Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth.

Andrea Bedei1, Ilaria Appolloni, Antonio Madesani, Alessia Pietrelli, Sara Franceschi, Lucio Barabesi.   

Abstract

PURPOSE: To evaluate the repeatability and agreement of 2 Scheimpflug tomographies (Pentacam HR, Oculus GmbH; Sirius, CSO Inc.) in measuring the central corneal thickness (CCT) and depth (ACD), volume (ACV), and angle (ACA) of the anterior chamber.
METHODS: Three consecutive measurements were performed in one eye of each subject, with both instruments. The following were evaluated: CCT, ACD, ACV, and ACA. Repeatability was assessed by the Friedman test and agreement was assessed by 4 different Wilcoxon tests for paired data.
RESULTS: The study enrolled 30 patients. With Sirius tomography the Friedman test p values for testing homogeneity across measurements was with respect to CCT 0.320, ACD 0.315, ACV 0.171, and ACA 0.020; and with Pentacam CCT 0.191, ACD 0.010, ACV 0.704, and ACA 0.150. The Wilcoxon test p values for testing the measurement homogeneity between the corneal tomography Sirius and Pentacam showed the following: ACD 0.000, ACD 0.000, ACV 0.853, and ACA 0.000.
CONCLUSIONS: The measurements may be considered substantially stable for each corneal tomography and each variable, but the 2 corneal tomographies do not produce statistically equivalent measurements.

Entities:  

Mesh:

Year:  2012        PMID: 22267453     DOI: 10.5301/ejo.5000102

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  3 in total

1.  Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes.

Authors:  Clara Ruiz-Belda; David P Piñero; Pedro Ruiz-Fortes; Roberto Soto-Negro; Myriam Moya; Rafael J Pérez-Cambrodí; Alberto Artola
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-07-15       Impact factor: 3.117

2.  Central and midperipheral corneal thickness measured with Scheimpflug imaging and optical coherence tomography.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2014-05-22       Impact factor: 3.240

Review 3.  Anterior Chamber Angle Assessment Techniques: A Review.

Authors:  Ivano Riva; Eleonora Micheletti; Francesco Oddone; Carlo Bruttini; Silvia Montescani; Giovanni De Angelis; Luigi Rovati; Robert N Weinreb; Luciano Quaranta
Journal:  J Clin Med       Date:  2020-11-25       Impact factor: 4.241

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.