Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation.

Xi-Fang Zhang1, Meng Li1, Yan Shi1, Xiu-Hua Wan1, Huai-Zhou Wang1.   

Abstract

AIM: To evaluate the intra-operator repeatability of time domain and swept-source Fourier domain anterior segment optical coherence tomography (AS-OCT), namely, Visante AS-OCT and Casia SS-1000 OCT, in measuring the preoperative parameters of implantable collamer lens (ICL) in myopic eyes, as well as the agreement between the two devices.
METHODS: A total of 97 eyes from 49 myopes were investigated in this prospective case series study. The anterior chamber depth (ACD), angle-to-angle distance (ATA), pupil diameter (PD) and crystalline lens rise (CLR) in all subjects were measured for three times during one session by the same operator. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability limits and intraclass correlation coefficients (ICC). The agreement between the two systems was evaluated using the Bland-Altman plots and 95% limits of agreement (LoA).
RESULTS: The repeatability limits of Visante AS-OCT in measuring ACD, ATA, PD and CLR were 0.099, 0.141, 0.304, and 0.079 mm, respectively. The repeatability limits of Casia SS-1000 OCT in measuring ACD, ATA, PD, and CLR were 0.105, 0.127, 0.357, and 0.082 mm, respectively. Excellent repeatability could be attained in both devices, with the ICC>0.8 for all the measured variables. The interdevice agreement was excellent (P>0.05) for ACD and ATA, but poor (P<0.05) for PD and CLR.
CONCLUSION: Good repeatability can be attained by time domain and swept-source Fourier-domain OCT for all the measured variables. Moreover, interdevice agreement analysis suggests that interchangeable measurements between two devices can be achieved for ACD and ATA, but not for PD and CLR; but the differences in measurements were not clinically significant. International Journal of Ophthalmology Press.

Entities:  

Keywords:  Casia SS-1000 OCT; Visante AS-OCT; agreement; implantable collamer; repeatability

Year:  2020        PMID: 32399415      PMCID: PMC7137697          DOI: 10.18240/ijo.2020.04.15

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  26 in total

1.  Repeatability, reproducibility, and agreement of central anterior chamber depth measurements in pseudophakic and phakic eyes: optical coherence tomography versus ultrasound biomicroscopy.

Authors:  Qi Zhang; Wanqing Jin; Qinmei Wang
Journal:  J Cataract Refract Surg       Date:  2010-06       Impact factor: 3.351

2.  Correcting Interdevice Bias of Horizontal White-to-White and Sulcus-to-Sulcus Measures Used for Implantable Collamer Lens Sizing.

Authors:  Ivo Guber; Ciara Bergin; Sylvie Perritaz; François Majo
Journal:  Am J Ophthalmol       Date:  2015-10-24       Impact factor: 5.258

3.  Assessment and reproducibility of anterior chamber depth measurement with anterior segment optical coherence tomography compared with immersion ultrasonography.

Authors:  Gabor Nemeth; Attila Vajas; Alexis Tsorbatzoglou; Bence Kolozsvari; Laszlo Modis; Andras Berta
Journal:  J Cataract Refract Surg       Date:  2007-03       Impact factor: 3.351

4.  Direct measurement of the ciliary sulcus diameter by 35-megahertz ultrasound biomicroscopy.

Authors:  Jaeryung Oh; Hyung-Ho Shin; Jun-Heon Kim; Hyo-Myung Kim; Jong-Suk Song
Journal:  Ophthalmology       Date:  2007-09       Impact factor: 12.079

5.  Anterior chamber angle imaging with optical coherence tomography.

Authors:  C K-S Leung; R N Weinreb
Journal:  Eye (Lond)       Date:  2011-01-14       Impact factor: 3.775

Review 6.  Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology.

Authors:  Colm McAlinden; Jyoti Khadka; Konrad Pesudovs
Journal:  Ophthalmic Physiol Opt       Date:  2011-05-26       Impact factor: 3.117

7.  Comparison of time-domain, spectral-domain and swept-source OCT in evaluating aqueous cells in vitro.

Authors:  Meng Li; Huaizhou Wang; Yue Liu; Xifang Zhang; Ningli Wang
Journal:  Sci China Life Sci       Date:  2016-11-04       Impact factor: 6.038

8.  Implantable Collamer Lens Sizing Method Based on Swept-Source Anterior Segment Optical Coherence Tomography.

Authors:  Tomoaki Nakamura; Naoki Isogai; Takashi Kojima; Yoko Yoshida; Yasuo Sugiyama
Journal:  Am J Ophthalmol       Date:  2017-12-30       Impact factor: 5.258

9.  Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices.

Authors:  Benjamin Y Xu; Derek D Mai; Rafaella C Penteado; Luke Saunders; Robert N Weinreb
Journal:  J Glaucoma       Date:  2017-11       Impact factor: 2.503

10.  Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Inmaculada Cabezos; Alberto López-Navarro; Dolores de Fez; María T Caballero; Vicente J Camps
Journal:  BMC Ophthalmol       Date:  2017-10-18       Impact factor: 2.209

View more
  4 in total

Review 1.  Advances in OCT Imaging in Myopia and Pathologic Myopia.

Authors:  Yong Li; Feihui Zheng; Li Lian Foo; Qiu Ying Wong; Daniel Ting; Quan V Hoang; Rachel Chong; Marcus Ang; Chee Wai Wong
Journal:  Diagnostics (Basel)       Date:  2022-06-08

2.  Inter-Eye Vault Differences of Implantable Collamer Lens Measured Using Anterior Segment Optical Coherence Tomography.

Authors:  Santiago Cerpa Manito; Angel Sánchez Trancón; Oscar Torrado Sierra; António M G Baptista; Pedro Miguel Serra
Journal:  Clin Ophthalmol       Date:  2020-10-29

3.  Anterior Ocular Biometrics Using Placido-scanning-slit System, Rotating Scheimpflug Tomography, and Swept-source Optical Coherence Tomography.

Authors:  Soyoung Ryu; Sook Hyun Yoon; Ikhyun Jun; Kyoung Yul Seo; Eung Kweon Kim; Tae-Im Kim
Journal:  Korean J Ophthalmol       Date:  2022-05-06

4.  Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography.

Authors:  Sisi Chen; Rongrong Gao; Colm McAlinden; Junming Ye; Yiran Wang; Min Chen; Jinhai Huang; Yong Sun; A-Yong Yu
Journal:  J Ophthalmol       Date:  2020-09-04       Impact factor: 1.909

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.