Estimated radiation dose associated with cardiac CT angiography.

Jörg Hausleiter1, Tanja Meyer, Franziska Hermann, Martin Hadamitzky, Markus Krebs, Thomas C Gerber, Cynthia McCollough, Stefan Martinoff, Adnan Kastrati, Albert Schömig, Stephan Achenbach.   

Abstract

CONTEXT: Cardiac computed tomography (CT) angiography (CCTA) has emerged as a useful diagnostic imaging modality in the assessment of coronary artery disease. However, the potential risks due to exposure to ionizing radiation associated with CCTA have raised concerns.
OBJECTIVES: To estimate the radiation dose of CCTA in routine clinical practice as well as the association of currently available strategies with dose reduction and to identify the independent factors contributing to radiation dose. DESIGN, SETTING, AND PATIENTS: A cross-sectional, international, multicenter, observational study (50 study sites: 21 university hospitals and 29 community hospitals) of estimated radiation dose in 1965 patients undergoing CCTA between February and December 2007. Linear regression analysis was used to identify independent predictors associated with dose. MAIN OUTCOME MEASURE: Dose-length product (DLP) of CCTA.
RESULTS: The median DLP of 1965 CCTA examinations performed at 50 study sites was 885 mGy x cm (interquartile range, 568-1259 mGy x cm), which corresponds to an estimated radiation dose of 12 mSv (or 1.2 x the dose of an abdominal CT study or 600 chest x-rays). A high variability in DLP was observed between study sites (range of median DLPs per site, 331-2146 mGy x cm). Independent factors associated with radiation dose were patient weight (relative effect on DLP, 5%; 95% confidence interval [CI], 4%-6%), absence of stable sinus rhythm (10%; 95% CI, 2%-19%), scan length (5%; 95% CI, 4%-6%), electrocardiographically controlled tube current modulation (-25%; 95% CI, -23% to -28%; applied in 73% of patients), 100-kV tube voltage (-46%; 95% CI, -42% to -51%; applied in 5% of patients), sequential scanning (-78%; 95% CI, -77% to -79%; applied in 6% of patients), experience in cardiac CT (-1%; 95% CI, -1% to 0%), number of CCTAs per month (0%; 95% CI, 0%-1%), and type of 64-slice CT system (for highest vs lowest dose system, 97%; 95% CI, 88%-106%). Algorithms for dose reduction were not associated with deteriorated diagnostic image quality in this observational study.
CONCLUSIONS: Median doses of CCTA differ significantly between study sites and CT systems. Effective strategies to reduce radiation dose are available but some strategies are not frequently used. The comparable diagnostic image quality may support an increased use of dose-saving strategies in adequately selected patients.

Entities:  

Mesh:

Year:  2009        PMID: 19190314     DOI: 10.1001/jama.2009.54

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  279 in total

1.  Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience.

Authors:  Zhaoping Cheng; Ximing Wang; Yanhua Duan; Lebin Wu; Dawei Wu; Baoting Chao; Cheng Liu; Zhuodong Xu; Hongxin Li; Fei Liang; Jian Xu; Jiuhong Chen
Journal:  Eur Radiol       Date:  2010-06-09       Impact factor: 5.315

Review 2.  Diagnostic value of coronary CT angiography with prospective ECG-gating in the diagnosis of coronary artery disease: a systematic review and meta-analysis.

Authors:  Zhonghua Sun; Kwan-Hoong Ng
Journal:  Int J Cardiovasc Imaging       Date:  2012-01-04       Impact factor: 2.357

3.  Three-dimensional anisotropic adaptive filtering of projection data for noise reduction in cone beam CT.

Authors:  Andreas Maier; Lars Wigstrom; Hannes G Hofmann; Joachim Hornegger; Lei Zhu; Norbert Strobel; Rebecca Fahrig
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

Review 4.  Low-dose cardiovascular computed tomography: where are the limits?

Authors:  Paul Schoenhagen; Carla M Thompson; Sandra S Halliburton
Journal:  Curr Cardiol Rep       Date:  2012-02       Impact factor: 2.931

5.  Evidence for prospective ECG-triggering coronary CT angiography in routine practice.

Authors:  Ali Salavati; Frank J Rybicki
Journal:  Int J Cardiovasc Imaging       Date:  2012-02-09       Impact factor: 2.357

6.  The effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography.

Authors:  F Tatsugami; M Matsuki; G Nakai; Y Inada; S Kanazawa; Y Takeda; H Morita; H Takada; S Yoshikawa; K Fukumura; Y Narumi
Journal:  Br J Radiol       Date:  2012-01-17       Impact factor: 3.039

7.  DNA double-strand breaks as potential indicators for the biological effects of ionising radiation exposure from cardiac CT and conventional coronary angiography: a randomised, controlled study.

Authors:  Dominik Geisel; Elke Zimmermann; Matthias Rief; Johannes Greupner; Michael Laule; Fabian Knebel; Bernd Hamm; Marc Dewey
Journal:  Eur Radiol       Date:  2012-04-17       Impact factor: 5.315

8.  Declining radiation dose of coronary computed tomography angiography: German cardiac CT registry experience 2009-2014.

Authors:  Axel Schmermund; Mohamed Marwan; Jörg Hausleiter; Sebastian Barth; Oliver Bruder; Sebastian Kerber; Grigorius Korosoglou; Alexander Leber; Werner Moshage; Stephen Schröder; Steffen Schneider; Jochen Senges; Stephan Achenbach
Journal:  Clin Res Cardiol       Date:  2017-07-19       Impact factor: 5.460

9.  Standardization and optimization of CT protocols to achieve low dose.

Authors:  Sigal Trattner; Gregory D N Pearson; Cynthia Chin; Dianna D Cody; Rajiv Gupta; Christopher P Hess; Mannudeep K Kalra; James M Kofler; Mayil S Krishnam; Andrew J Einstein
Journal:  J Am Coll Radiol       Date:  2014-03       Impact factor: 5.532

Review 10.  Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography.

Authors:  Zhonghua Sun
Journal:  Cardiovasc Diagn Ther       Date:  2012-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.