PURPOSE: To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA).Entities:
Mesh:
Year: 2014 PMID: 24854348 PMCID: PMC4031212 DOI: 10.1371/journal.pone.0098316
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Central corneal thickness, thinnest corneal thickness and midperipheral (2 and 5 mm diameter) corneal thickness in the superior, inferior, nasal, and temporal quadrants with the four investigated devices.
| Parameter | Mean (µm) ± Standard Deviation |
| |||
| Pentacam | Sirius | Galilei | RTVue | ||
| Center | 538.8±26.5 | 542.1±27.1 | 548.1±26.4 | 532.8±26.2 | <0.001 |
| Thinnest | 535.8±26.4 | 539.2±27.1 | 545.6±26.4 | 528.1±26.5 | <0.001 |
| Superior 2 mm | 555.4±27.3 | 559.3±27.8 | 563.1±27.0 | 542.4±26.6 | <0.001 |
| Inferior 2 mm | 542.5±26.8 | 544.0±27.2 | 551.3±26.4 | 533.4±26.8 | <0.001 |
| Nasal 2 mm | 549.4±26.8 | 552.8±27.4 | 558.7±26.5 | 540.0±26.8 | <0.001 |
| Temporal 2 mm | 540.1±26.7 | 545.4±27.5 | 550.7±27.1 | 533.2±26.3 | <0.001 |
| Superior 5 mm | 626.2±30.9 | 625.8±30.3 | 619.1±30.0 | 591.7±29.2 | <0.001 |
| Inferior 5 mm | 592.1±28.3 | 587.0±29.0 | 588.0±27.1 | 562.4±28.5 | <0.001 |
| Nasal 5 mm | 605.5±28.4 | 600.9±28.6 | 602.9±27.5 | 575.5±28.1 | <0.001 |
| Temporal 5 mm | 585.4±28.3 | 582.4±29.2 | 580.0±28.9 | 554.3±27.6 | <0.001 |
Mean difference of central corneal thickness, corresponding results of Bonferroni post hoc comparison and 95% limits of agreement (LoA) among the 4 investigated devices.
| Device Pairings | Mean Difference (µm) ± SD |
| 95% LoA (µm) |
| Pentacam - Sirius | −3.3±5.2 | <0.001 | −13.6 to 6.9 |
| Pentacam - Galilei | −9.3±3.7 | <0.001 | −16.6 to −2.0 |
| Pentacam - RTVue | 6.0±4.8 | <0.001 | −3.4 to 15.4 |
| Sirius - Galilei | −6.0±4.0 | <0.001 | −13.8 to 1.9 |
| Sirius - RTVue | 9.3±5.6 | <0.001 | −1.6 to 20.3 |
| Galilei - RTVue | 15.3±4.1 | <0.001 | 7.3 to 23.3 |
SD = Standard deviation.
Mean difference of superior 2% limits of agreement (LoA) among the 4 investigated devices.
| Device Pairings | Mean Difference (µm) ± SD |
| 95% LoA (µm) |
| Pentacam - Sirius | −4.0±5.5 | <0.001 | −14.7 to 6.8 |
| Pentacam - Galilei | −7.8±4.5 | <0.001 | −16.7 to 1.1 |
| Pentacam - RTVue | 12.9±5.0 | <0.001 | 3.2 to 22.6 |
| Sirius - Galilei | −3.8±4.7 | <0.001 | −13.0 to 5.4 |
| Sirius - RTVue | 16.9±5.6 | <0.001 | 5.9 to 28.0 |
| Galilei - RTVue | 20.7±4.8 | <0.001 | 11.3 to 30.1 |
SD = Standard deviation.
Mean difference of superior 5% limits of agreement (LoA) among the 4 investigated devices.
| Device Pairings | Mean Difference (µm) ± SD |
| 95% LoA (µm) |
| Pentacam - Sirius | 0.4±6.1 | 1.000 | −11.6 to 12.3 |
| Pentacam - Galilei | 7.1±7.1 | <0.001 | −6.9 to 21.0 |
| Pentacam - RTVue | 34.4±10.6 | <0.001 | 13.6 to 55.3 |
| Sirius - Galilei | 6.7±7.1 | <0.001 | −7.2 to 20.7 |
| Sirius - RTVue | 34.1±9.7 | <0.001 | 15.0 to 53.1 |
| Galilei - RTVue | 27.4±10.6 | <0.001 | 6.5 to 48.2 |
SD = Standard deviation.
Mean difference of thinnest corneal thickness, corresponding results of Bonferroni post hoc comparison and 95% limits of agreement (LoA) among the 4 investigated devices.
| Device Pairings | Mean Difference (µm) ± SD |
| 95% LoA (µm) |
| Pentacam - Sirius | −3.4±5.2 | <0.001 | −13.5 to 6.7 |
| Pentacam - Galilei | −9.8±3.7 | <0.001 | −17.0 to −2.7 |
| Pentacam - RTVue | 7.7±5.2 | <0.001 | −2.5 to 18.0 |
| Sirius - Galilei | −6.4±3.8 | <0.001 | −13.9 to 1.1 |
| Sirius - RTVue | 11.1±6.0 | <0.001 | −0.7 to 22.9 |
| Galilei - RTVue | 17.6±4.1 | <0.001 | 9.6 to 25.5 |