Comprehensive Comparison of Axial Length Measurement With Three Swept-Source OCT-Based Biometers and Partial Coherence Interferometry.

Jinhai Huang, Hao Chen, Yue Li, Zhongxing Chen, Rongrong Gao, Jinjin Yu, Yune Zhao, Weicong Lu, Colm McAlinden, Qinmei Wang.   

Abstract

PURPOSE: To compare axial length measurements (and failure rate) of three swept-source optical coherence tomography (SS-OCT)-based biometers: IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany), OA-2000 (Tomey, Nagoya, Japan), and Argos (Movu Inc., Komaki, Japan) to those provided by a partial coherence interferometry (PCI)-based optical biometer (IOLMaster v5.4 [Carl Zeiss Meditec]).
METHODS: A total of 119 patients (171 eyes) undergoing cataract surgery were enrolled. Axial length was measured with the four biometers in a random order. Chi-square analysis was used to determine whether statistically significant differences in success rates were found between biometers. Within-subject standard deviation (Sw), test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficients (ICCs) were calculated to analyze the intraobserver repeatability. Bland-Altman plots were used for agreement assessment.
RESULTS: Axial length measurements were successfully measured in 166 eyes (97.08%) with the IOLMaster 700, 166 eyes (97.08%) with the OA-2000, 170 eyes (99.42%) with the Argos, and 138 eyes (80.70%) with the IOLMaster v5.4. Chi-square analysis indicated a significant difference in the failure rate between PCI- and SS-OCT-based biometers (P < .001). Intraobserver repeatability for the IOLMaster 700, OA-2000, and Argos showed excellent repeatability with low TRT (0.03, 0.06, and 0.05 mm, respectively), low CoV (0.04%, 0.10%, 0.07%, respectively), and high ICC (1.000, 0.999, and 1.000, respectively). The Bland-Altman 95% limits of agreement were as narrow as 0.09 mm, indicating excellent agreement among the three SS-OCT biometers and the PCI biometer.
CONCLUSIONS: The three SS-OCT biometers showed a significantly higher success rate for axial length measurement than the IOLMaster v5.4 in various lens opacities. These SS-OCT biometers are likely to become the gold standard for axial measurement. [J Refract Surg. 2019;35(2):115-120.].
© 2019 Huang, Chen, Li, et al.

Entities:  

Mesh:

Year:  2019        PMID: 30742226     DOI: 10.3928/1081597X-20190109-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  16 in total

1.  Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.

Authors:  Shi-Ming Cheng; Jia-Sheng Zhang; Xu Shao; Ze-Tong Wu; Tian-Tian Li; Po Wang; Jun-Hai Lin; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-02-16       Impact factor: 3.117

2.  Refractive error characteristics and influence on ocular parameters in patients with unilateral congenital ptosis.

Authors:  Yingli Liu; Tingting Chen; Jingwen Huang; Wentao Li; Yilin Chen; Lijun Huo
Journal:  BMC Ophthalmol       Date:  2022-07-02       Impact factor: 2.086

3.  Comparison of acquisition rate and agreement of axial length with two swept-source optical coherence tomographers and a partial coherence interferometer.

Authors:  Shi-Ming Cheng; Wen-Tao Yan; Jia-Sheng Zhang; Tian-Tian Li; Xi Li; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-04-30       Impact factor: 3.535

4.  Comparative Analysis of Swept-Source Optical Coherence Tomography and Partial Coherence Interferometry Biometers in the Prediction of Cataract Surgery Refractive Outcomes.

Authors:  Arthur B Cummings; Stephanie Naughton; Ann-Marie Coen; Elizabeth Brennan; Gabrielle E Kelly
Journal:  Clin Ophthalmol       Date:  2020-12-02

5.  Effect of orthokeratology on precision and agreement assessment of a new swept-source optical coherence tomography biometer.

Authors:  Bao Shu; Fangjun Bao; Giacomo Savini; Weicong Lu; Ruixue Tu; Haisi Chen; Benhao Song; Qinmei Wang; Jinhai Huang
Journal:  Eye Vis (Lond)       Date:  2020-03-02

6.  Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devices.

Authors:  Richul Oh; Joo Youn Oh; Hyuk Jin Choi; Mee Kum Kim; Chang Ho Yoon
Journal:  BMC Ophthalmol       Date:  2021-01-27       Impact factor: 2.209

7.  Toric Intraocular Lenses for the Management of Corneal Astigmatism at the Time of Cataract Surgery.

Authors:  Colm McAlinden; David Janicek
Journal:  J Ophthalmol       Date:  2021-12-18       Impact factor: 1.909

8.  Agreement of ocular biometric measurements in young healthy eyes between IOLMaster 700 and OA-2000.

Authors:  Xuan Liao; Yue Peng; Bo Liu; Qing-Qing Tan; Chang-Jun Lan
Journal:  Sci Rep       Date:  2020-02-21       Impact factor: 4.379

9.  Changes in ocular biometric measurements after vitrectomy with silicone oil tamponade for rhegmatogenous retinal detachment repair.

Authors:  Rui Liu; Qingchen Li
Journal:  BMC Ophthalmol       Date:  2020-09-02       Impact factor: 2.209

10.  Development and Verification of an Adjustment Factor for Determining the Axial Length Using Optical Biometry in Silicone Oil-Filled Eyes.

Authors:  Gerd U Auffarth; Tadas Naujokaitis; Louise Blöck; Anna Daghbashyan; Jan Meis; Victor A Augustin; Ramin Khoramnia; Timur M Yildirim
Journal:  Diagnostics (Basel)       Date:  2022-01-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.