Repeatability and agreement of three Scheimpflug-based imaging systems for measuring anterior segment parameters in keratoconus.

Rohit Shetty1, Vishal Arora2, Chaitra Jayadev1, Rudy M M A Nuijts3, Mukesh Kumar1, Narendra K Puttaiah1, Mathew Kurian Kummelil2.   

Abstract

PURPOSE: To assess the repeatability and agreement of three rotating Scheimpflug cameras, Pentacam, Galilei, and Sirius, in measuring the mean keratometry (Km), thinnest corneal thickness (TCT), anterior chamber depth (ACD), and mean posterior keratometry (pKm) in keratoconus patients in a prospective study.
METHODS: Fifty-five eyes of 55 patients with keratoconus underwent three consecutive scans on each machine, performed by a single operator. Within-subject standard deviation (Sw), test-retest repeatability (TRT), and coefficient of variation (COV) for assessing repeatability and Bland-Altman plots for the agreement between the mean measurements of each machine were examined.
RESULTS: The Sw of Km and pKm measurements with Pentacam (0.23 and 0.10 diopters [D], respectively) were significantly lower (better) than those of Galilei (0.60 and 0.17) and Sirius (0.23 and 0.36). The Sw of TCT measurements with Sirius (8.88 μm) was significantly lower than that of Galilei (11.64 μm). The COV ranged between 0.5 for the Km measurements of Pentacam and 2.8 for the TCT measurements of Galilei. Significant proportional bias in agreement was detected for the pKm measurements with all the three device pairs and for the ACD measurements between Pentacam and Galilei and between Galilei and Sirius.
CONCLUSIONS: Though Pentacam, Galilei, and Sirius showed repeatable measurements for Km, TCT, ACD, and pKm, repeatabilities with Pentacam and Sirius were better than those with Galilei. There were significant differences in the measurements between the three devices; hence they cannot be used interchangeably for anterior segment measurements in keratoconus patients. Copyright 2014 The Association for Research in Vision and Ophthalmology, Inc.

Entities:  

Keywords:  Scheimpflug photography; keratoconus; refractive surgery; tomography

Mesh:

Year:  2014        PMID: 25074774     DOI: 10.1167/iovs.14-15055

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  35 in total

1.  Clinical Evaluation and Validation of the Dutch Crosslinking for Keratoconus Score.

Authors:  Robert P L Wisse; Rob W P Simons; Martijn J B van der Vossen; Marc B Muijzer; Nienke Soeters; Rudy M M A Nuijts; Daniel A Godefrooij
Journal:  JAMA Ophthalmol       Date:  2019-06-01       Impact factor: 7.389

2.  A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

3.  Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes.

Authors:  David P Piñero; Ainhoa Molina-Martín; Vicent J Camps; Dolores de Fez; María Teresa Caballero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-09-03       Impact factor: 3.117

4.  Prevalence of Subclinical Keratoconus Among Pediatric Egyptian Population with Astigmatism.

Authors:  Mohamed Karim Sidky; Dina Hosameldin Hassanein; Sherif A Eissa; Yehia M Salah; Nancy M Lotfy
Journal:  Clin Ophthalmol       Date:  2020-03-23

5.  Response to comments on: Repeatability of Orbscan III for anterior segment parameters in normal eyes.

Authors:  Mohita Sharma; Neha Jain; Ann Sarah Koshy; Vishal Arora; Vaitheeswaran Ganesan Lalgudi
Journal:  Indian J Ophthalmol       Date:  2021-05       Impact factor: 1.848

Review 6.  Current review and a simplified "five-point management algorithm" for keratoconus.

Authors:  Rohit Shetty; Luci Kaweri; Natasha Pahuja; Harsha Nagaraja; Kareeshma Wadia; Chaitra Jayadev; Rudy Nuijts; Vishal Arora
Journal:  Indian J Ophthalmol       Date:  2015-01       Impact factor: 1.848

7.  Comparison of two different scheimpflug devices in the detection of keratoconus, regular astigmatism, and healthy corneas.

Authors:  David Finis; Bernhard Ralla; Maria Karbe; Maria Borrelli; Stefan Schrader; Gerd Geerling
Journal:  J Ophthalmol       Date:  2015-04-14       Impact factor: 1.909

8.  Objective Assessment of Nuclear and Cortical Cataracts through Scheimpflug Images: Agreement with the LOCS III Scale.

Authors:  Alberto Domínguez-Vicent; Ulrika Birkeldh; Carl-Gustaf Laurell; Laurell Carl-Gustaf; Maria Nilson; Rune Brautaset
Journal:  PLoS One       Date:  2016-02-18       Impact factor: 3.240

9.  Inter-rater Reliability and Repeatability of Manual Anterior Segment Optical Coherence Tomography Image Grading in Keratoconus.

Authors:  Anna N Lin; Isa S K Mohammed; Wuqaas M Munir; Saleha Z Munir; Janet L Alexander
Journal:  Eye Contact Lens       Date:  2021-09-01       Impact factor: 3.152

10.  Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug+Placido device.

Authors:  Gaurav Prakash; Dhruv Srivastava
Journal:  J Optom       Date:  2015-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.