Central and peripheral pachymetry measurements according to age using the Pentacam rotating Scheimpflug camera.

Ramin Khoramnia1, Tanja M Rabsilber, Gerd U Auffarth.   

Abstract

PURPOSE: To investigate the mean values and standard deviations as well as the reliability of consecutive examinations of central and peripheral corneal thickness measurements according to age using the Pentacam rotating Scheimpflug camera (Oculus, Inc.).
SETTING: Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany.
METHODS: Seventy-six healthy volunteers were enrolled in a clinical prospective study. Three consecutive Pentacam measurements of 1 eye per subject were taken. Evaluated were the central corneal thickness (CCT), corneal thickness at 4 peripheral points (3.0 mm superior, inferior, nasal, and temporal), and the thinnest point of the cornea. The volunteers were then assigned to 3 groups to assess the influence of increasing age on the study parameters.
RESULTS: The mean age of the subjects was 46.6 years +/- 16.8 (SD). The mean CCT was 539.62 +/- 31.87 microm. Peripherally, the corneal thickness was between 11% and 19% higher than centrally, with the superior cornea being the thickest followed by the nasal, the inferior, and the temporal cornea. The thinnest point was located in the inferotemporal quadrant in 92% of eyes and in the superotemporal quadrant in 8%. There was no correlation between age and corneal thickness. Minor mean standard deviations of consecutive measurements were noted in the corneal center (4.33 microm), increasing significantly toward the periphery (mean 8.31 microm). Increasing age was not associated with decreasing reliability.
CONCLUSIONS: Using the Pentacam, it was possible to acquire information about corneal thickness across the entire cornea. No correlation was found between increasing age and change in peripheral or CCT. Good reliability was noted for pachymetry measurements decreasing slightly toward the periphery, which was independent of age.

Mesh:

Year:  2007        PMID: 17466857     DOI: 10.1016/j.jcrs.2006.12.025

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  24 in total

1.  [Scheimpflug and topography systems in ophthalmologic diagnostics].

Authors:  G U Auffarth; A F M Borkenstein; A Ehmer; A Mannsfeld; T M Rabsilber; M P Holzer
Journal:  Ophthalmologe       Date:  2008-09       Impact factor: 1.059

2.  [Influence of corneal pachymetric changes on functional results after cataract surgery].

Authors:  F N Auerbach; M P Holzer; G U Auffarth; R Khoramnia; B C Thomas; D Saure; T M Rabsilber
Journal:  Ophthalmologe       Date:  2015-10       Impact factor: 1.059

3.  Anterior segment parameters in Indian young adults using the Pentacam.

Authors:  Seyed Mahdi Ahmadi Hosseini; Fereshteh Abolbashari; Norhani Mohidin
Journal:  Int Ophthalmol       Date:  2013-03-01       Impact factor: 2.031

4.  Anatomical characterization of central, apical and minimal corneal thickness.

Authors:  Federico Saenz-Frances; Martha Cecilia Bermúdez-Vallecilla; Lara Borrego-Sanz; Luis Jañez; José María Martinez-de-la-Casa; Laura Morales-Fernandez; Enrique Santos-Bueso; Julián Garcia-Sanchez; Julián Garcia-Feijoo
Journal:  Int J Ophthalmol       Date:  2014-08-18       Impact factor: 1.779

5.  Comparison of central corneal thickness and endothelial cell measurements by Scheimpflug camera system and two noncontact specular microscopes.

Authors:  Irmak Karaca; Suzan Guven Yilmaz; Melis Palamar; Halil Ates
Journal:  Int Ophthalmol       Date:  2017-07-03       Impact factor: 2.031

6.  Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.

Authors:  Saleh Al-Ageel; Abdulrahman M Al-Muammar
Journal:  Saudi J Ophthalmol       Date:  2009-10-24

7.  Effects of corneal thickness distribution and apex position on postoperative refractive status after full-bed deep anterior lamellar keratoplasty.

Authors:  Bing-Hong Wang; Ye-Sheng Xu; Wen-Jia Xie; Yu-Feng Yao
Journal:  J Zhejiang Univ Sci B       Date:  2018 Nov.       Impact factor: 3.066

8.  Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes.

Authors:  Ayong Yu; Weiqi Zhao; Giacomo Savini; Zixu Huang; Fangjun Bao; Weicong Lu; Qinmei Wang; Jinhai Huang
Journal:  J Ophthalmol       Date:  2015-12-01       Impact factor: 1.909

9.  Evaluation of the shape symmetry of bilateral normal corneas in a Chinese population.

Authors:  Fangjun Bao; Hao Chen; Ye Yu; Jiguo Yu; Shi Zhou; Jing Wang; QinMei Wang; Ahmed Elsheikh
Journal:  PLoS One       Date:  2013-08-29       Impact factor: 3.240

10.  Central and midperipheral corneal thickness measured with Scheimpflug imaging and optical coherence tomography.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2014-05-22       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.