Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer.

Leyla Asena1, Sirel Gür Güngör2, Ahmet Akman2.   

Abstract

The aim of this study was to compare the keratometric measurements of Verion Image Guided System with an optical biometer (Zeiss IOLMaster 500, Carl Zeiss Meditec, Jena, Germany) and an automated keratorefractometer (AKR) (Topcon KR-8900, Topcon, Japan). In this prospective clinical trial, the right eyes of 52 patients with cataract were examined (mean age 62.25 ± 12.16 years). The measurements were taken by the three systems in a random order. Keratometric data, magnitude of astigmatism, and astigmatic axis measurements from all three instruments were compared. The results were evaluated using, intraclass correlation coefficients (ICC), Bland-Altman plots, and paired samples t tests. The mean flat/steep K of Verion, IOLMaster, and AKR were 43.22 ± 1.38D/44.23 ± 1.46D, 43.07 ± 1.26D/44.05 ± 1.34D, and 43.07 ± 1.31D/43.89 ± 1.42D, respectively. Flat K readings of Verion were higher than IOLMaster and AKR (p < 0.05 for both). Steep K readings were different for all three (p < 0.05). The magnitude of astigmatism by Verion and IOLMaster were 0.98 ± 0.65D and 0.98 ± 0.59D (p = 0.88). The mean astigmatism measured by the AKR was 0.82 ± 0.62D, less than the other two instruments (p < 0.001). Astigmatic axis measurements of Verion and AKR differed <10° in 38, between 10° and 20° in 5, and >20° in 9 eyes; the same difference was 30, 11, and 11 eyes, respectively, between Verion and IOLMaster. Although, keratometric and astigmatic results obtained from Verion were not completely interchangeable with IOLMaster and AKR, especially the agreement between Verion and IOLMaster was excellent with ICCs close to one. However, there were pronounced astigmatic axis measurement differences between three instruments.

Entities:  

Keywords:  Astigmatism; Biometry; Keratometry; Verion Image Guided System

Mesh:

Year:  2016        PMID: 27271763     DOI: 10.1007/s10792-016-0274-8

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  17 in total

1.  Simple method for accurate alignment in toric phakic and aphakic intraocular lens implantation.

Authors:  Joseph J K Ma; Santos S Tseng
Journal:  J Cataract Refract Surg       Date:  2008-10       Impact factor: 3.351

2.  Intraocular lens power calculation: clinical comparison of 2 optical biometry devices.

Authors:  Tanja M Rabsilber; Charlotte Jepsen; Gerd U Auffarth; Mike P Holzer
Journal:  J Cataract Refract Surg       Date:  2010-02       Impact factor: 3.351

3.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

4.  Repeatability data and agreement of keratometry with the VERION system compared to the IOLMaster.

Authors:  Gabor Nemeth; Eszter Szalai; Ziad Hassan; Agnes Lipecz; Andras Berta; Laszlo Modis
Journal:  J Refract Surg       Date:  2015-05       Impact factor: 3.573

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Agreement between a partial coherence interferometer and 2 manual keratometers.

Authors:  Mark A Bullimore; Tobias Buehren; Wilfried Bissmann
Journal:  J Cataract Refract Surg       Date:  2013-07-19       Impact factor: 3.351

7.  Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation.

Authors:  János Németh; Orsolya Fekete; Norbert Pesztenlehrer
Journal:  J Cataract Refract Surg       Date:  2003-01       Impact factor: 3.351

8.  Comparison and evaluation of ocular biometry using a new noncontact optical low-coherence reflectometer.

Authors:  Kaspar Rohrer; Beatrice E Frueh; Rudolf Wälti; Isabelle A Clemetson; Christoph Tappeiner; David Goldblum
Journal:  Ophthalmology       Date:  2009-09-10       Impact factor: 12.079

9.  Comparison of anterior segment measurements obtained by three different devices in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-06-02       Impact factor: 3.411

10.  Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts.

Authors:  Jinhai Huang; Na Liao; Giacomo Savini; Fangjun Bao; Ye Yu; Weicong Lu; Qingjie Hu; Qinmei Wang
Journal:  J Ophthalmol       Date:  2014-10-23       Impact factor: 1.909

View more
  4 in total

1.  Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer.

Authors:  Cecilio Velasco-Barona; Guadalupe Cervantes-Coste; Erick Mendoza-Schuster; Claudia Corredor-Ortega; Nadia L Casillas-Chavarín; Alejandro Silva-Moreno; Manuel Garza-León; Roberto Gonzalez-Salinas
Journal:  Int Ophthalmol       Date:  2017-04-25       Impact factor: 2.031

2.  Comparison of corneal power obtained from VERION image-guided surgery system and four other devices.

Authors:  Hung-Yuan Lin; Hsin-Yang Chen; Han Bor Fam; Ya-Jung Chuang; Ronald Yeoh; Pi-Jung Lin
Journal:  Clin Ophthalmol       Date:  2017-07-12

3.  Pseudophakic astigmatism reduction with femtosecond laser-assisted corneal arcuate incisions: a pilot study.

Authors:  Clayton Blehm; Richard Potvin
Journal:  Clin Ophthalmol       Date:  2017-01-23

4.  Clinical Outcomes After Femtosecond Laser-Assisted Arcuate Corneal Incisions versus Manual Incisions.

Authors:  Clayton Blehm; Richard Potvin
Journal:  Clin Ophthalmol       Date:  2021-06-23
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.