Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes.

Canan Asli Utine1, Feyza Altin, Hanefi Cakir, Irfan Perente.   

Abstract

PURPOSE: This study determined to assess the degree of agreement between anterior chamber depth (ACD) measurements obtained using three different devices and to analyse the relationship between ACD and spherical equivalent (SE) refraction.
METHODS: In this cross-sectional study, 42 eyes of 42 patients with a mean SE of - 4.69 +/- 4.61 D (range 0.00 D to - 14.88 D) were analysed. Measurements of ACD between the corneal epithelium and the anterior surface of the crystalline lens, obtained using the Pentacam, Orbscan IIz and IOLMaster, were compared. The relationships between SE and ACD measurements obtained with different devices were also investigated. The results were analysed using Bland-Altman analyses, single-sample t-test and Pearson's correlation test.
RESULTS: Orbscan ACD measurements were an average of 0.05 mm less than Pentacam measurements (p = 0.01). IOLMaster measurements were an average of 0.06 mm less than Orbscan measurements (p < 0.001). None of the ACD values measured by any of the devices were correlated with increasing SE (p > 0.05 for all). There was a weak positive correlation between SE and the difference in ACD measurements with Pentacam and Orbscan (p = 0.04); however, the differences between Pentacam and IOLMaster ACD measurements and Orbscan and IOLMaster ACD measurements seemed to be independent of SE (p = 0.17 and p = 0.54, respectively).
CONCLUSIONS: The ACD in clinically normal eyes is measured differently by various non-ultrasonic devices. However, the observed mean error between these modalities is too small to create any noticeable difference in refractive outcome. No significant relationship was found between SE and ACD measurements obtained by Pentacam, Orbscan or IOLMaster.

Entities:  

Mesh:

Year:  2008        PMID: 18778337     DOI: 10.1111/j.1755-3768.2008.01278.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  16 in total

1.  Comparison of the measurements of a novel optical biometry: Nidek AL-Scan with Sirius and a ultrasound biometry.

Authors:  Çağatay Çağlar; Sücattin İlker Kocamış; Emre Demir; Mustafa Durmuş
Journal:  Int Ophthalmol       Date:  2016-07-08       Impact factor: 2.031

2.  Anterior chamber depth in normal subjects by rotating scheimpflug imaging.

Authors:  Matthew T Feng; Michael W Belin; Renato Ambrósio; Satinder P S Grewal; Wang Yan; Mohamed S Shaheen; Charles McGhee; Naoyuki Maeda; Tobias H Neuhann; H Burkhard Dick; Saleh A Alageel; Andreas Steinmueller
Journal:  Saudi J Ophthalmol       Date:  2011-04-30

3.  Ocular biometry through fully refocused steady-state magnetic resonance imaging sequence: reliability and agreement with the IOLMaster® 500.

Authors:  Lorenzo Ismael Perez-Sanchez; Julia Gutierrez-Vazquez; Maria Satrustegui-Lapetra; Francisco Ferreira-Manuel; Juan Jose Arevalo-Manso; Juan Jesus Gomez-Herrera; Juan Jose Criado-Alvarez
Journal:  Int Ophthalmol       Date:  2021-02-23       Impact factor: 2.031

Review 4.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

5.  Comparison of techniques for measuring anterior chamber depth: Orbscan imaging, Smith's technique, and van Herick's method.

Authors:  Frank Eperjesi; Claire Holden
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-10       Impact factor: 3.117

6.  Comparison of 25 MHz and 50 MHz ultrasound biomicroscopy for imaging of the lens and its related diseases.

Authors:  Ming-Yu Shi; Xiao Han; Jin-Song Zhang; Qi-Chang Yan
Journal:  Int J Ophthalmol       Date:  2018-07-18       Impact factor: 1.779

7.  Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts.

Authors:  Jinhai Huang; Na Liao; Giacomo Savini; Fangjun Bao; Ye Yu; Weicong Lu; Qingjie Hu; Qinmei Wang
Journal:  J Ophthalmol       Date:  2014-10-23       Impact factor: 1.909

8.  Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes.

Authors:  Jing Dong; Maolong Tang; Yaqin Zhang; Yading Jia; Haining Zhang; Zhijie Jia; Xiaogang Wang
Journal:  PLoS One       Date:  2015-11-17       Impact factor: 3.240

9.  Biometry and visual function of a healthy cohort in Leipzig, Germany.

Authors:  Maria Teresa Zocher; Jos J Rozema; Nicole Oertel; Jens Dawczynski; Peter Wiedemann; Franziska G Rauscher
Journal:  BMC Ophthalmol       Date:  2016-06-07       Impact factor: 2.209

10.  Biomechanical Measurement of Rabbit Cornea by a Modified Scheimpflug Device.

Authors:  Bo Zhang; Jianjun Gu; Xiaoxiao Zhang; Bin Yang; Zheng Wang; Danying Zheng
Journal:  J Ophthalmol       Date:  2016-07-03       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.