Literature DB >> 16129300

Reliability of Orbscan II topography measurements in relation to refractive status.

Tanja M Rabsilber1, Klio A Becker, Gerd U Auffarth.   

Abstract

PURPOSE: To investigate the reliability of corneal topography measurements using the Orbscan II topography system (Bausch & Lomb) not only referring to a normal cohort but also to different refractive conditions.
SETTING: Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany.
METHODS: Eighty patients (mean age 46.4 +/- 19.0 years) were assigned to 4 refractive groups (Group A: emmetropia (n=20); Group B: astigmatism (n=20) (-1.98 +/- 1.77 diopters [D]); Group C: hyperopia (n=20) (+4.84 +/- 1.6 D); Group D: myopia (n=20) (-9.64 +/- 3.79 D). Three measurements were performed in a series. Thirteen defined, standardized points of the entire cornea (apex, 3.0 mm, 5.0 mm, and 7.0 mm zone) were evaluated for 3 different maps (anterior elevation, pachymetry, keratometry).
RESULTS: In all 80 patients, the following mean values were found in relation to analyzed zones: with regard to the anterior elevation map, the values decreased from 5.53 microm (center) to -6.52 microm (7.0 mm zone), the corneal thickness increased from 549.41 microm to 638.63 microm peripheral and the keratometry from 43.86 D (apex) to 45.4 D (7.0 mm zone), respectively. Analysis of the 3 different maps in all 4 refraction groups showed a tendency toward an increase in SD from the center to the 7.0 mm zone for anterior elevation and pachymetry maps. For keratometry, however, the lowest SD was found in the 7.0 mm zone. In particular, patients with hyperopia showed significant differences (P<.01), compared with emmetropic patients; the SD was higher with regard to anterior elevation as well as keratometry in peripheral zones. Group B (astigmatism) showed significantly higher SD for anterior elevation in zones 3.0, 5.0, and 7.0 mm when compared with emmetropic patients.
CONCLUSIONS: The repeated Orbscan II measurements showed SD in the micrometer range for anterior elevation and pachymetry but values between 0.48 D and 0.97 D for keratometric data. Patients with astigmatism and especially hyperopia showed significantly higher SD values in peripheral zones for anterior elevation and keratometry, indicating a lower reliability, compared with the emmetropic cohort. However, the different ages of the patients could also be a possible explanation for these findings. Therefore, the Orbscan II seems to be a predictable and useful device for measuring corneal topography.

Entities:  

Mesh:

Year:  2005        PMID: 16129300     DOI: 10.1016/j.jcrs.2005.01.013

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  [Scheimpflug and topography systems in ophthalmologic diagnostics].

Authors:  G U Auffarth; A F M Borkenstein; A Ehmer; A Mannsfeld; T M Rabsilber; M P Holzer
Journal:  Ophthalmologe       Date:  2008-09       Impact factor: 1.059

2.  Comparison of techniques for measuring anterior chamber depth: Orbscan imaging, Smith's technique, and van Herick's method.

Authors:  Frank Eperjesi; Claire Holden
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-10       Impact factor: 3.117

3.  Central and midperipheral corneal thickness measured with Scheimpflug imaging and optical coherence tomography.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2014-05-22       Impact factor: 3.240

4.  Evaluation of pachymetric measurements with scheimpflug photography-based system and optical coherence tomography pachymetry at different stages of keratoconus.

Authors:  Betül İlkay Sezgin Akçay; Engin Bilge Ozgürhan; Ercüment Bozkurt; Tuğba Kurt; Yusuf Yıldırım; Mediha Gülen Coşar; Aydın Yıldırım; Jülide Canan Umurhan Akkan; Ahmet Demirok
Journal:  J Ophthalmol       Date:  2014-07-20       Impact factor: 1.909

5.  Agreement between Pentacam and handheld Auto-Refractor/Keratometer for keratometry measurement.

Authors:  Hassan Hashemi; Samira Heydarian; Abbas Ali Yekta; Mohamadreza Aghamirsalim; Mahin Ahmadi-Pishkuhi; Mehrnaz Valadkhan; Hadi Ostadimoghaddam; Ahmad Ahmadzadeh Amiri; Mehdi Khabazkhoob
Journal:  J Optom       Date:  2019-07-10

6.  Calculation of Toric Intraocular Lens Power with the Barrett Calculator and Data from Three Keratometers.

Authors:  Jing Dong; Yaqin Zhang; Xiaogang Wang
Journal:  J Trop Med       Date:  2021-08-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.