Accuracy of RTVue optical coherence tomography, Pentacam, and ultrasonic pachymetry for the measurement of central corneal thickness.

Sang Min Nam1, Chan Young Im, Hyung Keun Lee, Eung Kweon Kim, Tae-Im Kim, Kyoung Yul Seo.   

Abstract

OBJECTIVE: To compare the reliability and the agreement in measuring central corneal thickness (CCT) using the following technologies: RTVue Fourier-domain optical coherence tomography (Optovue, Inc., Fremont, CA), Pentacam (Oculus, Inc., Wetzlar, Germany), and ultrasonic pachymetry (USP; Pocket-II; Quantel Medical, Inc., Bozeman, MT).
DESIGN: Evaluation of diagnostic test. PARTICIPANTS: One hundred four eyes of 52 healthy subjects (mean age ± standard deviation, 28.6 ± 4.8 years).
METHODS: One eye from each subject was assigned randomly for a repeatability test in which a single operator performed 3 successive measurements. The other eye underwent an interoperator reproducibility test by 3 operators. Two centering methods of RTVue and 3 types of CCT from Pentacam were investigated. For USP, 1 drop of topical anesthetic was administered, and measurement was initiated 90 seconds later. Agreement among the instruments was evaluated using Bland-Altman plots. MAIN OUTCOME MEASURES: Various types of CCT were compared: central zone average and minimum thickness of RTVue centering on the vertex and the pupil; corneal thickness at the pupil center, apex, and thinnest location from Pentacam; and mean CCT of 5 repeated measurements of USP. The reliability of measurement was assessed using the repeatability or reproducibility coefficient (Rco), the coefficient of variation, and the intraclass correlation coefficient. The limit of agreement was used to analyze concordance.
RESULTS: The Rco of RTVue was 4 to 5 μm, which was comparable with that of USP and better than that of Pentacam (10-11 μm). The Rco was not dependent on centering methods (RTVue) or types of CCT (Pentacam). The location of minimum thickness found by RTVue was less reliable than that of the Pentacam. The central zone average of RTVue was approximately 7 μm larger than the pupil center or apex thickness of Pentacam and approximately 13 μm larger than the CCT measurement of USP. Those discrepancies could be as high as 20 and 23 μm, respectively. The minimum thickness measured by the RTVue was similar to that of Pentacam.
CONCLUSIONS: The RTVue is a rapid and reliable noncontact means of measuring CCT; however, the characteristics of CCT measured by RTVue must be understood when comparing the CCT obtained by the Pentacam or USP. FINANCIAL DISCLOSURE(S): The author(s) have no proprietary or commercial interest in any materials discussed in this article.
Copyright © 2010 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20557940     DOI: 10.1016/j.ophtha.2010.03.002

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  26 in total

1.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

2.  Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT.

Authors:  David Kuerten; Niklas Plange; Eva Charlotte Koch; Antonis Koutsonas; Peter Walter; Matthias Fuest
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-04-21       Impact factor: 3.117

3.  Author reply: central corneal thickness determination in corneal edema.

Authors:  D Kuerten; N Plange; P Walter; M Fuest
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-03-22       Impact factor: 3.117

4.  Repeatability and agreement of central corneal thickness measurement with non-contact methods: a comparative study.

Authors:  Tarannum Mansoori; Nagalla Balakrishna
Journal:  Int Ophthalmol       Date:  2017-04-22       Impact factor: 2.031

5.  [Biometry of the anterior eye segment for implantation of phakic anterior chamber lenses. A comparison of current measurement devices].

Authors:  L Kiraly; G Duncker
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

Review 6.  The thick and thin of the central corneal thickness in glaucoma.

Authors:  Graham W Belovay; Ivan Goldberg
Journal:  Eye (Lond)       Date:  2018-02-15       Impact factor: 3.775

7.  In vivo evaluation of the cornea and conjunctiva of the normal laboratory beagle using time- and Fourier-domain optical coherence tomography and ultrasound pachymetry.

Authors:  Ann R Strom; Dennis E Cortés; Carol A Rasmussen; Sara M Thomasy; Kim McIntyre; Shwu-Fei Lee; Philip H Kass; Mark J Mannis; Christopher J Murphy
Journal:  Vet Ophthalmol       Date:  2015-02-10       Impact factor: 1.644

8.  Effects of corneal thickness distribution and apex position on postoperative refractive status after full-bed deep anterior lamellar keratoplasty.

Authors:  Bing-Hong Wang; Ye-Sheng Xu; Wen-Jia Xie; Yu-Feng Yao
Journal:  J Zhejiang Univ Sci B       Date:  2018 Nov.       Impact factor: 3.066

9.  Reliability of RTVue, Visante, and slit-lamp adapted ultrasonic pachymetry for central corneal thickness measurement.

Authors:  Jin Pyo Hong; Sang Min Nam; Tae-Im Kim; Kyoung Yul Seo; Sang Yeop Lee; Alessandro Meduri; Eung Kweon Kim
Journal:  Yonsei Med J       Date:  2012-05       Impact factor: 2.759

10.  Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound.

Authors:  J Jorge; Jl Rosado; Ja Díaz-Rey; Jm González-Méijome
Journal:  Clin Ophthalmol       Date:  2013-02-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.