Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery.

Oliver Findl1, Katharina Kriechbaum, Stefan Sacu, Barbara Kiss, Kaija Polak, Johannes Nepp, Gebtraud Schild, Georg Rainer, Saskia Maca, Vanessa Petternel, Birgit Lackner, Wolfgang Drexler.   

Abstract

PURPOSE: To compare measurements performed with the IOLMaster (Carl Zeiss, Meditec AG) with those obtained by applanation ultrasound (US) and manual keratometry and to evaluate the effect of operator experience on US biometry.
SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria.
METHODS: The axial length (696 eyes) and anterior chamber depth (ACD) (462 eyes) were measured in 377 patients with cataract using the IOLMaster and applanation US. To assess the effect of operator experience on the biometric results, the operators were divided into 2 groups: experienced and less experienced in performing US biometry. The difference in measurements between the methods and the variability of the difference were compared between the 2 groups.
RESULTS: Applanation US measured axial length and ACD shorter than the IOLMaster; the mean numerical difference was 0.13 mm and 0.19 mm, respectively (P<.01). For axial length, the absolute difference was smaller with experienced operators than with less experienced operators (0.15 mm versus 0.22 mm) (P<.01). For ACD, experienced operators obtained a smaller difference between measurement techniques (0.21 mm versus 0.29 mm; P<.05).
CONCLUSIONS: Experienced US operators had less difference and lower variability in the difference between applanation US and IOLMaster readings for axial length and ACD measurements. The noncontact optical method, which is essentially operator independent, gave significantly more reliable biometry before cataract surgery, especially in the case of less experienced operators.

Entities:  

Mesh:

Year:  2003        PMID: 14604716     DOI: 10.1016/s0886-3350(03)00243-8

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  18 in total

1.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

2.  Refractive outcomes comparison between the Lenstar LS 900® optical biometry and immersion A-scan ultrasound.

Authors:  Palanyraj Naicker; Siva Sundralingam; Mohammadreza Peyman; Azida Juana; Nor Fadhilah Mohamad; Maung Maung Win; Angela Loo; Visvaraja Subrayan
Journal:  Int Ophthalmol       Date:  2014-07-15       Impact factor: 2.031

3.  Refractive results of phacoemulsification in vitrectomized patients.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int Ophthalmol       Date:  2016-08-25       Impact factor: 2.031

4.  Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.

Authors:  Shi-Ming Cheng; Jia-Sheng Zhang; Xu Shao; Ze-Tong Wu; Tian-Tian Li; Po Wang; Jun-Hai Lin; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-02-16       Impact factor: 3.117

5.  Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry.

Authors:  Jeevan S Ladi; Nitant A Shah
Journal:  Indian J Ophthalmol       Date:  2010 Sep-Oct       Impact factor: 1.848

6.  Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int J Ophthalmol       Date:  2013-04-18       Impact factor: 1.779

7.  Outcomes of cataract operations in extreme high axial myopia.

Authors:  Jasmine K M Lam; Tommy C Y Chan; Alex L K Ng; Vanissa W S Chow; Victoria W Y Wong; Vishal Jhanji
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-06-16       Impact factor: 3.117

8.  Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts.

Authors:  Jinhai Huang; Na Liao; Giacomo Savini; Fangjun Bao; Ye Yu; Weicong Lu; Qingjie Hu; Qinmei Wang
Journal:  J Ophthalmol       Date:  2014-10-23       Impact factor: 1.909

9.  Agreement and relationship between ultrasonic and partial coherence interferometry measurements of axial length and anterior chamber depth.

Authors:  Amany R Wissa; Sherein S Wahba; Maged M Roshdy
Journal:  Clin Ophthalmol       Date:  2012-01-31

10.  Refractive predictability of partial coherence interferometry and factors that can affect it.

Authors:  Seung Mo Kim; Joohyun Choi; Sangkyung Choi
Journal:  Korean J Ophthalmol       Date:  2009-03-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.