Literature DB >> 15030827

Comparison of manual and automated methods to determine horizontal corneal diameter.

Martin Baumeister1, Evdoxia Terzi, Yurdagül Ekici, Thomas Kohnen.   

Abstract

PURPOSE: To compare 2 manual methods and 2 automated devices for measuring the horizontal corneal diameter (white-to-white [WTW] distance).
SETTING: Department of Ophthalmology, Johann Wolfgang Goethe-University, Frankfurt am Main, Germany.
METHODS: In 100 eyes of 61 patients, the WTW distance was measured independently by 2 examiners using the following techniques: the Holladay-Godwin gauge, a measuring caliper, Zeiss IOLMaster, and Orbscan II topography system (Bausch & Lomb). The results were compared with measurements on magnified slitlamp photographs of the anterior eye segment. Statistical evaluation was performed using the Bland-Altman method for comparison of measurement techniques.
RESULTS: The mean horizontal corneal diameter was 11.91 mm +/- 0.71 (SD) with the measuring caliper, 11.8 +/- 0.60 mm with the Holladay-Godwin gauge, 11.78 +/- 0.43 mm with the Orbscan II, and 12.02 +/- 0.38 mm with the IOLMaster. The coefficient of inter-rater repeatability was 1.30 for the caliper, 0.92 for the Holladay-Godwin gauge, 0.76 for the Orbscan II, 0.50 for the IOLMaster, and 1.16 for the manual measurement in anterior segment images. The results obtained with the caliper were statistically significantly different between the 2 examiners (P<.001). The measurements of examiner 2 using the caliper were significantly different from the measurements of the same examiner using the Holladay-Godwin gauge. This was not the case with the 2 automated devices.
CONCLUSIONS: Automated measurement of the WTW distance provides more precise results than measurements using manual methods. The Zeiss IOLMaster had the highest reliability in measuring corneal diameter.

Entities:  

Mesh:

Year:  2004        PMID: 15030827     DOI: 10.1016/j.jcrs.2003.06.004

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  30 in total

1.  Changes in the crystalline lens resulting from insertion of a phakic IOL (ICL) into the porcine eye.

Authors:  Kunitoshi Fujisawa; Kimiya Shimizu; Shigekazu Uga; Masanobu Suzuki; Koichi Nagano; Yuuki Murakami; Hiroko Goseki
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-04-25       Impact factor: 3.117

2.  Evaluation of myopic corneal diameter with the Orbscan II Topography System.

Authors:  Yi Zha; Wangqiang Feng; Xiaohui Han; Jianqiu Cai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-06-02       Impact factor: 3.117

3.  Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

Review 4.  [Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation].

Authors:  T Kohnen; T Kasper; E Terzi
Journal:  Ophthalmologe       Date:  2005-11       Impact factor: 1.059

Review 5.  [Intraocular lenses for the correction of refraction errors. Part 1: phakic anterior chamber lenses].

Authors:  T Kohnen; M Baumeister; M Cichocki
Journal:  Ophthalmologe       Date:  2005-10       Impact factor: 1.059

6.  [Biometry of the anterior eye segment for implantation of phakic anterior chamber lenses. A comparison of current measurement devices].

Authors:  L Kiraly; G Duncker
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

7.  Corneal white-to-white distance and mesopic pupil diameter.

Authors:  Hasan Basri Cakmak; Nurullah Cagil; Huseyin Simavli; Sabri Raza
Journal:  Int J Ophthalmol       Date:  2012-08-18       Impact factor: 1.779

8.  White-to-white corneal diameter: normal values in healthy Iranian population obtained with the Orbscan II.

Authors:  Hamid Gharaee; Mojtaba Abrishami; Masoud Shafiee; Asieh Ehsaei
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

9.  Novel Image-Based Analysis for Reduction of Clinician-Dependent Variability in Measurement of the Corneal Ulcer Size.

Authors:  Tapan P Patel; N Venkatesh Prajna; Sina Farsiu; Nita G Valikodath; Leslie M Niziol; Lakshey Dudeja; Kyeong Hwan Kim; Maria A Woodward
Journal:  Cornea       Date:  2018-03       Impact factor: 2.651

10.  Comparison of Eyemetrics and Orbscan automated method to determine horizontal corneal diameter.

Authors:  Arvind Venkataraman; Sapna K Mardi; Sarita Pillai
Journal:  Indian J Ophthalmol       Date:  2010 May-Jun       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.