Caryl E Richards1, Daniel R Obaid1,2.
Abstract
BACKGROUND: Coronary computed tomography angiography (CCTA) is now widely used in the diagnosis of coronary artery disease since it is a rapid, minimally invasive test with a diagnostic accuracy comparable to coronary angiography. However, to meet demands for increasing spatial and temporal resolution, higher x-ray radiation doses are required to circumvent the resulting increase in image noise. Exposure to high doses of ionizing radiation with CT imaging is a major health concern due to the potential risk of radiation-associated malignancy. Given its increasing use, a number of dose saving algorithms have been implemented to CCTA to minimize radiation exposure to "as low as reasonably achievable (ALARA)" without compromising diagnostic image quality.Entities:
Keywords: Coronary computed tomography angiography; effective radiation dose; image quality; iterative reconstruction; prospective electrocardiogram-zzm321990gating; tube current modulation; tube voltage reduction.
Mesh:
Year: 2019 PMID: 30806322 PMCID: PMC8142354 DOI: 10.2174/1573403X15666190222163737
Source DB: PubMed Journal: Curr Cardiol Rev ISSN: 1573-403X
Dose reduction strategies in coronary CT angiography.
|
|
|
|---|---|
| Iterative reconstruction | Improves signal-to-noise and contrast-to-noise ratio making it possible to maintain image quality, even when current is reduced. |
| Multi-row detectors | z-axis coverage of the scan is linearly proportional to radiation dose. |
| High-pitch prospective ECG-triggered helical acquisition (recommended for low and stable heart rates). | Projection data is acquired for only part of the complete gantry rotation (i.e., a partial scan). The minimum projection data required to construct a complete CT image is 180° plus the fan angle of the CT detectors in the axial plane. Full tube current is only applied during a single phase of the cardiac cycle. |
| Heart rate reduction with beta-blockers | Minimum cardiac motion is observed during diastolic phase; however, the diastolic phase narrows with increasing heart rate. |
| Automated tube current modulation | CT dose decreases linearly with tube current and tube current–time product. |
| Tube voltage | CT dose is approximately proportional to the square of the tube voltage. |
Studies with ultra-low-dose coronary CT angiography with mean effective radiation dose < 2.2 mSv.
|
|
|
|
|
|
|
| ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Yin | 21 | 50 ± 6 | 23.9 ± 3.2 | 128 detector row, second-generation dual Source CT (Definition Flash, Siemens Healthcare, Forchheim, Germany). | High-pitch spiral acquisition and raw data-based iterative reconstruction. Tube voltage was set to 80 kV and tube current was 50 mAs. | 0.06 ± 0.01 | ||||||||||
| Zhang | 58 | 60.4 ± 5.7 | 22.5 ± 1.9 | Dual-source CT system (Somatom Flash; Siemens Medical Solutions, Forchheim, Germany). | Prospective ECG-triggered high-pitch spiral acquisition (3.4) at 70 kVp with 30 mL of contrast agent. Automated tube current modulation (CAREDose 4D, Siemens). Image reconstruction with iterative reconstruction SAFIRE (SAFIRE, Siemens, strength-level 3). | 0.17 ± 0.02 | ||||||||||
| Zhang | 43 | 69.4 ± 13.6 | 23.3 ± 2.8 | 64-slice, second-generation dual-source CT system (Somatom Flash; Siemens Medical Solutions, Forchheim, Germany). | Prospective ECG-triggered high-pitch spiral acquisition (3.4) at 70 kVp. Automated tube current modulation (CAREDose 4D, Siemens) was enabled. | 0.20 ± 0.00 | ||||||||||
| Stehli | 36 | 57.6 ± 6.2 | 27.6 ± 4.7 | 64-slice CT scanner (Discovery HD 750, GE Healthcare). | Prospective ECG triggering. Body mass index (BMI)–adapted tube voltage and tube current. Images reconstructed using MBIR algorithms. | 0.29 ± 0.12 | ||||||||||
| Hell | 26 | 54 ± 5 | 27.7 ± 3.8 | 192-slice, third-generation dual-source CT system (Somatom Force; Siemens Healthcare, Forchheim, Germany). | A prospectively ECG-triggered high-pitch (3.2) spiral (flash) acquisition was performed. | 0.30 ± 0.03 | ||||||||||
| Gordic | 25* | 61 ± 5 | 25.3 ± 3.4 | 192-slice dual-source CT (SOMATOM Force, Siemens Healthcare). | Prospective ECG-triggered, high-pitch spiral acquisition. Automated attenuation-based tube voltage selection (CAREkV, Siemens) and tube current modulation (CAREDose, Siemens) was applied. In Images reconstructed with ADMIRE (ADMIRE; Siemens Healthcare, Forchheim, strength level 4). | 0.3 ± 0.1 | ||||||||||
| Wang | 40* | 55.4 ± 4.8 | 20.6 ± 1.4 | Dual source CT scanner (Definition Flash, Siemens AG, Forchheim, Germany). | Prospective ECG-triggering high-pitch spiral (3.4) at 80 kV. | 0.36 ± 0.03 | ||||||||||
| Yin | 40 | 54 ± 4 | 25.5 ± 3.1 | 64-slice, second-generation dual-source CT system (Somatom Definition Flash, Siemens Healthcare, Forchheim, Germany). | Prospective ECG-triggered high-pitch spiral acquisition (3.4). Attenuation-based tube current modulation (CareDose 4D, Siemens). | 0.58 ± 0.17 | ||||||||||
| Neefjes | 80* | 58 ± 7 | 28 ± 4 | 64-slice, dual-source CT scanner (Somatom Definition Flash; Siemens Healthcare, Forchheim, Germany). | Prospective ECG-triggered high-pitch spiral (3.4) | 0.81 ± 0.3 | ||||||||||
| Achenbach | 50 | 68 ± 9 | 25.4 ± 4.8 | 128 slice, dual-source CT system (‘Definition Flash’, Siemens Healthcare, Forchheim, Germany) | Prospective ECG-triggered high-pitch spiral acquisition ((pitch was 3.2 in first 28 patients and 3.4 in last 22 patients. Tube voltage was 100 kV and tube current was 320 mA s/rot. | 0.87 ± 0.07 | ||||||||||
| Alkadhi | 50* | 56 ± 10 | 25.9 ± 2.8 | 128-slice, second-generation dual-source CT scanner (Somatom Definition Flash; Siemens Healthcare, Forchheim, Germany). | Prospectively ECG-gated high-pitch mode (3.4). | 0.9 ± 0.1 | ||||||||||
| Chen | 107* | 57.1 ± 11.2 | 27.3 (24.6–32.3) | 320 row, second-generation CT scanner (Aquilion One Vision Edition; Toshiba Medical Systems, Otawara, Japan). | Prospectively ECG-gated. Tube potential and tube current were determined with use of automatic exposure control (SURE Exposure3D, Toshiba Medical Systems). Image reconstruction with iterative reconstruction AIDR-3D ((Toshiba Medical Systems). | 0.93 | ||||||||||
| Stolzmann | 100 | 66 ± 20 | 27.7 ± 4.2 | Second-generation dual-source 128-MDCT CT scanner (Somatom Definition Flash, Siemens Healthcare). | Prospective ECG-triggered high-pitch spiral acquisition (3.4). Tube voltage 100 kVp and tube current-time product 320 mAs per rotation | 1.0 ± 0.2 | ||||||||||
| Lell | 25 | 69 ± 9 | 26.8 ± 5.6 | 64 detector row, dual source CT system (Definition Flash, Siemens Healthcare, Forchheim, Germany). | Prospective ECG-triggered high-pitch spiral acquisition (3.2). Tube settings were 100 kV/320 mAs and 120 kV/400 mAs for patients below and above 100-kg weight, respectively. | 1.0 ± 0.3 | ||||||||||
| Sun | 134* | 79 ± 9 | 23.8 ± 2.7 | Second-generation DSCT system (Somatom Definition Flash, Siemens Healthcare, Forchheim, Germany). | Prospective ECG-gated high-pitch spiral acquisition (3.4). Tube voltage 100 kVp and tube current-time product 320 mAs per rotation | 1.04 ± 0.16 | ||||||||||
| Sommer | 33 | 57.3 ± 7.0 | 24.7 ± 2.0 | Second generation dual source CT scanner (Somatom Definition Flash, Siemens Healthcare). | Prospective ECG-triggered high-pitch spiral acquisition (3.4). | 1.11 ± 0.14 | ||||||||||
| Wang | 172* | 59.8 ± 7.6 | 24.2 ± 2.5 | Dual-source CT scanner (Somatom Definition Flash; Siemens, Munich, Germany). | Prospective ECG-triggered high-pitch helical mode (flash mode) d if the patient’s HR < 65 bpm and prospectively ECG-triggered sequential mode if the patient’s HR between 65 and 90 bpm. Automatic tube potential selection (CARE kV; Siemens) and automatic tube current selection algorithm (CARE Dose4D; Siemens). | 1.25 ± 1.24 | ||||||||||
| Wichmann | 25* | 53 ± 2 | … | 128-slice second-generation dual-source CT (Somatom Definition Flash, Siemens Healthcare). | Prosepctive ECG-gated high-pitch spiral. Automatic tube potential selection (CARE kV; Siemens) and automatic tube current selection algorithm (CARE Dose4D; Siemens). | 1.27 ± 0.62 | ||||||||||
| Huang | 70* | 58.6 ± 4.5 | 24.8 ± 2.8 | 64-slice DSCT scanner (Somatom Definition, Siemens Healthcare, Forchheim, Germany). | Single-phase, prospective ECG-triggered acquisition. | 1.27 ± 0.57 | ||||||||||
| Koplay | 186 | 66.52 ± 11 | 27.97 (19–40) | 128 slice dual-source CT (Somatom Definition Flash, Siemens, Germany). | Prospective ECG-triggered high-pitch spiral acquisition. Automatic tube potential selection (CARE kV; Siemens) and automatic tube current selection algorithm (CARE Dose4D; Siemens). | 1.3 ± 0.4 | ||||||||||
| Leipsic | 24* | 54 ± 5 | 27 ± 4 | Discovery HD 750 (GE Healthcare, Waukasha, WI, USA). | Prospective ECG-triggering. Reduced tube voltage was defined as 80 or 100 kVp for individuals with BMI < 25 kg/m2 or 25–35 kg/m2, respectively; whereas standard tube voltage was defined as 100 or 120 kVp for individuals with BMI < 25 kg/m2 or 25–35 kg/m2, respectively. | 1.3 ± 0.5 | ||||||||||
| Matsubara | 17* | 59.1 ± 6.0 | 21.0 ± 2.0 | 128-slice dual-source CT Somatom Definition Flash scanner (Siemens Healthcare) | Prospective ECG-triggered high-pitch spiral acquisition (3.4). | 1.5 ± 0.2 | ||||||||||
| Pflederer | 56* | … | … | Dual-source CT (Definition, Siemens Healthcare, Forchheim, Germany). | Prospective ECG-triggering. Tube settings were 100 kV/330 mAs. | 1.5 ± 0.4 | ||||||||||
| Mangold | 43* | 68.1 ± 18.4 | 23.5 ± 3.3 | 3rd generation dual-source CT (Somatom Force, Siemens Healthcare, Forchheim, Germany). | Prospective ECG-triggered spiral acquisition. Automated tube current selection and advanced iterative reconstruction. Tube voltage at 70 kV group. | 1.5 ± 1.2 | ||||||||||
| Kim | 23* | 59 ± 7 | 24.3 ± 3.5 | 128-slice MDCT (Definition AS Plus 128; Siemens, Forchheim, Germany). | Step-and-shoot prospective ECG-gated group. | 1.75 ± 0.83 | ||||||||||
| Leipsic | 109* | 59 ± 4 | 27.23 ± 4.27 | Discovery HD 750 (GE Healthcare, Waukasha, WI, USA) and a Toshiba Aquilion One (Toshiba Medical Systems, Tokyo, Japan). | Prospective ECG-triggering with a narrow window acquisition window. | 1.78 | ||||||||||
| Deseive | 150* | 55.5 ± 5.0 | 26.1 ± 3.5 | Second-generation dual-source CT scanner (Somatom, Definition Flash scanner, Siemens Medical Solutions) | PROTECTION IV study. Prospective ECG-gated high-pitch helical acquisition. 100-kV tube potentials up to a body mass index of 30 kg/m2. | 2.0 ± 2.4 | ||||||||||
| Yoo | 51 | 55 (39–65) | 25.3 (18.8–32.8) | 640-multi-slice CT scanner (Aquilion ONE; Toshiba Medical Systems, Tochiki-ken, Japan). | Prospective ECG-triggering. Automatic exposure control system (SUREExposure; Toshiba Medical Systems, Tochiki-ken, Japan). Image reconstruction with AIDR-3D (standard). | 2 ± 1 | ||||||||||
| Duarte | 40* | 60 ± 5 | Weight/kg (70 ± 10) | 128-MDCT (Somatom Definition AS128, Siemens Medical Solutions, Germany). | Prospective ECG-gating with full tube current at 70%. Automated tube current to patient-specific parameters such as size and attenuation of body region (CAREDose system). | 2.1 ± 0.9 | ||||||||||
| Husmann | 100 | 57 ± 6 | 27 ± 4 | 64-slice, LightSpeed VCT XT scanner (GE Healthcare). | Prospective ECG-triggering. | 2.2 ± 0.7 | ||||||||||