| Literature DB >> 26697213 |
Ayong Yu1, Weiqi Zhao1, Giacomo Savini2, Zixu Huang3, Fangjun Bao1, Weicong Lu1, Qinmei Wang1, Jinhai Huang1.
Abstract
Purpose. To assess the repeatability and reproducibility of central corneal thickness (CCT) measurements by corneal dynamic Scheimpflug analyzer Corvis ST in normal eyes and compare the agreement with Pentacam rotating Scheimpflug System and ultrasound pachymetry. Methods. 84 right eyes underwent Corvis ST measurements performed by two operators. The test-retest repeatability (TRT), within-subject coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to evaluate the intraoperator repeatability and interoperator reproducibility. CCT measurements also were obtained from Pentacam and ultrasound pachymetry by the first operator. The agreement between the three devices was evaluated with 95% limits of agreement (LoA) and Bland-Altman plots. Results. Corvis ST showed high repeatability as indicated by TRT ≤ 13.0 μm, CoV < 0.9%, and ICC > 0.97. The interoperator reproducibility was also excellent. The CoV was <0.9%, and ICC was >0.97. Corvis ST showed significantly lower values than Pentacam and ultrasound pachymetry (P < 0.001). The 95% LoA between Corvis ST and Pentacam or ultrasound pachymetry were -15.8 to 9.5 μm and -27.9 to 12.3 μm, respectively. Conclusions. Corvis ST showed excellent repeatability and interoperator reproducibility of CCT measurements in normal eyes. Corvis ST is interchangeable with Pentacam but not with ultrasound pachymetry.Entities:
Year: 2015 PMID: 26697213 PMCID: PMC4678087 DOI: 10.1155/2015/767012
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Intraobserver repeatability of the corneal dynamic Scheimpflug analyzer Corvis ST in measuring central corneal thickness (N = 84).
| Operator | Mean ( |
| TRT ( | CoV (%) | ICC (95% CI) |
|---|---|---|---|---|---|
| 1st | 535.9 ± 27.0 | 4.8 | 13.0 | 0.87 | 0.971 |
| 2nd | 537.4 ± 27.6 | 4.7 | 13.0 | 0.87 | 0.972 |
SD = standard deviation, S = within-subject standard deviation, TRT = test-retest repeatability (2.77S ), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Interobserver reproducibility of central corneal thickness readings using average (from average of 3 consecutive readings from each observer) and single (from the first reading from each observer) measurement by the corneal dynamic Scheimpflug analyzer Corvis ST.
| Parameter | Mean ( |
| TRT ( | CoV (%) | ICC (95% CI) |
|---|---|---|---|---|---|
| Average | 536.6 ± 27.2 | 3.5 | 9.7 | 0.65 | 0.984 |
| Single | 536.4 ± 27.5 | 4.5 | 12.6 | 0.85 | 0.973 |
SD = standard deviation, S = within-subject standard deviation, TRT = test-retest repeatability (2.77S ), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.
Comparison of the central corneal thickness readings obtained using the corneal dynamic Scheimpflug analyzer Corvis ST, Pentacam rotating Scheimpflug system, and ultrasound pachymetry.
| Device pairings | Mean difference ( | 95% LoA ( |
|
|---|---|---|---|
| Corvis-Pentacam | −3.2 ± 6.5 | −15.8 to 9.5 | <0.001 |
| Corvis-USP | −7.8 ± 10.3 | −27.9 to 12.3 | <0.001 |
USP = ultrasound pachymetry, SD = standard deviation.
Figure 1Bland-Altman plot demonstrating central corneal thickness measurements obtained using corneal dynamic Scheimpflug analyzer Corvis ST and Pentacam rotating Scheimpflug system against the mean values for both devices. The 95% limits of agreement are represented as the upper and lower lines.
Figure 2Bland-Altman plot demonstrating central corneal thickness measurements obtained using corneal dynamic Scheimpflug analyzer Corvis ST and ultrasound pachymetry against the mean values for both devices. The 95% limits of agreement are represented as the upper and lower lines.
Summary of previous studies for the agreement of CCT measurements obtained by other Scheimpflug systems, Orbscan II and specular microscopes in comparison to Pentacam or ultrasound pachymetry.
| Author (year) | Patients/eyes | Device pairings | Mean difference ( |
| 95% LoA ( |
|---|---|---|---|---|---|
| Lanza et al. [ | 102/102 | Pentacam-Orbscan II | 13.66 ± 16.53 | <0.0001 | −18.74 to 46.06 |
| Sirius-Orbscan II | 15.18 ± 17.16 | <0.0001 | −18.45 to 48.81 | ||
| Sirius-Pentacam | 1.52 ± 6.21 | 0.015 | −10.65 to 13.69 | ||
|
| |||||
| Khaja et al. [ | 32/32 | USP-Orbscan II | 2.8 ± 0.28 | NA | −30.15 to 24.40 |
| USP-specular microscopy | 8.69 ± 1.24 | NA | −8.82 to 27.4 | ||
|
| |||||
| Smedowski et al. [ | 76/152 | Corvis ST-Pentacam | NA | >0.05 | NA |
| Corvis ST-USP | NA | >0.05 | NA | ||
|
| |||||
| Huang et al. [ | 66/66 | Pentacam-Sirius | −3.3 ± 5.2 | <0.001 | −13.6 to 6.9 |
| Pentacam-Galilei | −9.3 ± 3.7 | <0.001 | −16.6 to −2.0 | ||
| Sirius-Galilei | −6.0 ± 4.0 | <0.001 | −13.8 to 1.9 | ||
|
| |||||
| Anayol et al. [ | 32/32 | Galilei-Pentacam | 13.93 ± 0.88 | <0.001 | 11.74 to 16.12 |
| Galilei-Sirius | 14.66 ± 0.69 | <0.001 | 12.96 to 16.37 | ||
| Pentacam-Sirius | 0.73 ± 0.93 | 1.0 | −1.50 to 3.02 | ||
|
| |||||
| Maresca et al. [ | 35/35 | Sirius-USP | −13.9 ± 14.4 | <0.001 | −42.2 to 14.4 |
|
| |||||
| Feizi et al. [ | 88/88 | USP-Orbscan | −14.5 ± 22.9 | <0.001 | −59.4 to 30.4 |
| USP-Galilei | −16.0 ± 19.6 | <0.001 | −54.5 to 22.5 | ||
| Orbscan-Galilei | −1.5 ± 17.0 | 0.99 | −34.8 to 31.9 | ||
|
| |||||
|
De La Parra-Colín et al. [ | 16/16 | Sirius-Pentacam | −10.1 ± 9.0 | NA | −27.7 to 7.5 |
|
| |||||
|
Jorgel et al. [ | 50/50 | Sirius-USP | 4.68 ± 10.47 | 0.003 | −15.84 to 25.20 |
|
| |||||
| Bayhan et al. [ | 50/50 | USP-Sirius | 17.58 ± 8.13 | <0.001 | 15.27 to 19.89 |
|
| |||||
| Huang et al. [ | 43/43 | Sirius-USP | 6.88 ± 6.77 | 0.000 | −6.39 to 20.14 |
|
| |||||
| Nassiri et al. [ | 32/61 | Pentacam-USP | −1 ± 9 | 0.32 | −20 to 17 |
| Orbscan II-USP | 6 ± 14 | <0.001 | −21 to 33 | ||
|
| |||||
| Al Farhan et al. [ | 30/30 | USP-specular microscopy | −2.40 ± 9.10 | 0.16 | −38.70 to 39.90 |
|
| |||||
| Tai et al. [ | 92/184 | Pentacam-USP | 10.08 ± 10.96 | 0.012 | −11.40 to 31.56 |
| Specular microscopy-USP | −20.49 ± 8.91 | <0.001 | −37.95 to −3.04 | ||
| Specular microscopy-Pentacam | −30.57 ± 10.26 | <0.001 | −59.69 to −10.45 | ||
|
| |||||
| Chen et al. [ | 35/35 | Pentacam-USP | 5.27 ± 9.55 | 0.007 | −24.0 to −13.4 |
|
| |||||
| Aramberri et al. [ | 35/35 | Pentacam-Galilei | −2.76 ± 4.52 | <0.01 | −6.1 to 11.6 |
|
| |||||
|
González-Pérez et al. [ | 22/22 | USP-Pentacam | 3 ± 10 | 0.233 | −16.2 to 21.2 |
| USP-Orbscan II | 32 ± 15 | <0.001 | 3.1 to 60.8 | ||
| USP-specular microscopy | 26 ± 37 | 0.004 | −46.2 to 97.8 | ||
| Pentacam-Orbscan II | 29 ± 11 | <0.001 | 7.2 to 51.6 | ||
| Pentacam-specular microscopy | 23 ± 32 | 0.003 | −40.2 to 86.2 | ||
| Orbscan II-specular microscopy | −6 ± 35 | 0.399 | −74.8 to 62.0 | ||
| USP-Orbscan II | −15 ± 17 | 0.001 | −47.9 to 18.7 | ||
| Pentacam-Orbscan II | −17 ± 14 | <0.001 | −43.9 to 9.8 | ||