Gaurav Prakash1, Dhruv Srivastava2.
Abstract
PURPOSE: To analyze single session, intrauser reliability of a Scheimpflug device for anterior chamber (AC) and corneal parameters.Entities:
Keywords: Adquisición de imágenes Scheimpflug; Anterior chamber; Capacidad de repetición; Cornea; Cámara anterior; Córnea; Repeatability; Scheimpflug imaging
Mesh:
Year: 2015 PMID: 26338543 PMCID: PMC4812002 DOI: 10.1016/j.optom.2015.07.001
Source DB: PubMed Journal: J Optom ISSN: 1989-1342
Figure 1Output screenshot of the glaucoma summary mode of Sirius Scheimpflug analysis system (CSO, Italy) showing the measured parameters.
Descriptive statistics for repeated measures.
| Parameter | Measurement number | Mean | Std. deviation | 95% confidence interval for mean | Range | P value, repeated measures ANOVA | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| Lower Bound | Upper Bound | Minimum | Maximum | |||||||
| Central corneal thickness (μ) | 1 | 100 | 538.83 | 32.4 | 532.4 | 545.3 | 469 | 649 | 0.6 | 0.9 |
| 2 | 100 | 538.77 | 32 | 532.4 | 545.1 | 470 | 647 | |||
| Total | 200 | 538.80 | 32.1 | 534.3 | 543.3 | 469 | 649 | |||
| Minimum corneal thickness (μ) | 1 | 100 | 534.94 | 32.8 | 528.4 | 541.4 | 465 | 646 | 0.6 | 0.4 |
| 2 | 100 | 534.58 | 32.4 | 528.2 | 541 | 465 | 641 | |||
| Total | 200 | 534.76 | 32.5 | 530.2 | 539.3 | 465 | 646 | |||
| Horizontal visible iris diameter (mm) | 1 | 100 | 12.15 | 0.4 | 12.1 | 12.2 | 11.2 | 13.7 | 0.1 | 0.4 |
| 2 | 100 | 12.15 | 0.4 | 12.1 | 12.2 | 11.2 | 13.7 | |||
| Total | 200 | 12.15 | 0.4 | 12.1 | 12.2 | 11.2 | 13.7 | |||
| Corneal apex thickness (μ) | 1 | 100 | 567.53 | 45.9 | 558.1 | 576.7 | 482 | 695 | 0.3 | 0.9 |
| 2 | 100 | 567.50 | 45.6 | 558.5 | 576.5 | 488 | 690 | |||
| Total | 200 | 567.52 | 45.7 | 561.1 | 573.9 | 482 | 695 | |||
| Corneal apex curvature (D) | 1 | 100 | 45.52 | 2.2 | 45.1 | 46 | 39.9 | 53.9 | 0.9 | 0.1 |
| 2 | 100 | 45.46 | 2.2 | 45 | 45.9 | 40 | 53.1 | |||
| Total | 200 | 45.49 | 2.2 | 45.2 | 45.8 | 39.9 | 53.9 | |||
| Anterior Chamber Depth (from endothelium) (mm) | 1 | 100 | 3.16 | 0.3 | 3.1 | 3.2 | 2.2 | 3.7 | 0.2 | 0.1 |
| 2 | 100 | 3.16 | 0.3 | 3.1 | 3.2 | 2.2 | 3.7 | |||
| Total | 200 | 3.16 | 0.3 | 3.1 | 3.2 | 2.2 | 3.7 | |||
| Anterior chamber volume (mm3) | 1 | 100 | 161.01 | 21.6 | 156.7 | 165.3 | 88 | 193 | 0.1 | 0.8 |
| 2 | 100 | 161.12 | 21.7 | 156.8 | 165.4 | 90 | 194 | |||
| Total | 200 | 161.07 | 21.6 | 158.1 | 164.1 | 88 | 194 | |||
| Iridocorneal angle (°) | 1 | 100 | 43.41 | 5.7 | 42.3 | 44.5 | 30 | 55 | 0.1 | 0.1 |
| 2 | 100 | 43.83 | 5.7 | 42.7 | 45 | 29 | 55 | |||
| Total | 200 | 43.62 | 5.7 | 42.8 | 44.4 | 29 | 55 | |||
| Horizontal anterior chamber distance (mm) | 1 | 100 | 12.21 | 0.5 | 12.1 | 12.3 | 10.7 | 13.7 | 0.2 | 0.6 |
| 2 | 100 | 12.24 | 0.5 | 12.1 | 12.4 | 10.7 | 13.8 | |||
| Total | 200 | 12.23 | 0.5 | 12.2 | 12.3 | 10.7 | 13.8 | |||
| Corneal volume (mm3) | 1 | 100 | 57.17 | 3.1 | 56.6 | 57.8 | 49.6 | 65.5 | 0.7 | 0.9 |
| 2 | 100 | 57.19 | 3 | 56.6 | 57.8 | 49.8 | 64.8 | |||
| Total | 200 | 57.18 | 3 | 56.8 | 57.6 | 49.6 | 65.5 | |||
Limits of agreement and intraclass correlations.
| Parameter | Intrameasurement standard deviation (Sw) | Precision (1.96 × Sw) | Repeatability (2.77 × Sw) | Intraclass correlation (ICC) | ICC 95% limits (lower) | ICC 95% limits (upper) |
|---|---|---|---|---|---|---|
| Central corneal thickness (μ) | 2.56 | 5.02 | 7.09 | 0.997 | 0.995 | 0.998 |
| Minimum corneal thickness (μ) | 2.87 | 5.62 | 7.95 | 0.996 | 0.994 | 0.997 |
| Horizontal visible iris diameter (mm) | 0.03 | 0.06 | 0.08 | 0.997 | 0.996 | 0.998 |
| Corneal apex thickness (μ) | 4.21 | 8.27 | 11.69 | 0.996 | 0.994 | 0.997 |
| Corneal apex curvature (dioptres) | 0.23 | 0.45 | 0.64 | 0.995 | 0.992 | 0.996 |
| Anterior chamber depth (mm) | 0.03 | 0.06 | 0.09 | 0.995 | 0.992 | 0.997 |
| Anterior chamber volume (mm3) | 3.34 | 6.55 | 9.26 | 0.988 | 0.982 | 0.992 |
| Iridocorneal angle (°) | 0.95 | 1.86 | 2.63 | 0.970 | 0.955 | 0.980 |
| Horizontal anterior chamber depth (mm) | 0.06 | 0.12 | 0.16 | 0.964 | 0.947 | 0.976 |
| Corneal volume (mm3) | 0.69 | 1.36 | 1.92 | 0.973 | 0.960 | 0.982 |
Sw is intrameasurement standard deviation, computed as: . Where, n is the number of cases, and di is the difference between two measurement for the same subject i.17, 18, 19
Statistical precision is the difference between a subject's measurement and the true value for 95% of observations and was defined as 1.96 × Sw.17, 18, 19
Repeatability was defined as 2.77 × Sw (= √2 * 1.96 * Sw). The difference between the two observations for the same candidate is expected to be less than the value noted for repeatability for 95% pairs.17, 18, 19