Repeatability of corneal thickness measured using an Oculus Pentacam.

Marco A Miranda1, Hema Radhakrishnan, Clare O'Donnell.   

Abstract

PURPOSE: To evaluate the repeatability of corneal thickness measurements over time, obtained with an Oculus Pentacam.
METHODS: Measurements of corneal thickness (apical and peripheral) were performed at three different sessions in one eye of 23 healthy subjects, using an Oculus Pentacam instrument. In each session, three consecutive measurements were taken (without realignment) by the same observer. Measurements were performed, immediately after a blink.
RESULTS: Mean values obtained were in agreement with published literature values for central corneal thickness (mean +/- SEM: 546.23 +/- 5.24 microm for first session over the first 1 min, 545.13 +/- 5.49 microm for second session 1 h later, and 543.23 +/- 4.96 microm for the third session measured 1 week later). The data showed an increase in corneal thickness towards the periphery with a decrease in intraobserver reproducibility. The confidence limits in the Bland and Altman charts for pachymetry at the corneal apex ranged between -10.91 microm and 13.00 microm within the first session measured over the first 1 min, -13.43 microm to 11.23 microm between the first and second sessions after 1 h and -11.99 microm to 10.02 microm between the first and third sessions after 1 week. The confidence limits were found to increase towards the peripheral cornea. There was no significant difference in within-subjects variance between the three time scales (p > 0.05 for all parameters measured).
CONCLUSION: Our data showed that the Oculus Pentacam provides repeatable measures of corneal thickness when measured within a few seconds, after 1 h, and 1 week.

Entities:  

Mesh:

Year:  2009        PMID: 19214132     DOI: 10.1097/OPX.0b013e318196a737

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  10 in total

1.  Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Authors:  Yong-Heng Luo; Qian Zhong; Ping-Bo Ouyang; Xiao-Jian Guo; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Reproducibility of optical quality parameters measured at objective and subjective best focuses in a double-pass system.

Authors:  Ai-Lian Hu; Li-Ya Qiao; Ye Zhang; Xiao-Gu Cai; Lei Li; Xiu-Hua Wan
Journal:  Int J Ophthalmol       Date:  2015-10-18       Impact factor: 1.779

3.  Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness.

Authors:  Xuepei Li; Yijing Zhou; Charlotte Aimee Young; Aiming Chen; Guangming Jin; Danying Zheng
Journal:  Ann Transl Med       Date:  2020-07

4.  Central and midperipheral corneal thickness measured with Scheimpflug imaging and optical coherence tomography.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2014-05-22       Impact factor: 3.240

5.  Evaluating the repeatability of corneal elevation through calculating the misalignment between Successive topography measurements during the follow up of LASIK.

Authors:  XiaoBo Zheng; WeiHua Yang; LiFang Huang; JunJie Wang; Si Cao; Brendan Geraghty; YiPing Zhao; QinMei Wang; FangJun Bao; Ahmed Elsheikh
Journal:  Sci Rep       Date:  2017-06-09       Impact factor: 4.379

6.  Aberrant ocular architecture and function in patients with Klinefelter syndrome.

Authors:  Cristin Brand; Michael Zitzmann; Nicole Eter; Sabine Kliesch; Joachim Wistuba; Maged Alnawaiseh; Peter Heiduschka
Journal:  Sci Rep       Date:  2017-10-13       Impact factor: 4.379

7.  Pachymetry and Elevation Back Map Changes in Keratoconus Patients After Crosslinking Procedure.

Authors:  Melisa Ahmedbegovic Pjano; Alma Biscevic; Senad Grisevic; Ivan Gabric; Ajla Skopljak Salkica; Nina Ziga
Journal:  Med Arch       Date:  2020-04

8.  The short-term effects of wearing swimming goggles on corneal biomechanics.

Authors:  Raimundo Jiménez; Rubén Molina; Jesús Vera; Beatriz Redondo
Journal:  Int Ophthalmol       Date:  2022-04-04       Impact factor: 2.029

9.  Effect of Misalignment between Successive Corneal Videokeratography Maps on the Repeatability of Topography Data.

Authors:  FangJun Bao; JunJie Wang; JinHai Huang; Ye Yu; ManLi Deng; LinNa Li; Ayong Yu; QinMei Wang; Pinakin Gunvant Davey; Ahmed Elsheikh
Journal:  PLoS One       Date:  2015-11-23       Impact factor: 3.240

10.  Limbal Metabolic Support Reduces Peripheral Corneal Edema with Contact-Lens Wear.

Authors:  Young Hyun Kim; Meng C Lin; Clayton J Radke
Journal:  Transl Vis Sci Technol       Date:  2020-06-30       Impact factor: 3.283

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.