Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output.

Berthold Bein1, Frank Worthmann, Peter H Tonner, Andrea Paris, Markus Steinfath, Jürgen Hedderich, Jens Scholz.   

Abstract

OBJECTIVE: Continuous measurement of cardiac output (CCO) is of great importance in the critically ill. However, pulmonary artery thermodilution has been questioned for possible complications associated with right heart catheterization. Furthermore, measurements are delayed in the continuous mode during rapid hemodynamic changes. A new pulmonary artery catheter CCO device (Aortech, Bellshill, Scotland) enabling real-time update of cardiac output was compared with 2 different, less-invasive methods of CCO determination, esophageal Doppler and pulse contour analysis.
DESIGN: Prospective, observational study.
SETTING: University hospital, single institution. PARTICIPANTS: Patients scheduled for elective coronary artery bypass grafting (CABG).
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: CCO measurements were analyzed using a Bland-Altman plot. Bias between CCO and pulse contour cardiac output (PCCO), and Doppler-derived cardiac output (UCCO) was (mean +/- 1 SD) -0.71 +/- 1 L/min versus -0.15 +/- 1.09 L/min, and between UCCO and PCCO -0.58 +/- 1.06 L/min. Bias was not significantly different among methods, nor were comparative values before and after cardiopulmonary bypass (p > 0.05).
CONCLUSIONS: Agreement between the CCO method and both less-invasive measurements was clinically acceptable. There were no adverse events associated with the use of either device.

Entities:  

Mesh:

Year:  2004        PMID: 15073709     DOI: 10.1053/j.jvca.2004.01.025

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  17 in total

1.  The 2-bodied continuous cardiac output catheter.

Authors:  Prabhat Tewari; Sushil Kumar Singh; Neeraj Kale
Journal:  MedGenMed       Date:  2005-07-21

Review 2.  [Pulmonary artery catheter in anaesthesiology and intensive care medicine].

Authors:  E E C de Waal; L de Rossi; W Buhre
Journal:  Anaesthesist       Date:  2006-06       Impact factor: 1.041

3.  Estimation of cardiac output in a pharmacological trial using a simple method based on arterial blood pressure signal waveform: a comparison with pulmonary thermodilution and echocardiographic methods.

Authors:  Jani Penttilä; Amir Snapir; Erkki Kentala; Juha Koskenvuo; Jussi Posti; Mika Scheinin; Harry Scheinin; Tom Kuusela
Journal:  Eur J Clin Pharmacol       Date:  2006-03-29       Impact factor: 2.953

Review 4.  Transesophageal Doppler devices: A technical review.

Authors:  Patrick Schober; Stephan A Loer; Lothar A Schwarte
Journal:  J Clin Monit Comput       Date:  2009-10-20       Impact factor: 2.502

5.  Comparison of the accuracy of transpulmonary thermodilution measurement using indicators of different temperatures.

Authors:  Shulan Chen; Pingdong Lin; Zhenshuang Du; Fangchen Lan; Shanshan Wu; Tiegang Zhong; Xiaohua Liang; Hongyu Liu; Cuiping Zeng; Chenghua Zhang
Journal:  Int J Clin Exp Med       Date:  2014-12-15

Review 6.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

Review 7.  [S3 guidelines on intensive medical care of cardiac surgery patients : Hemodynamic monitoring and cardiovascular system-an update].

Authors:  M Habicher; T Zajonz; M Heringlake; A Böning; S Treskatsch; U Schirmer; A Markewitz; M Sander
Journal:  Anaesthesist       Date:  2018-05       Impact factor: 1.041

8.  Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.

Authors:  Xavier Monnet; Nadia Anguel; Brice Naudin; Julien Jabot; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2010-06-10       Impact factor: 9.097

9.  Monitoring Cardiac Output and Transesophageal Echocardiography during Removal of a Ventricular Assist Device.

Authors:  Aslı Demir; Ümit Karadeniz; Bahar Aydınlı; Murat Taş; Özcan Erdemli
Journal:  Turk J Anaesthesiol Reanim       Date:  2013-05-23

10.  Transoesophageal Doppler monitoring for fluid and hemodynamic treatment during lung surgery.

Authors:  John Diaper; Christoph Ellenberger; Yann Villiger; John Robert; Jean-Marie Tschopp; Marc Licker
Journal:  J Clin Monit Comput       Date:  2008-10-19       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.