Comparison of corneal powers obtained from 4 different devices.

Mariko Shirayama1, Li Wang, Mitchell P Weikert, Douglas D Koch.   

Abstract

PURPOSE: To assess the repeatability and comparability of anterior corneal power values obtained from the Galilei Dual Scheimpflug Analyzer (Ziemer, Port, Switzerland), Humphrey Atlas corneal topographer (Carl Zeiss, Jena, Germany), IOLMaster (Carl Zeiss), and a manual keratometer (Bausch & Lomb Inc, Rochester, New York, USA).
DESIGN: Prospective, comparative study.
METHODS: Prospectively, 20 subjects were enrolled. Three sets of corneal power measurements were obtained by a single observer using the Galilei, Atlas topographer, IOLMaster, and manual keratometer. Repeatability of the 3 measurements from each device was evaluated by means of coefficient of variation, standard deviation (SD), and intraclass correlation coefficient. An analysis of variance was used to compare the differences in corneal powers among devices. The Bland and Altman method also was performed to assess agreement in measurements between devices. Vector analysis was used to compare the astigmatism values obtained from different devices.
RESULTS: For each device, the coefficient of variation of repeated measurements was lower than 0.22%. The SD of 3 repeated measurements ranged from 0.042 to 0.096 diopters (D). The intraclass correlation coefficients were higher than 0.99 in all devices. Mean central corneal powers were 43.80 D, 43.88 D, 43.92 D, and 43.76 D for the Galilei, Atlas, IOLMaster, and manual keratometer, respectively. SDs of the differences between devices ranged from 0.07 D for Galilei and IOLMaster to 0.14 D for Galilei and Atlas. For astigmatism, the mean astigmatism values for the Galilei, Atlas, IOLMaster, and manual keratometer were 0.54 D at 84 degrees, 0.51 D at 88 degrees, 0.62 D at 88 degrees, and 0.52 D at 87 degrees, respectively.
CONCLUSIONS: The corneal power measurements from these 4 devices were highly reproducible, comparable, and correlated.

Entities:  

Mesh:

Year:  2009        PMID: 19541287     DOI: 10.1016/j.ajo.2009.04.028

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  31 in total

1.  Predictability of Sirius dual-scanning corneal tomography in the measurement of corneal power after photorefractive surgery.

Authors:  Sameh M Fouda; Haitham Y Al-Nashar; Basem M Ibrahim; Ashraf Bor'i
Journal:  Int Ophthalmol       Date:  2015-05-19       Impact factor: 2.031

2.  Keratometry with five different techniques: a study of device repeatability and inter-device agreement.

Authors:  Shiva Mehravaran; Soheila Asgari; Sara Bigdeli; Ashkan Shahnazi; Hassan Hashemi
Journal:  Int Ophthalmol       Date:  2014-02-23       Impact factor: 2.031

3.  Precision of 5 different keratometry devices.

Authors:  Jonas Vejvad Nørskov Laursen; Peter Jeppesen; Thomas Olsen
Journal:  Int Ophthalmol       Date:  2015-03-29       Impact factor: 2.031

4.  Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer.

Authors:  Leyla Asena; Sirel Gür Güngör; Ahmet Akman
Journal:  Int Ophthalmol       Date:  2016-06-07       Impact factor: 2.031

5.  Corneal power measurement with Fourier-domain optical coherence tomography.

Authors:  Maolong Tang; Alex Chen; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

6.  Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam.

Authors:  Xiaochun Li; Xiaoguang Cao; Yongzhen Bao
Journal:  Biomed Res Int       Date:  2022-07-08       Impact factor: 3.246

7.  Optical coherence tomography accurately measures corneal power change from laser refractive surgery.

Authors:  Ryan P McNabb; Sina Farsiu; Sandra S Stinnett; Joseph A Izatt; Anthony N Kuo
Journal:  Ophthalmology       Date:  2014-12-06       Impact factor: 12.079

8.  Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging.

Authors:  Giacomo Savini; Leonardo Taroni; Domenico Schiano-Lomoriello; Kenneth J Hoffer
Journal:  Eye (Lond)       Date:  2020-11-02       Impact factor: 4.456

9.  Comparison of the Corneal Power Measurements with the TMS4-Topographer, Pentacam HR, IOL Master, and Javal Keratometer.

Authors:  Zahra Dehnavi; Mehdi Khabazkhoob; Ali Mirzajani; Mahmood Jabbarvand; Abbasali Yekta; Ebrahim Jafarzadehpur
Journal:  Middle East Afr J Ophthalmol       Date:  2015 Apr-Jun

10.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240