David P Piñero1, Inmaculada Cabezos2, Alberto López-Navarro3, Dolores de Fez2, María T Caballero2, Vicente J Camps2.
Abstract
BACKGROUND: To evaluate the intrasession repeatability of anterior chamber depth (ACD), central (CCT) and peripheral corneal thickness (PCT), white-to-white diameter (WTW), and irido-corneal angle (IA) measurements obtained with a multidiagnostic device in healthy eyes.Entities:
Keywords: Anterior chamber depth; Irido-corneal angle; Pachymetry; Scheimpflug imaging; VX120 system; White-to-white
Mesh:
Year: 2017 PMID: 29047369 PMCID: PMC5648429 DOI: 10.1186/s12886-017-0589-x
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1Horizontal scan obtained with the Scheimpflug camera of the VX120 system in which peripheral pachymetric measurements are done
Main characteristics of the sample evaluated
| Mean (SD) | Median | Range | |
|---|---|---|---|
| Age (years) | 47.8 (8.3) | 47.0 | 23.0 to 65.0 |
| Mean central corneal radius (mm) | 7.92 (0.30) | 7.86 | 7.26 to 8.84 |
| Mean central corneal astigmatism (D) | 0.82 (0.59) | 0.75 | 0.00 to 3.75 |
| Manifest sphere (D) | −0.43 (2.02) | 0.00 | −7.00 to 6.00 |
| Manifest cylinder (D) | −0.40 (0.75) | 0.00 | −4.00 to 0.00 |
| Manifest spherical equivalent (D) | −0.63 (2.03) | 0.00 | −7.33 to 5.50 |
| Corneal eccentricity | 0.34 (0.31) | 0.43 | −1.16 to 0.83 |
Abbreviations: SD standard deviation, D diopter
Summary of the intrasession repeatability outcomes for the anatomical measurements evaluated with the VX120 system
| Overall mean (SD) Overall median (Range) | Sw | Pr | CV (%) | ICC (Range 95% CI) | |
|---|---|---|---|---|---|
| ACD (mm) | 2.86 (0.29) | 0.03 | 0.06 | 1.16 | 0.992 |
| PCTN3 (μm) | 592.04 (39.87) | 8.98 | 17.59 | 1.52 | 0.976 |
| PCTN2 (μm) | 567.77 (37.83) | 7.05 | 13.81 | 1.25 | 0.984 |
| PCTN1 (μm) | 574.89 (45.35) | 7.70 | 15.09 | 1.34 | 0.984 |
| CCT (μm) | 545.88 (34.70) | 6.41 | 12.56 | 1.17 | 0.983 |
| PCTT1 (μm) | 553.27 (37.44) | 5.91 | 11.59 | 1.08 | 0.987 |
| PCTT2 (μm) | 568.70 (37.58) | 6.06 | 11.87 | 1.07 | 0.987 |
| PCTT3 (μm) | 593.76 (38.58) | 8.33 | 16.33 | 1.41 | 0.979 |
| IAN (°) | 37.17 (4.35) | 0.84 | 1.65 | 2.23 | 0.970 |
| IAT (°) | 39.63 (4.59) | 0.52 | 1.02 | 1.32 | 0.994 |
| WTW (mm) | 12.21 (0.57) | 0.24 | 0.46 | 1.95 | 0.873 |
Abbreviations: SD standard deviation, CV coefficient of variation, S within-subject standard deviation, Pr intrasubject precision: 1.96 x Sw, ICC intraclass correlation coefficient, ACD anterior chamber depth, PCT , PCT , and PCT peripheral corneal thickness at 1, 2 and 3 mm from vertex nasally, PCT , PCT , and PCT peripheral corneal thickness at 1, 2 and 3 mm from vertex temporally, CCT central corneal thickness, IA and IA nasal and temporal irido-corneal angles and WTW horizontal white-to-white corneal diameter
Summary of correlations between the different anatomical parameters and their within-subject standard deviation (Sw,) and coefficient of variation (CV) associated
| Sw | CV | |||
|---|---|---|---|---|
| Pearson correlation coefficient |
| Pearson correlation coefficient |
| |
| ACD (mm) | 0.033 | 0.734 | −0.058 | 0.550 |
| PCTN3 (μm) | 0.024 | 0.893 | −0.082 | 0.652 |
| PCTN2 (μm) | −0.125 | 0.487 | −0.220 | 0.218 |
| PCTN1 (μm) | 0.101 | 0.577 | 0.000 | 0.999 |
| CCT (μm) | 0.149 | 0.127 | 0.056 | 0.565 |
| PCTT1 (μm) | −0.069 | 0.704 | −0.158 | 0.381 |
| PCTT2 (μm) | −0.059 | 0.746 | −0.142 | 0.431 |
| PCTT3 (μm) | 0.000 | 0.997 | −0.100 | 0.580 |
| IAN (°) | 0.204 | 0.254 | 0.122 | 0.498 |
| IAT (°) | 0.039 | 0.827 | −0.102 | 0.572 |
| WTW (mm) | −0.109 | 0.282 | −0.154 | 0.125 |
Abbreviations: SD standard deviation, CV coefficient of variation, S within-subject standard deviation, Pr intrasubject precision: 1.96 x Sw, ICC intraclass correlation coefficient, ACD anterior chamber depth, PCT , PCT , and PCT peripheral corneal thickness at 1, 2 and 3 mm from vertex nasally, PCT , PCT , and PCT peripheral corneal thickness at 1, 2 and 3 mm from vertex temporally, CCT central corneal thickness, IA and IA nasal and temporal irido-corneal angles, and WTW horizontal white-to-white corneal diameter
Summary of the results of previous studies evaluating the consistency of anterior chamber depth (ACD), peripheral corneal thickness (PCT), central corneal thickness (CCT), iridocorneal angle (IA), and white-to-white corneal diameter (WTW)
| Authors | Year | Number of eyes | Type of eyes | System | Parameter | 1.96 Sw | Sw | ICC | CoV |
|---|---|---|---|---|---|---|---|---|---|
|
| 2016 | 100 | Healthy | IOL Master 700 | ACD | – | 0.04 mm | – | 1.22% |
|
| 2016 | 40 | Healthy | Pentacam HR | WTW | 0.5 | – | – | – |
|
| 2016 | 140 | Cataract | Galilei G6 | WTW | – | – | > 0.980 | – |
|
| 2016 | 120 | Cataract | IOL Master 700 | CCT | – | 19.5 μm | – | – |
|
| 2016 | 100 | Healthy | Sirius | CCT | 5 μm | – | – | – |
|
| 2015 | 52 | Healthy | Aladdin | WTW | – | 0.80 mm | > 0.94 | < 0.89% |
|
| 2015 | 100 | Cataract | IOL Master 500/700 | WTW | – | – | 0.93 to 1.00 | – |
|
| 2015 | 100 | Healthy | Sirius | IA | – | – | – | <2% |
|
| 2015 | 71 | Healthy | Galilei G2 | ACD | – | 0.04 to 0.07 mm | – | – |
|
| 2014 | 55 | Keratoconus | Pentacam | ACD | – | 0.03 mm | – | 1.1% |
|
| 2014 | 84 | Healthy | Galilei G2 | CCT | – | – | – | < 1% |
|
| 2013 | 56 | Healthy | Lenstar 900 | WTW | – | 0.274 mm | – | – |
|
| 2013 | 61 | Keratoconus | Sirius | WTW | --- | 0.07 mm | 0.989 | 0.56% |
|
| 2012 | 40 | Healthy | Sirius | WTW | – | 0.04 mm | – | – |
|
| 2012 | 117 | Healthy | Sirius | WTW | – | 0.06 mm | 0.974 | 0.48% |
|
| 2012 | 45 | Healthy | Sirius | ACD | – | – | – | 0.56% |
|
| 2011 | 45 | Healthy | Galilei | CCT | – | – | > 0.94 | < 3.5% |
Abbreviations: SD standard deviation, S within-subject standard deviation, ICC intraclass correlation coefficient, CoV coefficient of variation, D diopter