLUSIONS: Anterior keratometry obtained using 3 devices showed high degrees of agreement. Posterior keratometry and eccentricity showed greater agreement between the dual rotating Scheimpflug-Placido and swept-source OCT systems than with the Placido-scanning-slit system. The dual rotating Scheimpflug-Placido and swept-source OCT systems were equivalent in detecting the shape of the cornea and could be considered interchangeable.
Copyright © 2015. Published by Elsevier Inc.

Entities:  

Mesh:

Year:  2015        PMID: 26049835     DOI: 10.1016/j.jcrs.2014.08.040

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  13 in total

1.  Comparison of Scheimpflug and swept-source anterior segment optical coherence tomography in normal and keratoconus eyes.

Authors:  Seyed Mohammad Ghoreishi; Seyed Ali Akbar Mortazavi; Zahra-Alsadat Abtahi; Mohammad-Ali Abtahi; Seyed Ali Sonbolestan; Seyed-Hossein Abtahi; Mohadeseh Mohammadinia; Kimia Nekouie Isfahani
Journal:  Int Ophthalmol       Date:  2016-09-26       Impact factor: 2.031

2.  Correlation between visual function and refractive, topographic, pachymetric and aberrometric data in eyes with keratoconus.

Authors:  Neslihan Bayraktar Bilen; Ibrahim F Hepsen; Carlos G Arce
Journal:  Int J Ophthalmol       Date:  2016-08-18       Impact factor: 1.779

3.  Pediatric Corneal Structural Development During Childhood Characterized by Ultrasound Biomicroscopy.

Authors:  Snehaa Maripudi; Julia Byrd; Azam Qureshi; Gianna Stoleru; Moran Roni Levin; Osamah J Saeedi; Wuqaas Munir; Marlet Bazemore; Bethany Karwoski; Camilo Martinez; Mohamad S Jaafar; William P Madigan; Janet Leath Alexander
Journal:  J Pediatr Ophthalmol Strabismus       Date:  2020-07-01       Impact factor: 1.402

4.  Measurements of Corneal Thickness in Eyes with Pseudoexfoliation Syndrome: Comparative Study of Different Image Processing Protocols.

Authors:  Katarzyna Krysik; Dariusz Dobrowolski; Katarzyna Polanowska; Anita Lyssek-Boron; Edward A Wylegala
Journal:  J Healthc Eng       Date:  2017-09-07       Impact factor: 2.682

5.  The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging.

Authors:  Yune Zhao; Ding Chen; Giacomo Savini; Qing Wang; Hongfang Zhang; Yili Jin; Benhao Song; Rui Ning; Jinhai Huang; Chenyang Mei
Journal:  Eye Vis (Lond)       Date:  2020-06-09

6.  Impact of the anterior-posterior corneal radius ratio on intraocular lens power calculation errors.

Authors:  Asato Hasegawa; Takashi Kojima; Mana Yamamoto; Yukihito Kato; Akeno Tamaoki; Kazuo Ichikawa
Journal:  Clin Ophthalmol       Date:  2018-08-27

Review 7.  Galilei Corneal Tomography for Screening of Refractive Surgery Candidates: A Review of the Literature, Part II.

Authors:  Majid Moshirfar; Mahsaw N Motlagh; Michael S Murri; Hamed Momeni-Moghaddam; Yasmyne C Ronquillo; Phillip C Hoopes
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2019

8.  Comparison of simulated keratometric changes following wavefront-guided and wavefront-optimized myopic laser-assisted in situ keratomileusis.

Authors:  Wen-Shin Lee; Edward E Manche
Journal:  Clin Ophthalmol       Date:  2018-03-29

9.  Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography.

Authors:  Sisi Chen; Rongrong Gao; Colm McAlinden; Junming Ye; Yiran Wang; Min Chen; Jinhai Huang; Yong Sun; A-Yong Yu
Journal:  J Ophthalmol       Date:  2020-09-04       Impact factor: 1.909

10.  Comparison of Color Light-Emitting Diode Corneal Topographer and Dual Rotating Scheimpflug-Placido Topographer.

Authors:  Jae Hyuck Lee; Yong Woo Lee; Jong Soo Lee; Michael C Knorz; Gerd U Auffarth; Chul Young Choi
Journal:  J Ophthalmol       Date:  2018-12-27       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.