Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry.

Julio C Hernández-Camarena, Patricia Chirinos-Saldaña, Alejandro Navas, Arturo Ramirez-Miranda, Angle de la Mota, Aida Jimenez-Corona, Enrique O Graue-Hernindez.   

Abstract

PURPOSE: To assess the reliability of three different Scheimpflug systems and their agreement.
METHODS: Eighty-four eyes of 42 patients were examined with three Scheimpflug devices: the Galilei G2 Dual Scheimpflug Analyzer (Ziemer Ophthalmic Systems AG, Port, Switzerland), Pentacam HR system (Oculus Optikgerate GmbH, Wetzlar, Germany), and Sirius 3D imaging system (Costruzione Strumenti Oftalmici, Florence, Italy). The central corneal thickness, maximum anterior and posterior corneal elevation, anterior radius of curvature, total higher-order aberrations, and anterior chamber depth were evaluated. Repeatability and re-producibility were evaluated using coefficients of variation and intraclass correlation coefficients. Interdevice agreement was assessed by Bland-Altman comparison analysis.
RESULTS: The three Scheimpflug systems had excellent repeatability for central corneal thickness, anterior radius of curvature, and anterior chamber depth (coefficients of variation being less than 1%). Repeatability for anterior and posterior corneal evaluation was good for the Pentacam HR and Galilei G2 (coefficients of variation < 10%) and excellent for the Sirius 3D (coefficients of variation = 0). Reproducibility was excellent in the three devices, (intraclass correlation coefficient of more than 0.9). Interdevice agreement was excellent (P > .05) for anterior radius of curvature, central corneal thickness, and anterior chamber depth, but was poor (P < .05) for corneal elevation and higher-order aberrations.
CONCLUSIONS: Reliability was good to excellent in the three devices for all measured variables. Interdevice agreement analysis suggests that measurements for anterior radius of curvature, central corneal thickness, and anterior chamber depth, but not for maximum anterior and posterior corneal elevation and total higher-order aberrations, from the Sirius 3D and Galilei G2 can be interchangeable with the Pentacam HR.

Entities:  

Mesh:

Year:  2014        PMID: 25250418     DOI: 10.3928/1081597X-20140815-02

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  30 in total

1.  Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes.

Authors:  Clara Ruiz-Belda; David P Piñero; Pedro Ruiz-Fortes; Roberto Soto-Negro; Myriam Moya; Rafael J Pérez-Cambrodí; Alberto Artola
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-07-15       Impact factor: 3.117

2.  Validation of Fourier analysis of videokeratographic data.

Authors:  Haris Sideroudi; Georgios Labiris; Fienke Ditzel; Efi Tsaragli; Kimonas Georgatzoglou; Haralampos Siganos; Vassilios Kozobolis
Journal:  Int Ophthalmol       Date:  2017-06-15       Impact factor: 2.031

3.  Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation.

Authors:  Xi-Fang Zhang; Meng Li; Yan Shi; Xiu-Hua Wan; Huai-Zhou Wang
Journal:  Int J Ophthalmol       Date:  2020-04-18       Impact factor: 1.779

4.  Precision of corneal thickness measurements obtained using the scheimpflug-placido imaging and agreement with ultrasound pachymetry.

Authors:  Jinhai Huang; Giacomo Savini; Chengfang Wang; Weicong Lu; Rongrong Gao; Yuanguang Li; Qinmei Wang; Yune Zhao
Journal:  J Ophthalmol       Date:  2015-02-24       Impact factor: 1.909

5.  Comparison of two different scheimpflug devices in the detection of keratoconus, regular astigmatism, and healthy corneas.

Authors:  David Finis; Bernhard Ralla; Maria Karbe; Maria Borrelli; Stefan Schrader; Gerd Geerling
Journal:  J Ophthalmol       Date:  2015-04-14       Impact factor: 1.909

Review 6.  Applications of Scheimpflug Imaging in Glaucoma Management: Current and Potential Applications.

Authors:  Alexander T Nguyen; Tiffany Liu; Ji Liu
Journal:  J Ophthalmol       Date:  2016-12-04       Impact factor: 1.909

7.  Distribution and internal correlations of corneal astigmatism in cataract patients.

Authors:  Yuanfeng Jiang; Ying Qin; Shaochong Bu; Hong Zhang; Xiaomin Zhang; Fang Tian
Journal:  Sci Rep       Date:  2021-06-01       Impact factor: 4.379

8.  Posterior corneal curvature changes following Refractive Small Incision Lenticule Extraction.

Authors:  Sri Ganesh; Utsav Patel; Sheetal Brar
Journal:  Clin Ophthalmol       Date:  2015-07-21

9.  Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug+Placido device.

Authors:  Gaurav Prakash; Dhruv Srivastava
Journal:  J Optom       Date:  2015-09-01

10.  Changes in posterior corneal elevations after combined transepithelial photorefractive keratectomy and accelerated corneal collagen cross-linking: retrospective, comparative observational case series.

Authors:  Hun Lee; David Sung Yong Kang; Byoung Jin Ha; Jin Young Choi; Eung Kweon Kim; Kyoung Yul Seo; Tae-Im Kim
Journal:  BMC Ophthalmol       Date:  2016-08-08       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.