Literature DB >> 21070998

Image quality and radiation exposure with a low tube voltage protocol for coronary CT angiography results of the PROTECTION II Trial.

Jörg Hausleiter1, Stefan Martinoff, Martin Hadamitzky, Eugenio Martuscelli, Iris Pschierer, Gudrun M Feuchtner, Paz Catalán-Sanz, Benedikt Czermak, Tanja S Meyer, Franziska Hein, Bernhard Bischoff, Miriam Kuse, Albert Schömig, Stephan Achenbach.   

Abstract

OBJECTIVES: The purpose of this study was to evaluate image quality and radiation dose using a 100 kVp tube voltage scan protocol compared with standard 120 kVp for coronary computed tomography angiography (CTA).
BACKGROUND: Concerns have been raised about radiation exposure during coronary CTA. The use of a 100 kVp tube voltage scan protocol effectively lowers coronary CTA radiation dose compared with standard 120 kVp, but it is unknown whether image quality is maintained.
METHODS: We enrolled 400 nonobese patients who underwent coronary CTA: 202 patients were randomly assigned to a 100 kVp protocol and 198 patients to a 120 kVp protocol. The primary end point was to demonstrate noninferiority in image quality with the 100 kVp protocol, which was assessed by a 4-point grading score (1 = nondiagnostic, 4 = excellent image quality). For the noninferiority analysis, a margin of -0.2 image quality score points for the difference between both scan protocols was pre-defined. Secondary end points included radiation dose and need for additional diagnostic tests during follow-up.
RESULTS: The mean image quality scores in patients scanned with 100 kVp and 120 kVp were 3.30 ± 0.67 and 3.28 ± 0.68, respectively (p = 0.742); image quality of the 100 kVp protocol was not inferior, as demonstrated by the 97.5% confidence interval of the difference, which did not cross the pre-defined noninferiority margin of -0.2. The 100 kVp protocol was associated with a 31% relative reduction in radiation exposure (dose-length product: 868 ± 317 mGy × cm with 120 kVp vs. 599 ± 255 mGy × cm with 100 kVp; p < 0.0001). At 30-day follow-up, the need for additional diagnostic studies did not differ (13.4% vs. 19.2% for 100 kVp vs. 120 kVp, respectively; p = 0.114).
CONCLUSIONS: A coronary CTA protocol using 100 kVp tube voltage maintained image quality, but reduced radiation exposure by 31% as compared with the standard 120 kVp protocol. Thus, 100 kVp scan protocols should be considered for nonobese patients to keep radiation exposure as low as reasonably achievable. (Prospective Randomized Trial on Radiation Dose Estimates of Cardiac CT Angiography in Patients Scanned With a 100 kVp Protocol [PROTECTION II]; NCT00611780).
Copyright © 2010 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21070998     DOI: 10.1016/j.jcmg.2010.08.016

Source DB:  PubMed          Journal:  JACC Cardiovasc Imaging        ISSN: 1876-7591


  67 in total

1.  Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose.

Authors:  Eun-Ah Park; Whal Lee; Kwang Woo Kim; Kwang Gi Kim; Allmendinger Thomas; Jin Wook Chung; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-21       Impact factor: 2.357

Review 2.  Low-dose cardiovascular computed tomography: where are the limits?

Authors:  Paul Schoenhagen; Carla M Thompson; Sandra S Halliburton
Journal:  Curr Cardiol Rep       Date:  2012-02       Impact factor: 2.931

3.  Feasibility study of low tube voltage (80 kVp) coronary CT angiography combined with contrast medium reduction using iterative model reconstruction (IMR) on standard BMI patients.

Authors:  Fan Zhang; Li Yang; Xiang Song; Ying-Na Li; Yan Jiang; Xing-Hua Zhang; Hai-Yue Ju; Jian Wu; Rui-Ping Chang
Journal:  Br J Radiol       Date:  2015-11-26       Impact factor: 3.039

4.  Application of intelligent optimal kV scanning technology (CARE kV) in dual-source computed tomography (DSCT) coronary angiography.

Authors:  Jun Zhang; Shaolei Kang; Dan Han; Xiaojie Xie; Yaming Deng
Journal:  Int J Clin Exp Med       Date:  2015-10-15

5.  Cyclic changes in area- and perimeter-derived effective dimensions of the aortic annulus measured with multislice computed tomography and comparison with metric intraoperative sizing.

Authors:  Won-Keun Kim; Alexander Meyer; Helge Möllmann; Andreas Rolf; Susanne Möllmann; Johannes Blumenstein; Arnaud Van Linden; Christian W Hamm; Thomas Walther; Jörg Kempfert
Journal:  Clin Res Cardiol       Date:  2016-02-18       Impact factor: 5.460

6.  Adenosine-stress dynamic myocardial perfusion imaging using 128-slice dual-source CT in patients with normal body mass indices: effect of tube voltage, tube current, and iodine concentration on image quality and radiation dose.

Authors:  Sung Mok Kim; Young Kwon Cho; Yeon Hyeon Choe
Journal:  Int J Cardiovasc Imaging       Date:  2014-08-26       Impact factor: 2.357

7.  Quantitative analysis of left ventricular dyssynchrony using cardiac computed tomography versus three-dimensional echocardiography.

Authors:  Sebastian J Buss; Felix Schulz; David Wolf; Waldemar Hosch; Christian Galuschky; Georg Schummers; Evangelos Giannitsis; Hans-Ulrich Kauczor; Christian Zugck; Ruediger Becker; Stefan E Hardt; Hugo A Katus; Grigorios Korosoglou
Journal:  Eur Radiol       Date:  2012-01-22       Impact factor: 5.315

8.  Radiation dose of coronary CT angiography with a third-generation dual-source CT in a "real-world" patient population.

Authors:  Aleksander Kosmala; Bernhard Petritsch; Andreas Max Weng; Thorsten Alexander Bley; Tobias Gassenmaier
Journal:  Eur Radiol       Date:  2018-11-30       Impact factor: 5.315

9.  Effect of bismuth breast shielding on radiation dose and image quality in coronary CT angiography.

Authors:  Andrew J Einstein; Carl D Elliston; Daniel W Groves; Bin Cheng; Steven D Wolff; Gregory D N Pearson; M Robert Peters; Lynne L Johnson; Sabahat Bokhari; Gary W Johnson; Ketan Bhatia; Theodore Pozniakoff; David J Brenner
Journal:  J Nucl Cardiol       Date:  2011-11-09       Impact factor: 5.952

10.  Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study.

Authors:  Felix G Meinel; Christian Canstein; U Joseph Schoepf; Martin Sedlmaier; Bernhard Schmidt; Brett S Harris; Thomas G Flohr; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.