Dose performance of a 64-channel dual-source CT scanner.

Cynthia H McCollough1, Andrew N Primak, Osama Saba, Herbert Bruder, Karl Stierstorfer, Rainer Raupach, Christoph Suess, Bernhard Schmidt, Bernd M Ohnesorge, Thomas G Flohr.   

Abstract

PURPOSE: To prospectively compare the dose performance of a 64-channel multi-detector row computed tomographic (CT) scanner and a 64-channel dual-source CT scanner from the same manufacturer.
MATERIALS AND METHODS: To minimize dose in the cardiac (dual-source) mode, the evaluated dual-source CT system uses a cardiac beam-shaping filter, three-dimensional adaptive noise reduction, heart rate-dependent pitch, and electrocardiographically based modulation of the tube current. Weighted CT dose index per 100 mAs was measured for the head, body, and cardiac beam-shaping filters. Kerma-length product was measured in the spiral cardiac mode at four pitch values and three electrocardiographic modulation temporal windows. Noise was measured in an anthropomorphic phantom. Data were compared with data from a 64-channel multi-detector row CT scanner.
RESULTS: For the multi-detector row and dual-source CT systems, respectively, weighted CT dose index per 100 mAs was 14.2 and 12.2 mGy (head CT), 6.8 and 6.4 mGy (body CT), and 6.8 and 5.3 mGy (cardiac CT). In the spiral cardiac mode (no electrocardiographically based tube current modulation, 0.2 pitch), equivalent noise occurred at volume CT dose index values of 23.7 and 35.0 mGy (coronary artery calcium CT) and 58.9 and 61.2 mGy (coronary CT angiography) for multi-detector row CT and dual-source CT, respectively. The use of heart rate-dependent pitch values reduced volume CT dose index to 46.2 mGy (0.265 pitch), 34.0 mGy (0.36 pitch), and 26.6 mGy (0.46 pitch) compared with 61.2 mGy for 0.2 pitch. The use of electrocardiographically based tube current-modulation and temporal windows of 110, 210, and 310 msec further reduced volume CT dose index to 9.1-25.1 mGy, dependent on the heart rate.
CONCLUSION: For electrocardiographically gated coronary CT angiography, image noise equivalent to that of multi-detector row CT can be achieved with dual-source CT at doses comparable to or up to a factor of two lower than the doses at multi-detector row CT, depending on heart rate of the patient. (c) RSNA, 2007.

Entities:  

Mesh:

Year:  2007        PMID: 17446525     DOI: 10.1148/radiol.2433061165

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  63 in total

1.  Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose.

Authors:  Eun-Ah Park; Whal Lee; Kwang Woo Kim; Kwang Gi Kim; Allmendinger Thomas; Jin Wook Chung; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-21       Impact factor: 2.357

2.  A strategy to decrease partial scan reconstruction artifacts in myocardial perfusion CT: phantom and in vivo evaluation.

Authors:  Juan C Ramirez-Giraldo; Lifeng Yu; Birgit Kantor; Erik L Ritman; Cynthia H McCollough
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

3.  Usefulness of dual-energy CT scanning at 80 kVp for identifying hilar and mediastinal structures: evaluation of contrast enhancement of the pulmonary vessels and lymph nodes.

Authors:  Ayano Imafuji; Masaki Hara; Shigeru Sasaki; Toshinao Arakawa; Yoshiyuki Ozawa; Yuta Shibamoto
Journal:  Jpn J Radiol       Date:  2011-12-03       Impact factor: 2.374

4.  Low dose dual-source CT angiography of the thoracic aorta.

Authors:  Cormac Farrelly; Amir Davarpanah; Aoife N Keeling; John Sheehan; Ann Ragin; Vahid Yaghmai; James C Carr
Journal:  Int J Cardiovasc Imaging       Date:  2010-11-03       Impact factor: 2.357

5.  Effect of CT scan protocols on x-ray-induced DNA double-strand breaks in blood lymphocytes of patients undergoing coronary CT angiography.

Authors:  M A Kuefner; S Grudzenski; J Hamann; S Achenbach; Michael Lell; K Anders; S A Schwab; L Häberle; M Löbrich; M Uder
Journal:  Eur Radiol       Date:  2010-07-13       Impact factor: 5.315

6.  Dose levels at coronary CT angiography--a comparison of Dual Energy-, Dual Source- and 16-slice CT.

Authors:  J Matthias Kerl; Ralf W Bauer; Tobias B Maurer; Rene Aschenbach; Huedayi Korkusuz; Thomas Lehnert; Simon Deseive; Hanns Ackermann; Thomas J Vogl
Journal:  Eur Radiol       Date:  2010-09-23       Impact factor: 5.315

7.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

Review 8.  Paediatric cardiac computed tomography: a review of imaging techniques and radiation dose consideration.

Authors:  Carolyn Young; Andrew M Taylor; Catherine M Owens
Journal:  Eur Radiol       Date:  2010-12-28       Impact factor: 5.315

Review 9.  Dual-source cardiac computed tomography: image quality and dose considerations.

Authors:  Stephan Achenbach; Katharina Anders; Willi A Kalender
Journal:  Eur Radiol       Date:  2008-02-26       Impact factor: 5.315

Review 10.  Functional imaging: CT and MRI.

Authors:  Edwin J R van Beek; Eric A Hoffman
Journal:  Clin Chest Med       Date:  2008-03       Impact factor: 2.878

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.