Literature DB >> 27100654

Comparison of Corneal Diameter and Anterior Chamber Depth Measurements Using 4 Different Devices.

Mehdi Shajari1, Ulrike C Lehmann, Thomas Kohnen.   

Abstract

PURPOSE: To evaluate the reliability and validity of anterior chamber depth (ACD) and corneal diameter [white-to-white (WTW) distance] measurements with 4 devices as well as applicability of these values in predicting the anterior chamber angle diameter evaluated by optical coherence tomography for implantation of lenses in the anterior chamber.
METHODS: In this prospective randomized case series, we measured ACD and WTW distance in one eye of 40 healthy subjects with Pentacam HR, IOL Master 500, LenStar 900, and Visante OCT.
RESULTS: When comparing WTW distance with anterior chamber angle diameter measured by Visante OCT coefficient of repeatability and limits of agreement ranges were 0.5 and 1.1 mm for IOL Master, 0.5 and 1.1 mm for Pentacam HR, and 0.6 and 1.1 mm for LenStar (P < 0.05 in all cases), respectively. Coefficient of repeatability and limits of agreement ranges were 0.3 and 0.6 mm for IOL Master, 0.3 and 0.5 mm for Pentacam HR, and 0.3 and 0.6 mm for LenStar, respectively, compared with ACD measured by Visante OCT; however, P was not significant in any of the cases. P value was also not significant when comparing IOL Master, LenStar, and Pentacam in terms of ACD and WTW distance with each other.
CONCLUSIONS: Unlike previous groups analyzing the ACD and WTW distance we found no significant difference between the measurements by different automatic devices. We consequently expect no difference in clinical outcome when measurements for WTW distance or ACD are performed with any of the devices listed above.

Entities:  

Mesh:

Year:  2016        PMID: 27100654     DOI: 10.1097/ICO.0000000000000840

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  10 in total

1.  Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.

Authors:  Noelia Martínez-Albert; Jose J Esteve-Taboada; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-02       Impact factor: 3.117

2.  Comparison of anterior segment measurements with LenStar and Pentacam in patients with newly diagnosed glaucoma.

Authors:  Emine Sen; Merve Inanc; Ufuk Elgin; Pelin Yilmazbas
Journal:  Int Ophthalmol       Date:  2017-01-21       Impact factor: 2.031

3.  Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements.

Authors:  Sattar Rajabi; Amir Asharlous; Abbas Riazi; Mehdi Khabazkhoob; Alireza Moalej
Journal:  J Curr Ophthalmol       Date:  2022-04-16

4.  Comparison of two Nd:YAG laser posterior capsulotomy: cruciate pattern vs circular pattern with vitreous strand cutting.

Authors:  Jin-Soo Kim; Jung Yeol Choi; Ji-Won Kwon; Won Ryang Wee; Young Keun Han
Journal:  Int J Ophthalmol       Date:  2018-02-18       Impact factor: 1.779

5.  Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Inmaculada Cabezos; Alberto López-Navarro; Dolores de Fez; María T Caballero; Vicente J Camps
Journal:  BMC Ophthalmol       Date:  2017-10-18       Impact factor: 2.209

6.  Comparative study of anterior segment measurements using 3 different instruments in myopic patients after ICL implantation.

Authors:  Ting Wan; Houfa Yin; Yi Yang; Fang Wu; Zhiyi Wu; Yabo Yang
Journal:  BMC Ophthalmol       Date:  2019-08-13       Impact factor: 2.209

7.  Comparison of a New Optical Biometer That Combines Scheimpflug Imaging With Partial Coherence Interferometry With That of an Optical Biometer Based on Swept-Source Optical Coherence Tomography and Placido-Disk Topography.

Authors:  Shihao Chen; Qiaoyue Zhang; Giacomo Savini; Shuangzhe Zhang; Xiaomin Huang; Jinjin Yu; Yirang Wang; Rui Ning; Jinhai Huang; Ruixue Tu
Journal:  Front Med (Lausanne)       Date:  2022-02-10

8.  Agreement of implantable collamer lens sizes using parameters from different devices.

Authors:  Manachai Nonpassopon; Passara Jongkhajornpong; Prae Phimpho; Nontawat Cheewaruangroj; Kaevalin Lekhanont; Varintorn Chuckpaiwong
Journal:  BMJ Open Ophthalmol       Date:  2022-03-08

9.  Commentary: Reliability of different biometric measurements in implantable collamer lense procedure.

Authors:  Hariprasad V Hebri; Ramya Nayak; Roopashree Rao
Journal:  Indian J Ophthalmol       Date:  2022-05       Impact factor: 2.969

10.  Daily Variations of Corneal White-to-White Diameter Measured with Different Methods.

Authors:  Stefan Bandlitz; Makram Nakhoul; Konstantin Kotliar
Journal:  Clin Optom (Auckl)       Date:  2022-09-20
  10 in total