Reliability and agreement of the central and mid-peripheral corneal thickness measured by a new Scheimpflug based imaging.

A-Yong Yu1,2, Junming Ye1, Giacomo Savini3, Yiran Wang1, Tianjiao Zhang1, Min Chen1, Qinmei Wang1,2, Jinhai Huang1,2.   

Abstract

BACKGROUND: To assess the intra-observer repeatability and inter-observer reproducibility of central corneal thickness (CCT) and mid-peripheral corneal thickness (MPCT) measurements using a new Scheimpflug imaging instrument (Scansys) and compare the agreement with the rotating Scheimpflug corneal tomographer (Pentacam HR).
METHODS: The same well-trained operator performed the measuring using the two devices, after which Scansys measurements were repeated by another operator. Both instruments required three consecutive measurements per subject. Corneal thickness measurements were obtained by each instrument, including CCT, thinnest corneal thickness (TCT), pupil corneal thickness (PCT), and MPCT. Test-retest repeatability (TRT), within-subject coefficient of variation (CoV), and intra-class correlation coefficient (ICC) were used to evaluate repeatability and reproducibility. A paired t-test was used to compare the differences between Scansys and Pentacam, and the agreement was compared with Bland-Altman plots.
RESULTS: This study enrolled 112 healthy subjects. The CoV were <0.91% and 0.55% for repeatability and reproducibility, respectively. The ICC was close to 1 in all measurements. For intra-observer repeatability in the CT2mm region, TRT was <10.30 µm. Moreover, TRT was <15.26 µm within the CT5mm region. The paired t-test showed significant differences in all corneal thickness measurements (P<0.001). The central region and CT2mm agreement were high, but the largest range of 95% limits of agreement (LoA) appeared in the CTnasal-5mm.
CONCLUSIONS: The new Scheimpflug imaging instrument showed excellent intra-observer repeatability and inter-observer reproducibility for corneal thickness measurements. The agreement analysis suggested that Scansys and Pentacam could be interchangeably used between the central region and CT2mm, except CT5mm. 2021 Annals of Translational Medicine. All rights reserved.

Entities:  

Keywords:  Scheimpflug; agreement; biometry; corneal thickness; reliability

Year:  2021        PMID: 34430577      PMCID: PMC8350684          DOI: 10.21037/atm-20-7895

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


Introduction

Corneal thickness has an important role in predicting glaucoma (1), the diagnosis of keratoconus (2), and the assessment of corneal edema (3). In refractive surgery, accurate preoperative measurement of corneal thickness, such as central corneal thickness (CCT) and mid-peripheral corneal thickness (MPCT) can be used to assess the feasibility of surgery, thereby reducing the incidence of postoperative ectasia (2,4). Yet, there is a significant difference in the thickness of the cornea from the center to the periphery when measured using different devices, such that measurements further away from the center result in the greater the difference (5). In addition, peripheral corneal thickness has to be obtained for the diagnosis of keratoconus and other corneal ectasia, in the postoperative follow-up of corneal edema, and in the planification of corneal relaxing incisions (6-8). Scheimpflug image analysis is a precise technique that is commonly applied for light scattering measurements and biometric measurements of the anterior segment (9). Repeatable data on the characteristics of the anterior segment have been provided in clinical and experimental studies. Several instruments based on this technique have been marketed, including Pentacam (Oculus, Wetzlar, Germany), Sirius (CSO, Florence, Italy), and Galilei (Ziemer, Port, Switzerland), while their precision, repeatability, and agreement have been well validated (10-12). Scansys (MediWorks, Shanghai, China) is a novel device based on Scheimpflug imaging that can be used for anterior segment diagnosis. However, so far, no studies reported on this Scheimpflug based imaging. Consequently, the aim of this study was to evaluate the intra-observer repeatability and inter-observer reproducibility of the central and mid-peripheral regional corneal thickness and compare the agreement of these measurements between Scansys and Pentacam to determine the equivalence and interchangeability in healthy subjects. We present the following article in accordance with the STARD reporting checklist (available at https://dx.doi.org/10.21037/atm-20-7895).

Methods

Subjects

This prospective study recruited healthy subjects from the Eye Hospital of Wenzhou Medical University. After a detailed explanation of the purpose of the study, informed consent was provided by all patients. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013) and was approved by the Office of Research Ethical Committee, Eye Hospital of Wenzhou Medical University (No. KYK2013–21). Informed consent was obtained from all individual participants. Exclusion criteria included active ocular disease or trauma, dry eye, previous ocular surgery, and the history of wearing contact lenses (for soft contact lenses less than two weeks, rigid contact lenses less than four weeks).

Instruments

The Scansys is a new corneal topography instrument that uses the Scheimpflug camera, which can obtain 107,520 data points and 28 slit images in high resolution (1,280×1,024) of the anterior and posterior surface of the cornea in just 1 second per measurement, and has a faster scan speed. This instrument uses a slit-light source with a wavelength of 470 nm. The horizontal measurement range is up to 14 mm, and the vertical measurement range is up to 10 mm. Meanwhile, it can track the inevitable micromovements of the eye and reduce motion error by correcting eye motion through a software algorithm. It provides anterior and posterior corneal topography, including cornea curvature maps, cornea thickness maps, cornea elevation maps, etc. The Pentacam HR (Oculus, Wetzlar, Germany) is one of the first commercially available rotating Scheimpflug cameras, which takes 25 slit images composed of 138,000 points per measurement in less than 2 seconds. It can generate a 3-dimensional image of the anterior segment and provide measurements of the corneal curvature, central and peripheral corneal thickness, and anterior chamber depth.

Procedures

The two devices were used to measure the right eye of all subjects in random order. To assess reliability and agreement, the measurements by the two devices were performed by the same well-trained operator. Then, Scansys measurements were repeated by another operator to investigate the inter-operator reproducibility. Both devices were used to perform three consecutive measurements from each subject. Before each measurement, the patient was asked to blink to smooth the tear film. To ensure the independence of each measurement, subjects were asked to sit back while the operator readjusted the joystick. All measurements were taken between 10 AM and 5 PM to minimize the diurnal change of corneal thickness. The operating procedure strictly followed the manufacturer's guidelines. If the “quality specification” was “OK”, the examination was included in the analysis; otherwise, the procedure was repeated. Corneal thickness measurements included CCT, thinnest corneal thickness (TCT), and pupil corneal thickness (PCT). At positions 1 mm and 2.5 mm away from the corneal apex, 8 points were collected in the nasal, superior, temporal, and inferior quadrants: these were defined as CT2mm (CTnasal-2mm, CTsuperior-2mm, CTtemporal-2mm, and CTinferior-2mm) and CT5mm (CTnasal-5mm, CTsuperior-5mm, CTtemporal-5mm, and CTinferior-5mm), respectively.

Statistical analysis

The data were analyzed by SPSS (version 21.0, IBM Corp.) and EXCEL (2010, Microsoft Corp.). Mean and standard deviation (SD) were used to indicate corneal thickness at different locations. Within-subject standard deviation (Sw), test-retest repeatability (TRT =2.77 Sw), within-subject coefficient of variation (CoV), and intra-class correlation coefficient (ICC) were used to evaluate repeatability and reproducibility. A paired t-test was used to compare the measurements by the two devices, and agreement between Scansys and Pentacam was evaluated with Bland-Altman plots and the 95% limits of agreement (LoA). These were defined as the average difference between the two devices ±1.96 SD (13). A P value <0.05 was considered statistically significant.

Results

A total of 112 right eyes from 112 subjects (29 female and 83 male) were included in the current study. The mean age ± SD was 26±5.58 (range, 18 to 43) years, and the equivalent spherical power ± SD was −5.78±2.39 (range, −1.25 to −11.75) diopters (D). shows the intra-observer repeatability of corneal thickness measurements for each observer with the new Scheimpflug camera. For all measurements, the ICC was close to 1, which indicated high intra-observer repeatability. The repeatability of central and CT2mm regions was higher than that of CT5mm. In CT2mm measurements, the TRT was lower than 10.3 µm, whereas, in CT5mm, it was less than 15.5 µm. The CoV within CT2mm was less than 0.7% and within CT5mm was less than 1%.
Table 1

Intra-observer repeatability of the new Scheimpflug imaging by each observer in measuring corneal thickness

ParameterObserverMean ± SD (μm)Sw (μm)TRT (μm)CoV (%)ICC (95% CI)
Center1st545.36±29.863.309.140.610.988 (0.983 to 0.991)
2nd544.78±29.953.238.940.590.988 (0.983 to 0.992)
Thinnest1st543.18±29.813.238.950.590.988 (0.984 to 0.992)
2nd542.52±29.953.389.370.620.987 (0.981 to 0.991)
Pupil1st545.48±29.833.329.200.610.988 (0.983 to 0.991)
2nd544.89±29.963.238.940.590.989 (0.983 to 0.992)
Nasal 2 mm1st550.45±29.893.399.390.620.987 (0.983 to 0.991)
2nd550.07±29.953.409.420.620.987 (0.982 to 0.991)
Superior 2 mm1st558.54±30.253.6510.110.650.986 (0.980 to 0.990)
2nd558.32±30.253.439.510.620.987 (0.981 to 0.991)
Temporal 2 mm1st549.85±29.933.7210.300.680.985 (0.979 to 0.989)
2nd549.21±29.943.389.360.620.987 (0.981 to 0.991)
Inferior 2 mm1st547.65±29.853.479.620.630.987 (0.982 to 0.990)
2nd546.98±29.853.529.760.640.986 (0.980 to 0.990)
Nasal 5 mm1st582.93±30.644.2911.890.740.981 (0.974 to 0.986)
2nd582.69±30.383.8610.700.660.984 (0.978 to 0.989)
Superior 5 mm1st608.64±31.115.5115.260.910.969 (0.958 to 0.978)
2nd608.23±31.564.8613.460.800.977 (0.968 to 0.983)
Temporal 5 mm1st580.64±31.045.0013.850.860.975 (0.965 to 0.982)
2nd580.10±30.924.6012.730.790.978 (0.970 to 0.985)
Inferior 5 mm1st584.54±30.804.4712.370.760.979 (0.972 to 0.985)
2nd583.88±30.444.4712.380.770.978 (0.970 to 0.984)

SD, standard deviation; Sw, within-subject standard deviation; TRT, test-retest repeatability (2.77 Sw); CoV, within-subject coefficient of variation; ICC, intraclass correlation coefficient; USP, ultrasound pachymetry.

SD, standard deviation; Sw, within-subject standard deviation; TRT, test-retest repeatability (2.77 Sw); CoV, within-subject coefficient of variation; ICC, intraclass correlation coefficient; USP, ultrasound pachymetry. shows the inter-observer reproducibility of corneal thickness measurements. TRT showed an increasing trend from the center to the periphery. Nevertheless, the reproducibility of all the positions was excellent, with only slight deviations in the CTsuperior-5mm and CTinferior-5mm (CTsuperior-5mm: TRT =9.09 µm, COV =0.54%; CTinferior-5mm: TRT =8.90 µm, COV =0.55%). The ICC values for all measurements exceeded 0.989.
Table 2

Inter-observer reproducibility of the new Scheimpflug imaging in measuring corneal thickness

ParameterSw (μm)TRT (μm)COV (%)ICC
Center2.105.820.390.995 (0.993 to 0.997)
Thinnest2.256.230.410.994 (0.992 to 0.996)
Pupil2.105.820.390.995 (0.993 to 0.997)
Nasal 2 mm2.045.660.370.995 (0.993 to 0.997)
Superior 2 mm2.296.350.410.994 (0.992 to 0.996)
Temporal 2 mm2.336.460.420.994 (0.991 to 0.996)
Inferior 2 mm2.466.810.450.993 (0.990 to 0.995)
Nasal 5 mm2.366.530.400.994 (0.991 to 0.996)
Superior 5 mm3.289.090.540.989 (0.984 to 0.992)
Temporal 5 mm2.637.270.450.993 (0.990 to 0.995)
Inferior 5 mm3.218.900.550.989 (0.984 to 0.992)

SD, standard deviation; Sw, within-subject standard deviation; TRT, test-retest reproducibility (2.77 Sw); COV, within-subject coefficient of variation; ICC, intraclass correlation coefficient.

SD, standard deviation; Sw, within-subject standard deviation; TRT, test-retest reproducibility (2.77 Sw); COV, within-subject coefficient of variation; ICC, intraclass correlation coefficient. indicates the differences and agreement between Scansys and Pentacam. The paired t-test of the two instruments showed statistically significant differences at all locations (P<0.001). In the central and CT2mm region, the value of Scansys was slightly larger than Pentacam in measuring corneal thickness, except for the CTnasal-2mm. In CT5mm, the value of Scansys was smaller than Pentacam.
Table 3

Difference and agreement between the new Scheimpflug imaging Scansys and Pentacam HR in measuring corneal thickness

Device pairingsMean ± SD (μm)P value95% LoA
Center3.36±5.11<0.001−6.65 to 13.37
Thinnest4.47±5.29<0.001−5.89 to 14.83
Pupil3.10±4.94<0.001−6.58 to 12.79
Nasal 2 mm−2.68±5.29<0.001−13.05 to 7.69
Superior 2 mm0.58±5.12<0.001−9.45 to 10.61
Temporal 2 mm5.78±4.90<0.001−3.81 to 15.38
Inferior 2 mm2.95±6.09<0.001−8.98 to 14.88
Nasal 5 mm−22.35±6.61<0.001−35.30 to −9.40
Superior 5 mm−17.83±7.71<0.001−32.94 to −2.72
Temporal 5 mm−7.44±5.83<0.001−18.87 to 3.99
Inferior 5 mm−7.56±7.56<0.001−22.38 to 7.26

SD, standard deviation.

SD, standard deviation. The agreement analysis of Bland-Altman plots is shown in . The central and CT2mm region had a smaller range than the CT5mm. Yet, the most significant difference appeared in the CTnasal-5mm.
Figure 1

Corneal thickness maps obtained with the new Scheimpflug imaging instrument.

Figure 2

Bland-Altman plots for central corneal thickness (CCT), thinnest corneal thickness (TCT), and pupil corneal thickness (PCT) between the new Scheimpflug imaging Scansys and Pentacam HR. The solid lines = the mean difference; the upper and lower lines = the 95% LoA; SD, standard deviation.

Figure 3

Bland-Altman plots for corneal thickness2mm-Nasal (CTnasal-2mm), thickness2mm-Superior (CTsuperior-2mm), thickness2mm-Temporal (CTtemporal-2mm), and thickness2mm-Inferior (CTinferior-2mm) between the new Scheimpflug imaging Scansys and Pentacam HR. The solid lines = the mean difference; the upper and lower lines = the 95% LoA; SD, standard deviation.

Figure 4

Bland-Altman plots for corneal thickness5mm-Nasal (CTnasal-5mm), thickness5mm-Superior (CTsuperior-5mm), thickness5mm-Temporal (CTtemporal-5mm), and thickness5mm-Inferior (CTinferior-5mm) between the new Scheimpflug imaging Scansys and Pentacam HR. The solid lines = the mean difference; the upper and lower lines = the 95% LoA; SD, standard deviation.

Corneal thickness maps obtained with the new Scheimpflug imaging instrument. Bland-Altman plots for central corneal thickness (CCT), thinnest corneal thickness (TCT), and pupil corneal thickness (PCT) between the new Scheimpflug imaging Scansys and Pentacam HR. The solid lines = the mean difference; the upper and lower lines = the 95% LoA; SD, standard deviation. Bland-Altman plots for corneal thickness2mm-Nasal (CTnasal-2mm), thickness2mm-Superior (CTsuperior-2mm), thickness2mm-Temporal (CTtemporal-2mm), and thickness2mm-Inferior (CTinferior-2mm) between the new Scheimpflug imaging Scansys and Pentacam HR. The solid lines = the mean difference; the upper and lower lines = the 95% LoA; SD, standard deviation. Bland-Altman plots for corneal thickness5mm-Nasal (CTnasal-5mm), thickness5mm-Superior (CTsuperior-5mm), thickness5mm-Temporal (CTtemporal-5mm), and thickness5mm-Inferior (CTinferior-5mm) between the new Scheimpflug imaging Scansys and Pentacam HR. The solid lines = the mean difference; the upper and lower lines = the 95% LoA; SD, standard deviation.

Discussion

Scheimpflug technology is widely used in the pre-operative and post-operative settings as a reliable method for imaging and measuring the anterior ocular segment. The Scansys is the latest model based on Scheimpflug technology; however, before its clinical application can be approved, it is necessary to evaluate whether it provides repeatable and reproducible measurements and whether its measurements are consistent with those given by previously available instruments based on a similar principle. To the best of our knowledge, this is the first study that evaluated the intra-observer repeatability and inter-observer reproducibility of corneal central and mid-peripheral thickness measurements using the new Scheimpflug imaging Scansys. Since the CoV did not exceed 1% and the ICC was close to 1 in all quadrants, we concluded that the measurements provided by this new instrument have excellent repeatability and reproducibility. The values of corneal thickness obtained in the current study were similar to other studies on the use of Scheimpflug technology on unoperated eyes (14-16). In the four quadrants, the superior measurements had poorer repeatability, which is similar to previous studies (16,17). Meanwhile, other studies have found that the nasal quadrant is better than the temporal quadrant (18), which was consistent with our results. The current study found excellent precision of central corneal thickness measurements, which was greatly improved compared with that of Pentacam rotating Scheimpflug camera (CoV =0.6–1.7%) (15,18) and was similar to that of Sirius Scheimpflug camera-Placido topographer (Sw =2.78–3.17 µm) (16,19). Besides, Kumar et al. (20) reported higher Sw (7.15 µm) and TRT (19.81 µm) in TCT measurements with Orbscan II (Bausch & Lomb, Rochester, NY) compared to our results. However, according to Huang et al. (19), the repeatability for central region with Scansys was lower than Galilei (CoV =0.24–0.33%, TRT =3.64–5.05 µm) and RTVue (Optovue, Freemont, CA, CoV =0.31–0.41%, TRT =4.57–6.01 µm). Repeatability was reduced when moving measurements from 2 to 5 mm. The reliability of the MPCT was inferior to that of CCT (17,21), which was consistent with our findings. This may be due to the rotation of the camera around the optical axis, as the overlap of the Scheimpflug images and the number of analyzed points at the center of the cornea are greater than those at the corneal periphery, so the number of points captured at the same location decreased from the center of the cornea to the periphery (17). In a previous study, Huang et al. compared 3 rotating Scheimpflug cameras (Pentacam, Sirius, Galilei) and 1 Fourier-domain optical coherence tomography system (RTVue) (19). In the 5 mm region, the intra-observer TRT ranged from 14.08 to 19.01 µm with Pentacam, from 9.47 to 18.21 µm with Sirius, from 11.14 to 19.37 µm with Galilei, and from 11.63 to 19.01 µm with RTVue, respectively. Moreover, Lu et al. evaluated the mid-peripheral region with RTVue and reported the TRT of 11.77–17.23 µm (22). In the repeatability of Orbscan II assessed by Martin and his team, the CoV of MPCT was 1.11–1.67% (23). In general, the measurements obtained by the new Scheimpflug instrument in the mid-peripheral region had slightly higher repeatability and advantage compared with several other instruments mentioned above. Subsequently, we compared the new Scheimpflug imaging with the Pentacam to analyze their agreement. Many previous studies demonstrated the reliability of Pentacam measurements (6,15). In terms of CCT and TCT measurements, Scansys provided slightly higher measurements compared to those obtained by the Pentacam, which was in line with the previous studies showing that the Pentacam underestimated the values of CCT and TCT (24). Furthermore, Anayol et al. reported that the difference between Pentacam and Galilei was −13.93 and −5.5 µm in the CCT and TCT, respectively (25). Moreover, Huang et al. found that the difference between Pentacam and Sirius was −3.3 and −3.4 µm in the CCT and TCT, respectively (26). Thus, compared to other instruments that use a similar principle, the difference between Scansys and Pentacam was slight and could be considered clinically insignificant. The only exceptions were measurements at 5 mm in the nasal and superior quadrants, where the difference was higher than 15µm. As far as the agreement is concerned, central measurements revealed a high agreement between the two devices, thus suggesting they could be used interchangeably. Previous studies with other Scheimpflug-based instruments reported wider 95% LoAs than our present results. De la Parra-Colin et al. reported that the 95% LoAs ranged from −27.7 to 7.5 µm in the CCT and from −31.7 to −6.8 µm in the TCT when comparing the measurements by Pentacam and Sirius, respectively (27). Similarly, Nasser et al. reported that the range of LoA was from −9.61 to 33.44 µm in the TCT (28). Additionally, 95% LoAs (from 11.74 to 16.12 µm in the CCT and from 3.34 to 7.65 µm in the TCT) were also reported by Anayol et al. when comparing Galilei with Pentacam (25). In addition, we also evaluated the agreement of CT5mm measurements performed with the two instruments. According to results from a previous study that compared Pentacam and Visante OCT (Zeiss, Dublin, CA), Pentacam provided a larger value of MPCT (33.4±2.7 µm) (6). In the present study, Pentacam also provided a larger corneal thickness measurement in the CT5mm region. Bland-Altman plots revealed that there was only a moderated agreement in the CT5mm region between the two devices so that their measurements were not interchangeable. The lower level of agreement in the mid-peripheral region may be due to the image distortion of the Scheimpflug technique in the peripheral region, as shown by previous studies (17,29), and to the different number of scans (28 vs. 25) and analyzed data points (107,520 vs. 138,000 points). According to the results of Huang et al., the 95% LoAs between Scansys and Pentacam were narrower than the results among the RTVue OCT and three Scheimpflug cameras (range approximately 40 µm) (26). Comparing Sirius and Visante OCT, Milla et al. reported that the wider 95% LoAs were −7.6 to 59.1 µm, −9.2 to 77.5 µm, −24.0 to 42.8 µm, −0.4 to 51.6 µm at nasal, superior, temporal, inferior 2.5 mm from the center, respectively (16). In another study by Bourges et al. (30), a wider range of the 95% LoAs was also found between Orbscan II and Pentacam (the maximum ranged from −6.2 to 59.8 µm, the minimum ranged −22.3 to 26.9 µm). Martin et al. also found a poor agreement (spanning over 130 µm) between Orbscan II and Ultrasound (23). This study has a few limitations. First, only young, healthy subjects with normal corneas were examined. Future studies should involve patients who received refractive surgery or with keratoconus. Second, only two devices were evaluated. Future studies should evaluate several other instruments based on Scheimpflug technology at the same time. In conclusion, the new Scheimpflug imaging Scansys showed high intra-observer repeatability and inter-observer reproducibility of corneal thickness measurements. In terms of agreement with Pentacam, their measurements can be considered interchangeable in the central region and at 2 mm, whereas they cannot be considered interchangeable for measurements at 5 mm. The article’s supplementary files as
  30 in total

1.  Accuracy of RTVue optical coherence tomography, Pentacam, and ultrasonic pachymetry for the measurement of central corneal thickness.

Authors:  Sang Min Nam; Chan Young Im; Hyung Keun Lee; Eung Kweon Kim; Tae-Im Kim; Kyoung Yul Seo
Journal:  Ophthalmology       Date:  2010-06-16       Impact factor: 12.079

2.  Repeatability and reproducibility of pachymetric mapping with Visante anterior segment-optical coherence tomography.

Authors:  Shaheeda Mohamed; Gary K Y Lee; Srinivas K Rao; Amy L Wong; Arthur C K Cheng; Emmy Y M Li; Stanley C C Chi; Dennis S C Lam
Journal:  Invest Ophthalmol Vis Sci       Date:  2007-12       Impact factor: 4.799

3.  Discrepancy between central and midperipheral corneal thickness measurements obtained with slit-scanning pachymetry and noncontact specular microscopy.

Authors:  Sven Jonuscheit; Michael J Doughty
Journal:  J Cataract Refract Surg       Date:  2009-12       Impact factor: 3.351

4.  Repeatability of Pentacam peripheral corneal thickness measurements.

Authors:  Raul Martin; Sven Jonuscheit; Ana Rio-Cristobal; Michael J Doughty
Journal:  Cont Lens Anterior Eye       Date:  2015-06-11       Impact factor: 3.077

5.  A Comparison between Scheimpflug imaging and optical coherence tomography in measuring corneal thickness.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Chao Pan; Yifan Feng; Dan Cheng; Yanjun Hua; Xiuli Hu; Qinmei Wang
Journal:  Ophthalmology       Date:  2013-05-11       Impact factor: 12.079

6.  Comparison of central corneal thickness, thinnest corneal thickness, anterior chamber depth, and simulated keratometry using galilei, Pentacam, and Sirius devices.

Authors:  Mustafa A Anayol; Emre Güler; Ramazan Yağci; Mehmet A Şekeroğlu; Meltem Ylmazoğlu; Hakan Trhş; Ali E Kulak; Pelin Ylmazbaş
Journal:  Cornea       Date:  2014-06       Impact factor: 2.651

7.  The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma.

Authors:  Mae O Gordon; Julia A Beiser; James D Brandt; Dale K Heuer; Eve J Higginbotham; Chris A Johnson; John L Keltner; J Philip Miller; Richard K Parrish; M Roy Wilson; Michael A Kass
Journal:  Arch Ophthalmol       Date:  2002-06

8.  Comparison of three methods of measuring corneal thickness and anterior chamber depth.

Authors:  Wolf Buehl; Danijela Stojanac; Stefan Sacu; Wolfgang Drexler; Oliver Findl
Journal:  Am J Ophthalmol       Date:  2006-01       Impact factor: 5.258

9.  Risk assessment for ectasia after corneal refractive surgery.

Authors:  J Bradley Randleman; Maria Woodward; Michael J Lynn; R Doyle Stulting
Journal:  Ophthalmology       Date:  2007-07-12       Impact factor: 12.079

10.  Central and midperipheral corneal thickness measured with Scheimpflug imaging and optical coherence tomography.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2014-05-22       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.