PURPOSE: To comprehensively assess the reliability of a new optical biometer (IOLMaster 700), based on swept-source optical coherence tomography (SS-OCT) and comparison with a standard biometer (IOLMaster 500), in healthy children, adults, and cataract patients.Entities:
Year: 2020 PMID: 32509343 PMCID: PMC7246393 DOI: 10.1155/2020/8946364
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Intraobserver repeatability outcomes for biometric measurements obtained using IOLMaster 700 swept-source optical coherence tomography in children.
| Parameter | Observer | Mean ± SD |
| TRT | CoV (%) | ICC (95% CI) |
|---|---|---|---|---|---|---|
| AL (mm) | 1st | 24.7 ± 1.07 | 0.01 | 0.03 | 0.04 | 1.000 (1.000 to 1.000) |
| 2nd | 24.7 ± 1.07 | 0.01 | 0.02 | 0.03 | 1.000 (1.000 to 1.000) | |
| CCT ( | 1st | 552.27 ± 33.6 | 2.80 | 7.76 | 0.51 | 0.993 (0.990 to 0.995) |
| 2nd | 551.78 ± 33.2 | 2.86 | 7.93 | 0.52 | 0.993 (0.989 to 0.995) | |
| ACD (mm) | 1st | 3.76 ± 0.24 | 0.02 | 0.05 | 0.48 | 0.994 (0.992 to 0.996) |
| 2nd | 3.76 ± 0.23 | 0.01 | 0.04 | 0.40 | 0.996 (0.994 to 0.997) | |
| LT (mm) | 1st | 3.38 ± 0.15 | 0.02 | 0.07 | 0.72 | 0.973 (0.962 to 0.981) |
| 2nd | 3.38 ± 0.15 | 0.02 | 0.06 | 0.61 | 0.981 (0.974 to 0.987) | |
| Km (D) | 1st | 42.73 ± 1.55 | 0.09 | 0.26 | 0.22 | 0.996 (0.995 to 0.997) |
| 2nd | 42.73 ± 1.55 | 0.11 | 0.30 | 0.25 | 0.995 (0.993 to 0.997) | |
| WTW (mm) | 1st | 12.28 ± 0.43 | 0.09 | 0.26 | 0.75 | 0.955 (0.938 to 0.968) |
| 2nd | 12.27 ± 0.42 | 0.09 | 0.26 | 0.76 | 0.952 (0.934 to 0.966) |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, K = keratometry, WTW = white to white, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Intraobserver repeatability outcomes for biometric measurements obtained using IOLMaster 700 swept-source optical coherence tomography in normal subjects.
| Parameter | Observer | Mean ± SD |
| TRT | CoV (%) | ICC (95% CI) |
|---|---|---|---|---|---|---|
| AL (mm) | 1st | 25.1 ± 1.85 | 0.01 | 0.02 | 0.02 | 1.000 (1.000 to 1.000) |
| 2nd | 25.1 ± 1.85 | 0.01 | 0.02 | 0.03 | 1.000 (1.000 to 1.000) | |
| CCT ( | 1st | 542.84 ± 31.96 | 2.21 | 6.11 | 0.41 | 0.995 (0.993 to 0.997) |
| 2nd | 542.54 ± 32.13 | 2.20 | 6.10 | 0.41 | 0.995 (0.993 to 0.997) | |
| ACD (mm) | 1st | 3.56 ± 0.27 | 0.01 | 0.04 | 0.40 | 0.997 (0.996 to 0.998) |
| 2nd | 3.56 ± 0.28 | 0.02 | 0.04 | 0.43 | 0.997 (0.996 to 0.998) | |
| LT (mm) | 1st | 3.75 ± 0.27 | 0.02 | 0.04 | 0.41 | 0.997 (0.996 to 0.998) |
| 2nd | 3.75 ± 0.27 | 0.02 | 0.05 | 0.45 | 0.996 (0.995 to 0.997) | |
| Km (D) | 1st | 43.72 ± 1.36 | 0.10 | 0.28 | 0.23 | 0.994 (0.992 to 0.996) |
| 2nd | 43.73 ± 1.35 | 0.09 | 0.25 | 0.21 | 0.996 (0.994 to 0.997) | |
| WTW (mm) | 1st | 12.04 ± 0.35 | 0.09 | 0.24 | 0.72 | 0.940 (0.917 to 0.958) |
| 2nd | 12.04 ± 0.37 | 0.09 | 0.25 | 0.74 | 0.943 (0.920 to 0.960) |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, K = keratometry, WTW = white to white, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Intraobserver repeatability outcomes for biometric measurements obtained using IOLMaster 700 swept-source optical coherence tomography in cataract patients.
| Parameter | Observer | Mean ± SD |
| TRT | CoV (%) | ICC (95% CI) |
|---|---|---|---|---|---|---|
| AL (mm) | 1st | 23.6 ± 1.73 | 0.02 | 0.04 | 0.07 | 1.000 (1.000 to 1.000) |
| 2nd | 23.6 ± 1.73 | 0.01 | 0.02 | 0.04 | 1.000 (1.000 to 1.000) | |
| CCT ( | 1st | 534.09 ± 32.64 | 2.48 | 6.88 | 0.46 | 0.994 (0.992 to 0.996) |
| 2nd | 533.69 ± 32.44 | 2.61 | 7.24 | 0.49 | 0.994 (0.991 to 0.995) | |
| ACD (mm) | 1st | 2.87 ± 0.36 | 0.01 | 0.02 | 0.25 | 1.000 (0.999 to 1.000) |
| 2nd | 2.88 ± 0.36 | 0.02 | 0.06 | 0.73 | 0.997 (0.995 to 0.998) | |
| LT (mm) | 1st | 4.61 ± 0.37 | 0.01 | 0.02 | 0.13 | 1.000 (1.000 to 1.000) |
| 2nd | 4.61 ± 0.37 | 0.01 | 0.02 | 0.16 | 1.000 (0.999 to 1.000) | |
| Km (D) | 1st | 44.29 ± 1.57 | 0.09 | 0.26 | 0.21 | 0.996 (0.995 to 0.997) |
| 2nd | 44.28 ± 1.57 | 0.10 | 0.28 | 0.23 | 0.996 (0.994 to 0.997) | |
| WTW (mm) | 1st | 11.76 ± 0.41 | 0.12 | 0.33 | 1.01 | 0.922 (0.895 to 0.944) |
| 2nd | 11.75 ± 0.43 | 0.15 | 0.41 | 1.25 | 0.893 (0.856 to 0.922) |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, K = keratometry, WTW = white to white, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Interobserver reproducibility outcomes for biometric measurements obtained using an IOLMaster 700 swept-source optical coherence tomography-based biometer in children.
| Parameter | Methods |
| TRT | CoV (%) | ICC |
|---|---|---|---|---|---|
| AL (mm) | Average | 0.005 | 0.01 | 0.02 | 1.000 (1.000 to 1.000) |
| Single | 0.01 | 0.02 | 0.03 | 1.000 (1.000 to 1.000) | |
| CCT ( | Average | 1.96 | 5.42 | 0.35 | 0.997 (0.995 to 0.998) |
| Single | 2.59 | 7.18 | 0.47 | 0.994 (0.991 to 0.996) | |
| ACD (mm) | Average | 0.01 | 0.04 | 0.34 | 0.997 (0.996 to 0.998) |
| Single | 0.02 | 0.05 | 0.52 | 0.993 (0.989 to 0.995) | |
| LT (mm) | Average | 0.02 | 0.05 | 0.52 | 0.986 (0.979 to 0.991) |
| Single | 0.03 | 0.07 | 0.75 | 0.969 (0.954 to 0.979) | |
| Km (D) | Average | 0.05 | 0.14 | 0.12 | 0.999 (0.998 to 0.999) |
| Single | 0.09 | 0.25 | 0.21 | 0.997 (0.995 to 0.998) | |
| CD (mm) | Average | 0.07 | 0.18 | 0.54 | 0.976 (0.965 to 0.984) |
| Single | 0.10 | 0.29 | 0.85 | 0.943 (0.917 to 0.962) |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, Km = mean keratometry, CD = corneal diameter, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Interobserver reproducibility outcomes for biometric measurements obtained using an IOLMaster 700 swept-source optical coherence tomography-based biometer in normal subjects.
| Parameter | Methods |
| TRT | CoV (%) | ICC |
|---|---|---|---|---|---|
| AL (mm) | Average | 0.01 | 0.01 | 0.02 | 1.000 (1.000 to 1.000) |
| Single | 0.01 | 0.02 | 0.03 | 1.000 (1.000 to 1.000) | |
| CCT ( | Average | 1.97 | 5.47 | 0.36 | 0.996 (0.994 to 0.997) |
| Single | 2.63 | 7.30 | 0.49 | 0.993 (0.990 to 0.995) | |
| ACD (mm) | Average | 0.01 | 0.03 | 0.28 | 0.999 (0.998 to 0.999) |
| Single | 0.02 | 0.04 | 0.43 | 0.997 (0.995 to 0.998) | |
| LT (mm) | Average | 0.01 | 0.03 | 0.32 | 0.998 (0.997 to 0.999) |
| Single | 0.02 | 0.05 | 0.47 | 0.996 (0.994 to 0.997) | |
| Km (D) | Average | 0.07 | 0.19 | 0.15 | 0.998 (0.996 to 0.998) |
| Single | 0.10 | 0.27 | 0.23 | 0.995 (0.992 to 0.996) | |
| CD (mm) | Average | 0.06 | 0.18 | 0.53 | 0.968 (0.952 to 0.979) |
| Single | 0.10 | 0.26 | 0.79 | 0.929 (0.895 to 0.953) |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, Km = mean keratometry, CD = corneal diameter, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Interobserver reproducibility outcomes for biometric measurements obtained using an IOLMaster 700 swept-source optical coherence tomography-based biometer in cataract patients.
| Parameter | Methods |
| TRT | CoV (%) | ICC |
|---|---|---|---|---|---|
| AL (mm) | Average | 0.01 | 0.02 | 0.03 | 1.000 (1.000 to 1.000) |
| Single | 0.02 | 0.05 | 0.08 | 1.000 (1.000 to 1.000) | |
| CCT ( | Average | 1.93 | 5.36 | 0.36 | 0.996 (0.995 to 0.998) |
| Single | 2.80 | 7.77 | 0.53 | 0.993 (0.989 to 0.995) | |
| ACD (mm) | Average | 0.01 | 0.03 | 0.34 | 0.999 (0.999 to 0.999) |
| Single | 0.03 | 0.07 | 0.92 | 0.995 (0.992 to 0.996) | |
| LT (mm) | Average | 0.00 | 0.01 | 0.09 | 1.000 (1.000 to 1.000) |
| Single | 0.01 | 0.02 | 0.16 | 1.000 (0.999 to 1.000) | |
| Km (D) | Average | 0.07 | 0.18 | 0.15 | 0.998 (0.997 to 0.999) |
| Single | 0.09 | 0.26 | 0.21 | 0.996 (0.995 to 0.998) | |
| CD (mm) | Average | 0.10 | 0.27 | 0.82 | 0.948 (0.925 to 0.964) |
| Single | 0.16 | 0.44 | 1.36 | 0.863 (0.805 to 0.904) |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, Km = mean keratometry, CD = corneal diameter, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Figure 1Bland–Altman plot of the axial length, anterior chamber depth, mean keratometry and corneal diameter measurements using the IOLMaster 700 swept-source optical coherence tomography against IOLMaster partial coherence interferometry. The solid line represents the mean difference. The dotted lines on the side represent the upper and lower 95% LoA.
Figure 2Bland–Altman plot of the axial length measurements and the corneal diameter measurements using IOLMaster 700 swept-source optical coherence tomography against IOLMaster partial coherence interferometry in children (in the first row), normal subjects (in the second row), and cataract patients (in the third row). The solid line represents the mean difference. The dotted lines on the side represent the upper and lower 95% LoA.