Jinhai Huang1, Na Liao2, Giacomo Savini3, Fangjun Bao1, Ye Yu1, Weicong Lu1, Qingjie Hu2, Qinmei Wang1.
Abstract
Purpose. To assess the consistency of anterior segment measurements obtained using a Sirius Scheimpflug/Placido photography-based topography system (CSO, Italy) and IOLMaster partial coherence interferometry (Carl Zeiss Meditec, Germany) in eyes with cataracts. Methods. A total of 90 eyes of 90 patients were included in this prospective study. The anterior chamber depth (ACD), keratometry (K), corneal astigmatism axis, and white to white (WTW) values were randomly measured three times with Sirius and IOLMaster. Concordance between them was assessed by calculating 95% limits of agreement (LoA). Results. The ACD and K taken with the Sirius were statistically significantly higher than that taken with the IOLMaster; however, the Sirius significantly underestimated the WTW values compared with the IOLMaster. Good agreement was found for Km and ACD measurements, with 95% LoA of -0.20 to 0.54 mm and -0.16 to 0.34 mm, respectively. Poor agreement was observed for astigmatism axis and WTW measurements, as the 95% LoA was -23.96 to 23.36° and -1.15 to 0.37 mm, respectively. Conclusion. With the exception of astigmatism axis and WTW, anterior segment measurements taken by Sirius and IOLMaster devices showed good agreement and may be used interchangeably in patients with cataracts.Entities:
Year: 2014 PMID: 25400939 PMCID: PMC4226174 DOI: 10.1155/2014/540760
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Descriptive statistics of ocular measurements by Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry.
| Parameter | Mean ± SD | Minimum | Maximum |
|---|---|---|---|
| Sirius | |||
| ACD (mm) | 3.13 ± 0.41 | 2.20 | 4.35 |
| Kf (D) | 44.19 ± 1.50 | 40.79 | 48.16 |
| Ks (D) | 44.95 ± 1.55 | 41.36 | 49.24 |
| Km (D) | 44.57 ± 1.50 | 41.09 | 48.59 |
| Axis (°) | 99.85 ± 52.50 | 3.7 | 180.0 |
| WTW (mm) | 11.42 ± 0.46 | 10.41 | 12.95 |
| IOLMaster | |||
| ACD (mm) | 3.04 ± 0.40 | 2.25 | 4.08 |
| Kf (D) | 43.97 ± 1.51 | 40.56 | 48.21 |
| Ks (D) | 44.83 ± 1.57 | 41.46 | 49.34 |
| Km (D) | 44.40 ± 1.51 | 41.06 | 48.64 |
| Axis (°) | 100.16 ± 53.46 | 3.7 | 180.0 |
| WTW (mm) | 11.81 ± 0.42 | 10.33 | 13.10 |
| AL (mm) | 23.47 ± 1.11 | 21.28 | 28.36 |
ACD = anterior chamber depth, Kf = flattest keratometry, Ks = steepest keratometry, Km = mean keratometry, WTW = white to white, AL = axial length, and SD = standard deviation.
The mean difference, limits of agreement (LoA), paired t-test for these differences, and their significance for difference parameters between Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry.
| Device pairings | Mean difference ± SD |
| 95% LoA |
|---|---|---|---|
| ACD (mm) | 0.09 ± 0.13 | <0.05 | −0.16 to 0.34 |
| Kf (D) | 0.22 ± 0.24 | <0.05 | −0.26 to 0.69 |
| Ks (D) | 0.12 ± 0.24 | <0.05 | −0.34 to 0.58 |
| Km (D) | 0.17 ± 0.19 | <0.05 | −0.20 to 0.54 |
| Axis (°) | −0.30 ± 12.07 | 0.813 | −23.96 to 23.36 |
| WTW (mm) | −0.39 ± 0.39 | <0.05 | −1.15 to 0.37 |
ACD = anterior chamber depth, Kf = flattest keratometry, Ks = steepest keratometry, Km = mean keratometry, WTW = white to white, and SD = standard deviation.
Figure 1Bland-Altman plots of agreement of anterior chamber depth measurements taken with the Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry. The solid line indicates the mean difference (bias). The upper and lower lines represent the 95% limits of agreement.
Figure 2Bland-Altman plots of agreement of keratometry values along the flattest meridian taken with the Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry. The solid line indicates the mean difference (bias). The upper and lower lines represent the 95% limits of agreement.
Figure 3Bland-Altman plots of agreement of keratometry values along the steepest meridian taken with the Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry. The solid line indicates the mean difference (bias). The upper and lower lines represent the 95% limits of agreement.
Figure 4Bland-Altman plots of agreement of mean Keratometry readings taken with the Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry. The solid line indicates the mean difference (bias). The upper and lower lines represent the 95% limits of agreement.
Figure 5Bland-Altman plots of agreement of mean corneal astigmatism axis readings taken with the Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry. The solid line indicates the mean difference (bias). The upper and lower lines represent the 95% limits of agreement.
Figure 6Bland-Altman plots of agreement of white to white readings taken with the Sirius Scheimpflug/Placido photography-based topography system and IOLMaster partial coherence interferometry. The solid line indicates the mean difference (bias). The upper and lower lines represent the 95% limits of agreement.