Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: repeatability and agreement.

Wuhe Chen1, Colm McAlinden, Konrad Pesudovs, Qinmei Wang, Fan Lu, Yifan Feng, Jie Chen, Jinhai Huang.   

Abstract

PURPOSE: To assess the repeatability of common measurements with the Sirius Scheimpflug-Placido topographer and Lenstar LS900 optical low-coherence reflectometry (OLCR) biometer and the limits of agreement (LoA) between the devices.
SETTING: Eye Hospital of Wenzhou Medical College, Wenzhou, China.
DESIGN: Comparative evaluation of a diagnostic test or technology.
METHODS: One randomly healthy eye of subjects was scanned 3 times with both devices. The parameters assessed were central corneal thickness (CCT), anterior chamber depth (ACD) from the corneal epithelium and from the endothelium, mean keratometry (K), and white-to-white (WTW) corneal diameter. The repeatability of scans was calculated using the within-subject standard deviation after 1-way analysis of variance was performed. The agreement between devices was assessed using the Bland-Altman LoA method, which equals the mean difference between devices ± 1.96 × standard deviation of the differences. The mean of 3 scans of each device was used to assess the LoA.
RESULTS: Forty subjects were evaluated. The repeatability of the Scheimpflug-Placido topographer and OLCR biometer was 3.10 μm and 3.32 μm for CCT, 0.04 mm and 0.05 mm for WTW corneal diameter, and 0.17 D and 0.10 D for mean K, respectively. The repeatability for both devices was 0.02 mm for the ACD from the corneal epithelium and the ACD from the corneal endothelium. On Bland-Altman LoA analysis, all parameters were within clinically acceptable limits.
CONCLUSION: Both devices had excellent repeatability for all parameters assessed. Good LoAs were found between the 2 devices, indicating they can be used interchangeably for the parameters assessed. Crown
Copyright © 2012. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22763002     DOI: 10.1016/j.jcrs.2012.04.031

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  18 in total

1.  The effect of pupillary dilatation on IOL power measurement by using the IOLMaster.

Authors:  Bharkbhum Khambhiphant; Nattida Chatbunchachai; Krit Pongpirul
Journal:  Int Ophthalmol       Date:  2015-03-26       Impact factor: 2.031

2.  Agreement of keratometric readings measured using rotating Scheimpflug imaging, auto-refractokeratometer, and biograph in eyes with keratoconus.

Authors:  Meltem Guzin Altınel; Hasim Uslu
Journal:  Int Ophthalmol       Date:  2021-02-05       Impact factor: 2.031

3.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

4.  Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation.

Authors:  Xi-Fang Zhang; Meng Li; Yan Shi; Xiu-Hua Wan; Huai-Zhou Wang
Journal:  Int J Ophthalmol       Date:  2020-04-18       Impact factor: 1.779

5.  Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes.

Authors:  Xiao-Xiao Guo; Ran You; Shan-Shan Li; Xiu-Fen Yang; Lu Zhao; Fan Zhang; Yan-Ling Wang; Xi Chen
Journal:  Int J Ophthalmol       Date:  2019-10-18       Impact factor: 1.779

6.  Growth curves of myopia-related parameters to clinically monitor the refractive development in Chinese schoolchildren.

Authors:  Pablo Sanz Diez; Li-Hua Yang; Mei-Xia Lu; Siegfried Wahl; Arne Ohlendorf
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-03-23       Impact factor: 3.117

7.  Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract.

Authors:  Colm McAlinden; Qinmei Wang; Konrad Pesudovs; Xin Yang; Fangjun Bao; Ayong Yu; Shishi Lin; Yifan Feng; Jinhai Huang
Journal:  PLoS One       Date:  2015-06-10       Impact factor: 3.240

8.  Repeatability of ocular measurements with a dual-Scheimpflug analyzer in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Lorena Fuentes-Broto; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-04-24       Impact factor: 3.411

9.  Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.

Authors:  Jinhai Huang; Giacomo Savini; Hao Chen; Fangjun Bao; Yuanguang Li; Haisi Chen; Weicong Lu; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2015-01-05       Impact factor: 3.240

10.  Impact of tear optics on the repeatability of Pentacam AXL wave and iTrace in measuring anterior segment parameters and aberrations.

Authors:  Gairik Kundu; Rohit Shetty; Pooja Khamar; Sneha Gupta; Ritika Mullick; Vaitheeswaran L Ganesan; Sharon D'Souza
Journal:  Indian J Ophthalmol       Date:  2022-04       Impact factor: 2.969

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.