Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Paola De la Parra-Colín1, Manuel Garza-León, Tonatiuh Barrientos-Gutierrez.   

Abstract

To assess the repeatability and comparability of six anterior segment biometry parameters obtained with a novel Scheimpflug camera with a Placido disc topographer (Sirius) and slit-scanning tomography with a Scheimpflug camera (Pentacam), in a sample of 16 unoperated eyes of healthy subjects. The anterior segment was analyzed by a single examiner using the Sirius and the Pentacam analyzers. Mean simulated keratometry (Sim K), flat and steep axis keratometry (K f and K s ), central and thinnest corneal thicknesses (CCT and TCT), and anterior chamber depth (ACD) measurements were evaluated. Repeatability of three sets of measurements from each device was assessed using the coefficient of variation (CV), within-subject standard deviation, and intraclass correlation coefficient (ICC). Bonferroni-adjusted t-tests, and Bland and Altman plots were used to establish agreement between devices. For both devices the CV of repeated measurements was <0.79 %. The ICC was >0.95 in all measurements except for Sirius K s (ICC = 0.869). For all parameters evaluated, the Pentacam systematically yielded higher values, although differences were statistically significant in only three parameters-0.31 diopters for K s , 10.1 μm for CCT and 12.4 μm for TCT. In the assessment of normal corneas both devices showed overall high repeatability. Although good agreement was found in three parameters (Sim K, K f and ACD) these devices do not seem to be interchangeable for pachymetric determination.

Mesh:

Year:  2013        PMID: 23616225     DOI: 10.1007/s10792-013-9780-0

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  17 in total

1.  A corneal model for slit-scanning elevation topography.

Authors:  G Cairns; A Collins; C N J McGhee
Journal:  Ophthalmic Physiol Opt       Date:  2003-05       Impact factor: 3.117

2.  Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound.

Authors:  Birgit Lackner; Gerald Schmidinger; Stefan Pieh; Martin A Funovics; Christian Skorpik
Journal:  Optom Vis Sci       Date:  2005-10       Impact factor: 1.973

3.  Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan.

Authors:  Birgit Lackner; Gerald Schmidinger; Christian Skorpik
Journal:  Optom Vis Sci       Date:  2005-09       Impact factor: 1.973

4.  Anterior chamber depth measurements in phakic and pseudophakic eyes: Pentacam versus ultrasound device.

Authors:  Gabor Nemeth; Attila Vajas; Bence Kolozsvari; Andras Berta; Laszlo Modis
Journal:  J Cataract Refract Surg       Date:  2006-08       Impact factor: 3.351

5.  Pentacam pachometry: comparison with non-contact specular microscopy on the central cornea and inter-session repeatability on the peripheral cornea.

Authors:  Andrew K C Lam; Davie Chen
Journal:  Clin Exp Optom       Date:  2007-03       Impact factor: 2.742

6.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

7.  Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements.

Authors:  Hema Shankar; Deepa Taranath; Chandramalar T Santhirathelagan; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2008-01       Impact factor: 3.351

8.  Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes.

Authors:  Giacomo Savini; Michele Carbonelli; Piero Barboni; Kenneth J Hoffer
Journal:  J Cataract Refract Surg       Date:  2011-02       Impact factor: 3.351

9.  Diurnal variation of corneal and corneal epithelial thickness measured using optical coherence tomography.

Authors:  Y Feng; J Varikooty; T L Simpson
Journal:  Cornea       Date:  2001-07       Impact factor: 2.651

10.  Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry.

Authors:  Ugo de Sanctis; Alessandro Missolungi; Bernardo Mutani; Lorenzo Richiardi; Federico M Grignolo
Journal:  Am J Ophthalmol       Date:  2007-09-17       Impact factor: 5.258

View more
  22 in total

1.  Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens.

Authors:  Manuel Garza-Leon; Hugo A Fuentes-de la Fuente; Ana V García-Treviño
Journal:  Int Ophthalmol       Date:  2016-10-21       Impact factor: 2.031

2.  Corneal topographic alterations after selective laser trabeculoplasty.

Authors:  Kursat Atalay; Ahmet Kirgiz; Kubra Serefoglu Cabuk; Havva Erdogan Kaldirim
Journal:  Int Ophthalmol       Date:  2016-09-15       Impact factor: 2.031

3.  Topographic indices and pachymetry in healthy adolescents obtained with Sirius topographer.

Authors:  Emilia Cantera; Magdalena Cortes; Roberto Sacco; Gabriele Vestri; Alessandra Micera
Journal:  Int Ophthalmol       Date:  2017-10-30       Impact factor: 2.031

4.  Accuracy of optimized Sirius ray-tracing method in intraocular lens power calculation.

Authors:  Li-Qing Wei; Ying-Hui Fu; Wei-Hua Pan; Li Nie; Yu Chen; Gui-Fang Liu; Zhen-Bin Qian
Journal:  Int J Ophthalmol       Date:  2022-02-18       Impact factor: 1.779

5.  Comparison of two different scheimpflug devices in the detection of keratoconus, regular astigmatism, and healthy corneas.

Authors:  David Finis; Bernhard Ralla; Maria Karbe; Maria Borrelli; Stefan Schrader; Gerd Geerling
Journal:  J Ophthalmol       Date:  2015-04-14       Impact factor: 1.909

6.  Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes.

Authors:  Ayong Yu; Weiqi Zhao; Giacomo Savini; Zixu Huang; Fangjun Bao; Weicong Lu; Qinmei Wang; Jinhai Huang
Journal:  J Ophthalmol       Date:  2015-12-01       Impact factor: 1.909

7.  Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts.

Authors:  Jinhai Huang; Na Liao; Giacomo Savini; Fangjun Bao; Ye Yu; Weicong Lu; Qingjie Hu; Qinmei Wang
Journal:  J Ophthalmol       Date:  2014-10-23       Impact factor: 1.909

8.  Color light-emitting diode reflection topography: validation of keratometric repeatability in a large sample of wide cylindrical-range corneas.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2015-02-05

9.  Outcomes of photorefractive keratectomy in patients with posterior corneal steepening.

Authors:  Yumna Busool Abu-Eta; Achia Nemet; Igor Vainer; Michael Mimouni; Tzahi Sela; Gur Munzer; Igor Kaiserman
Journal:  Eye (Lond)       Date:  2020-10-06       Impact factor: 4.456

10.  Distribution and Repeatability of Corneal Astigmatism Measurements (Magnitude and Axis) Evaluated With Color Light Emitting Diode Reflection Topography.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Cornea       Date:  2015-08       Impact factor: 2.651

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.