Repeatability and agreement of two A-scan ultrasonic biometers and IOLMaster in non-orthokeratology subjects and post-orthokeratology children.

Ben Chan1, Pauline Cho, Sin Wan Cheung.   

Abstract

PURPOSE: Our aim was to determine the repeatability of measurements of axial length (AL) and anterior chamber depth (ACD) made with two ultrasonic biometers and the IOLMaster in a group of non-orthokeratology (ortho-k) adult subjects and to investigate the agreement among instruments in children undergoing ortho-k therapy and in children wearing spectacles.
METHODS: To determine repeatability, AL and ACD were measured twice in 22 non-ortho-k young adults using two A-scan ultrasonic biometers (A-5500 and A-2500) and the IOLMaster. To determine agreement, AL and ACD were measured with the same instruments in 30 children undergoing ortho-k therapy and 30 spectacle-wearing children.
RESULTS: In the adult subjects, there were no significant differences in ACD and AL measurements obtained from the three instruments (repeated measures ANOVAs, p > 0.05). There was also no significant between-measurement difference for each instrument. The between-measurement agreement was better for the IOLMaster (95% limits of agreement (LA): -0.04 and +0.05 mm for both AL and ACD) than for the two A-scan ultrasonic biometers (95% LA: -0.12 and +0.11 mm for AL; -0.22 and +0.27 mm for ACD). Among the children, AL measurements with all three instruments were not significantly different from each other for both the children undergoing ortho-k therapy and those wearing spectacles (repeated measures ANOVAs, p > 0.05). The 95% LA of differences obtained from any two instruments were also comparable for both groups of subjects (within -0.20 mm and +0.20 mm). ACD measurements of the children were significantly different among the three instruments (repeated measures ANOVAs, p < 0.05). No significant differences in ACD measurements were found between A-5500 and A-2500 for both groups of children (paired t tests, p > 0.017).
CONCLUSIONS: The repeatability of AL and ACD measurements with the IOLMaster was very good, and was better than with the A-scan ultrasonic biometers. The agreements in AL measurements between A-scan ultrasonic biometers and IOLMaster were comparable in both the ortho-k and the spectacle-wearing subjects, and were comparable to the repeatability of the A-scan ultrasonic biometers. ACD measurements between A-scan ultrasonic biometry and the IOLMaster were not comparable. AL measurements with the IOLMaster can replace the measurements from the two A-scan ultrasonic biometers used, however, the reverse is not true. AL and ACD measurements with all three instruments were unaffected by the flattened cornea following ortho-k lens wear.

Entities:  

Mesh:

Year:  2006        PMID: 16637971     DOI: 10.1111/j.1444-0938.2006.00029.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  11 in total

1.  Errors associated with IOLMaster biometry as a function of internal ocular dimensions.

Authors:  Miguel Faria-Ribeiro; Daniela Lopes-Ferreira; Norberto López-Gil; Jorge Jorge; José Manuel González-Méijome
Journal:  J Optom       Date:  2014-01-28

2.  Additive effects of orthokeratology and atropine 0.01% ophthalmic solution in slowing axial elongation in children with myopia: first year results.

Authors:  Nozomi Kinoshita; Yasuhiro Konno; Naoki Hamada; Yoshinobu Kanda; Machiko Shimmura-Tomita; Akihiro Kakehashi
Journal:  Jpn J Ophthalmol       Date:  2018-07-04       Impact factor: 2.447

3.  Comparison of the biometric values obtained by two different A-mode ultrasound devices (Eye Cubed vs. PalmScan): a transversal, descriptive, and comparative study.

Authors:  Raul Velez-Montoya; Eugene Mark Shusterman; Miriam Jessica López-Miranda; Mariana Mayorquin-Ruiz; Guillermo Salcedo-Villanueva; Hugo Quiroz-Mercado; Virgilio Morales-Cantón
Journal:  BMC Ophthalmol       Date:  2010-03-24       Impact factor: 2.209

4.  Semiautomated Assessment of Anterior Segment Structures in Pediatric Glaucoma Using Quantitative Ultrasound Biomicroscopy.

Authors:  Janet L Alexander; Snehaa Maripudi; Karun Kannan; Jennifer Drechsler; Moran R Levin; Osamah J Saeedi; Mona Kaleem; Marlet Bazemore; Bethany Karwoski; Camilo Martinez; Mohamad Jaafar; William P Madigan
Journal:  J Glaucoma       Date:  2021-05-01       Impact factor: 2.503

5.  Corneal Structural Changes in Congenital Glaucoma.

Authors:  Jennifer Drechsler; Adrianna Lee; Snehaa Maripudi; Laura Kueny; Moran R Levin; Osamah J Saeedi; Marlet Bazemore; Bethany Karwoski; Richard Birdsong; Camilo Martinez; Mohamad S Jaafar; Sairah Yousaf; Zubair M Ahmed; William P Madigan; Janet Leath Alexander
Journal:  Eye Contact Lens       Date:  2022-01-01       Impact factor: 3.152

6.  Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents.

Authors:  Mehmet Kola; Hikmet Duran; Adem Turk; Suleyman Mollamehmetoglu; Ahmet Kalkisim; Hidayet Erdol
Journal:  J Ophthalmol       Date:  2014-07-22       Impact factor: 1.909

7.  Factors related to axial length elongation and myopia progression in orthokeratology practice.

Authors:  Bingjie Wang; Rajeev K Naidu; Xiaomei Qu
Journal:  PLoS One       Date:  2017-04-18       Impact factor: 3.240

8.  Effect of orthokeratology on precision and agreement assessment of a new swept-source optical coherence tomography biometer.

Authors:  Bao Shu; Fangjun Bao; Giacomo Savini; Weicong Lu; Ruixue Tu; Haisi Chen; Benhao Song; Qinmei Wang; Jinhai Huang
Journal:  Eye Vis (Lond)       Date:  2020-03-02

9.  Short-term stability in refractive status despite large fluctuations in glucose levels in diabetes mellitus type 1 and 2.

Authors:  Byki Huntjens; W Neil Charman; Helena Workman; Sarah L Hosking; Clare O'Donnell
Journal:  PLoS One       Date:  2012-12-28       Impact factor: 3.240

Review 10.  Efficacy and Acceptability of Orthokeratology for Slowing Myopic Progression in Children: A Systematic Review and Meta-Analysis.

Authors:  Daizong Wen; Jinhai Huang; Hao Chen; Fangjun Bao; Giacomo Savini; Antonio Calossi; Haisi Chen; Xuexi Li; Qinmei Wang
Journal:  J Ophthalmol       Date:  2015-06-11       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.