David P Piñero1, Dolores de Fez2, Inmaculada Cabezos2, Alberto López-Navarro3, María T Caballero2, Vicente J Camps2.
Abstract
BACKGROUND: The measurement of the pupillary function is an indispensable test in some eye examinations, being necessary the evaluation of the precision of instruments performing such measures. The aim of this study was to evaluate the intrasession repeatability of pupil size measurements provided by a multidiagnostic platform in a large sample of healthy eyes.Entities:
Keywords: Mesopic; Photopic; Pupil diameter; Pupillometry; Scheimpflug imaging; Scotopic; VX120 system
Mesh:
Year: 2020 PMID: 32867721 PMCID: PMC7457779 DOI: 10.1186/s12886-020-01625-4
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Mean, median, standard deviation (SD), variance and range for each repeated pupillometric measurement (100 patients) under the four illumination conditions evaluated. S: Scotopic level, L: Low level, M: mesopic level, P: Photopic level
| S1 | S2 | S3 | L1 | L2 | L3 | M1 | M2 | M3 | P1 | P2 | P3 | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | 5.38 | 5.32 | 5.33 | 3.23 | 3.24 | 3.23 | 2.88 | 2.96 | 2.93 | 2.57 | 2.60 | 2.60 |
| Median | 5.50 | 5.30 | 5.35 | 3.20 | 3.20 | 3.10 | 2.80 | 2.90 | 2.85 | 2.60 | 2.55 | 2.60 |
| SD | 0.96 | 0.87 | 0.90 | 0.65 | 0.57 | 0.59 | 0.47 | 0.47 | 0.49 | 0.38 | 0.40 | 0.38 |
| Variance | 0.93 | 0.76 | 0.81 | 0.43 | 0.33 | 0.35 | 0.22 | 0.22 | 0.24 | 0.15 | 0.16 | 0.14 |
| Range | 4.00 | 3.80 | 3.70 | 2.80 | 2.90 | 2.70 | 2.60 | 2.90 | 2.70 | 2.30 | 2.50 | 2.40 |
Median, standard deviation (Sw) and maximum difference between repeated measurements under each light condition evaluated. Likewise, the percentage of eyes with a range of variation of repeated measurements of 0.5 mm or lower was also provided
| S | L | M | P | |
|---|---|---|---|---|
| Median (mm) | 5.3 | 3.2 | 2.9 | 2.6 |
| Sw (mm) | 0.2 | 0.1 | 0.1 | 0.1 |
| Max difference (mm) | 1.0 | 0.9 | 1.0 | 0.5 |
| Max difference < =0.5 mm (%) | 82 | 97 | 96 | 100 |
Fig. 1Passing-Bablok interchangeability analysis of the first and second measurements of pupil size under mesopic conditions. The confidence interval of x-intercept (coloured area) and slope of the adjusted lines does not include the values of 0 and 1, respectively
Fig. 2Passing-Bablok interchangeability analysis of the first and third measurements of pupil size under mesopic conditions. The confidence interval of x-intercept (coloured area) and slope of the adjusted lines includes the values of 0 and 1, respectively
Mean, median, standard deviation (SD), variance and range for each eye pupillometric measurement (75 patients) under the four illumination conditions evaluated. S: Scotopic level, L: Low level, M: mesopic level, P: Photopic level
| S_RE | S_LE | L_RE | L_LE | M_RE | M_LE | P_RE | P_LE | |
|---|---|---|---|---|---|---|---|---|
| Mean | 5.17 | 5.17 | 3.17 | 3.15 | 2.94 | 2.93 | 2.59 | 2.57 |
| Median | 5.20 | 5.20 | 3.10 | 3.10 | 2.90 | 2.90 | 2.50 | 2.50 |
| SD | 0.92 | 0.89 | 0.61 | 0.61 | 0.48 | 0.49 | 0.40 | 0.38 |
| Variance | 0.85 | 0.79 | 0.38 | 0.37 | 0.23 | 0.24 | 0.16 | 0.15 |
| Range | 3.90 | 4.00 | 2.80 | 2.90 | 2.50 | 2.70 | 2.30 | 2.10 |