Evaluation of the repeatability of the Lenstar and comparison with two other non-contact biometric devices in myopes.

Jing Zhao1, Zhi Chen, Zimei Zhou, Lan Ding, Xingtao Zhou.   

Abstract

BACKGROUND: The aim was to evaluate the repeatability of a new optical low-coherence reflectometric device Lenstar, as well as its agreement with the IOL Master and the Pentacam-HR system in myopes.
METHODS: In a prospective study, 56 eyes of 28 myopes were examined with the Lenstar, the IOL Master and the Pentacam-HR system. The repeatability of the Lenstar was assessed. Bland-Altman plots were used to analyse the agreement between the Lenstar and the IOL Master for axial length (AL), anterior chamber depth (ACD from corneal epithelium to lens surface), keratometry readings of flattest and steepest meridian (K1 and K2) and horizontal iris width (white-to-white distance; WTW) measurements and between the Lenstar and the Pentacam-HR system for central corneal thickness (CCT), ACD (from corneal endothelium to lens surface) and K1 and K2 readings.
RESULTS: The variability of the Lenstar was small for all recorded measurements. CCT measurements with the Lenstar and the Pentacam-HR system and ACD measurements with the Lenstar and the other two devices showed narrow 95% limits of agreement (LoA), which imply good agreement. In contrast, range and 95% LoA were clinically significantly different for AL, K1, K2 and WTW values between the Lenstar and the IOL Master as well as for K readings between the Lenstar and the Pentacam-HR system.
CONCLUSIONS: The repeatability of the Lenstar is excellent. In clinical practice, the Lenstar and the Pentacam-HR system can be used interchangeably for CCT and ACD measurements but not for K readings. The Lenstar and the IOL Master can be used interchangeably for ACD measurements but not for AL, K readings or WTW distance values.
© 2012 The Authors. Clinical and Experimental Optometry © 2012 Optometrists Association Australia.

Entities:  

Mesh:

Year:  2012        PMID: 22943766     DOI: 10.1111/j.1444-0938.2012.00793.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  9 in total

1.  Axial biometry of the entire eye using ultra-long scan depth optical coherence tomography.

Authors:  Jianguang Zhong; Yilei Shao; Aizhu Tao; Hong Jiang; Che Liu; Huicheng Zhang; Jianhua Wang
Journal:  Am J Ophthalmol       Date:  2013-10-07       Impact factor: 5.258

2.  Bifocal & Atropine in Myopia Study: Baseline Data and Methods.

Authors:  Juan Huang; Donald O Mutti; Lisa A Jones-Jordan; Jeffrey J Walline
Journal:  Optom Vis Sci       Date:  2019-05       Impact factor: 1.973

3.  Comparison axial length measurements from three biometric instruments in high myopia.

Authors:  Xiao-Gang Wang; Jing Dong; Yu-Lan Pu; Hui-Jun Liu; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

4.  Versatile optical coherence tomography for imaging the human eye.

Authors:  Aizhu Tao; Yilei Shao; Jianguang Zhong; Hong Jiang; Meixiao Shen; Jianhua Wang
Journal:  Biomed Opt Express       Date:  2013-06-04       Impact factor: 3.732

5.  Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Inmaculada Cabezos; Alberto López-Navarro; Dolores de Fez; María T Caballero; Vicente J Camps
Journal:  BMC Ophthalmol       Date:  2017-10-18       Impact factor: 2.209

6.  Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices.

Authors:  Remzi Karadag; Murat Unluzeybek; Ozgur Cakici; Ayse Yagmur Kanra; Huseyin Bayramlar
Journal:  Int Ophthalmol       Date:  2016-10-08       Impact factor: 2.031

7.  Validation of refraction and anterior segment parameters by a new multi-diagnostic platform (VX120).

Authors:  Ariela Gordon-Shaag; David P Piñero; Cyril Kahloun; David Markov; Tzadok Parnes; Liat Gantz; Einat Shneor
Journal:  J Optom       Date:  2018-03-09

8.  Is Hyperopia an Important Risk Factor for PACG in the Dutch Population?-A Case Control Study.

Authors:  Saskia H M van Romunde; Gijs Thepass; Hans G Lemij
Journal:  J Ophthalmol       Date:  2013-09-15       Impact factor: 1.909

9.  Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device.

Authors:  Adam Wylęgała; Bartłomiej Bolek; Robert Mazur; Edward Wylęgała
Journal:  Sci Rep       Date:  2020-09-02       Impact factor: 4.379

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.