| Literature DB >> 20537159 |
Xavier Monnet1, Nadia Anguel, Brice Naudin, Julien Jabot, Christian Richard, Jean-Louis Teboul.
Abstract
INTRODUCTION: We compared the ability of two devices estimating cardiac output from arterial pressure-curve analysis to track the changes in cardiac output measured with transpulmonary thermodilution induced by volume expansion and norepinephrine in sepsis patients.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20537159 PMCID: PMC2911755 DOI: 10.1186/cc9058
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Patient characteristics at baseline
| Age (years) | 59 | (53 to 72) |
| Gender (M/F) | 37:43 | |
| SAPS II | 41 | (37 to 80) |
| ARDS ( | 38 | (47) |
| Source of infection | ||
| Pneumonia ( | 64 | (80) |
| Peritonitis ( | 6 | (7) |
| Endocarditis ( | 5 | (6) |
| Fasciitis ( | 2 | (3) |
| Unknown | 3 | (4) |
| CItd (L/min/m2) | 3.1 | (2.2 to 3.5) |
| Vasopressors | ||
| Norepinephrine ( | 53 | (66) |
| Dosage of norepinephrine (μg/kg/min) | 0.43 | (0.21 to 0.71) |
| Dobutamine ( | 6 |
n = 80.
Data are expressed as median [25% to 75% interquartile] or as number (%).
SAPS, Simplified Acute Physiologic Score; ARDS, acute respiratory distress syndrome; MAP, mean arterial pressure; CItd, cardiac index measured with transpulmonary thermodilution.
Evolution of hemodynamic parameters during therapeutic interventions
| Group 1 | Group 2 | |||||||
|---|---|---|---|---|---|---|---|---|
| Baseline | After volume expansion | Baseline | After introduction/increase of norepinephrine | |||||
| Heart rate (beats/min) | 105 | (82-130) | 100 | (89-123) | 92 | (82-120) | 95 | (79-115) |
| MAP (mm Hg) | 69 | (67-75) | 76 | (70-85)a | 52 | (47-58)b | 72 | (64-78)a |
| CItd (L/min/m2) | 3.2 | (2.4-3.6) | 3.7 | (3.1-4.1)a | 2.9 | (2.1-3.5) | 2.9 | (2.5-3.7)a |
| CIpc (L/min/m2) | 3.2 | (2.4-3.6) | 3.4 | (2.8-4.0)a | 2.9 | (2.1-3.5) | 3.0 | (2.4-3.6)a |
| CIpw (L/min/m2) | 2.9 | (2.6-3.2) | 3.2 | (2.8-3.7)a | 2.7 | (2.5-3.0) | 3.0 | (2.8-3.3)a |
| SVRi (dyne/sec/m2/cm5) | 1,741 | 1,355-2,267) | 1,724 | 1,495-1,996) | 1,832 | (1,323-2,675) | 2,229 | (1,740-2,874)a |
| GEDVi (ml/m2) | 649 | (554-715) | 725 | (649-803)a | 690 | (602-474) | 728 | (674-790)a |
n = 40 in Group 1, and n = 40 in Group 2.
Data are expressed as median (25% to 75% interquartile). aP < 0.05, before versus after intervention; bP < 0.05, Group 2 versus Group 1.
MAP, mean arterial pressure; CItd, cardiac index measured by thermodilution; CIpc, pulse contour-based cardiac index measured with the PiCCO device; CIpw, arterial-pressure waveform-based cardiac index measured with the Vigileo device; SVRi, systemic vascular resistance indexed for body surface; GEDVi, global end-diastolic volume indexed for body surface.
Summary of the comparisons between the different techniques used for measuring cardiac index
| Changes induced by volume expansion (Group 1) | Changes induced by introduction/increase in norepinephrine (Group 2) | |
|---|---|---|
| CIpc versus CItd | ||
| Bland-Altman analysis for the changes in absolute value | Bias, -0.20 ± 0.63 L/min/m2 | Bias, -0.05 ± 0.74 L/min/m2 |
| Linear regression for the changes in percentage | ||
| Ability of CIpc to detect an increase in CItd ≥ 15% | Cut-off, CIpc increase ≥ 12% | Cut-off, CIpc increase ≥ 15% |
| Specificity, 74 (49 to 91)% | Specificity, 93 (68 to 99)% | |
| Specificity, 95 (76 to 99)% | Specificity, 88 (69 to 97)% | |
| Area under the ROC curve, 0.878 (0.736 to 0.960) | Area under the ROC curve, 0.924 (0.795 to 0.983) | |
| CIpw versus CItd | ||
| Bland-Altman analysis for the changes in absolute value | Bias, -0.23 ± 0.95 L/min/m2 | Bias, -0.01 ± 1.75 L/min/m2 |
| Linear regression for the changes in percentage | ||
| Ability of CIpw to detect an increase in CItd ≥ 15% | Cut-off, CIpw increase ≥8% | Cut-off, CIpw increase ≥34% |
| Sensitivity, 56 (33 to 80)% | Sensitivity, 27 (8 to 55)% | |
| Specificity, 71 (48 to 89)% | Specificity, 96 (80 to 99)% | |
| Area under the ROC curve, 0.564 (0.398 to 0.720)a | Area under the ROC curve, 0.541 (0.377 to 0.700)a | |
n = 40 in Group 1, and n = 40 in Group 2.
aP < 0.05 vs. CIpc.
CItd, cardiac index measured by transpulmonary thermodilution; CIpc, pulse contour-based cardiac index measured with the PiCCO device; CIpw, arterial pressure waveform-based cardiac index measured with the Vigileo device; AUC, area under the curve; ROC, receiving operating characteristics.
Figure 1Bland-Altman plots. (a, b) Bland-Altman plot for the changes (in absolute values) and correlation (for the percentage changes) induced by volume expansion on cardiac index obtained by transpulmonary thermodilution (CItd) and cardiac index obtained by the pulse-contour analysis (CIpc). (c, d) Bland-Altman plot for the changes (in absolute values) and correlation (for the percentage changes) induced by volume expansion on cardiac index obtained by transpulmonary thermodilution (CItd) and cardiac index obtained by the arterial pressure waveform analysis (CIpw). Bland-Altman plots: straight line, bias; dashed line: +2 SD/-2 SD limits of agreement); dashed line, correlation line.
Figure 2Receiver operating characteristic (ROC) curves. (a) ROC curves constructed for testing the ability of the changes in cardiac index obtained by the pulse-contour analysis (CIpc) (straight line) and of the changes in cardiac index obtained by the arterial-pressure waveform analysis (CIpw) (dashed line) to detect an increase in cardiac index obtained by transpulmonary thermodilution (CItd) ≥15% induced by me expansion. (b) ROC curves constructed for testing the ability of the changes in cardiac index obtained by the pulse-contour analysis (CIpc) (straight line) and of the changes in cardiac index obtained by the arterial-pressure waveform analysis (CIpw) (dashed line) to detect an increase in cardiac index obtained by transpulmonary thermodilution (CItd) ≥15% induced by the introduction/increase in norepinephrine.
Figure 3Bland-Altman plots. (a, b) Bland-Altman plot for the changes (in absolute values) and correlation (for the percentage changes) induced by the introduction/increase in norepinephrine on cardiac index obtained by transpulmonary thermodilution (CItd) and cardiac index obtained by the pulse-contour analysis (CIpc). (c, d) Bland-Altman plot for the changes (in absolute values) and correlation (for the percentage changes) induced by the introduction/increase in norepinephrine on cardiac index obtained by transpulmonary thermodilution (CItd) and cardiac index obtained by the arterial-pressure waveform analysis (CIpw). Plots: straight line, bias; dashed line: +2SD/-2SD limits of agreement); dashed line, correlation line.