Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery.

J Mayer1, J Boldt, T Schöllhorn, K D Röhm, A M Mengistu, S Suttner.   

Abstract

BACKGROUND: Thermodilution technique using a pulmonary artery catheter (PAC) is a widely used method to determine cardiac output (CO). It is increasingly criticized because of its invasiveness and its unclear risk-benefit ratio. Thus, less invasive techniques for measuring CO are highly desirable. We compared a new, semi-invasive device (FloTrac/Vigileo) using arterial pressure waveform analysis for CO measurement in patients undergoing cardiac surgery with bolus thermodilution measurements.
METHODS: Forty patients undergoing coronary artery bypass grafting or valve repair were enrolled. A PAC was inserted and routine radial arterial access was used for semi-invasive determination of CO with the Vigileo. CO was measured simultaneously by bolus thermodilution and the Vigileo technique after induction of anaesthesia (T1), before cardiopulmonary bypass (CPB) (T2), after CPB (T3), after sternal closure (T4), on arrival in the intensive care unit (ICU) (T5), and 4 h (T6), 8 h (T7), and 24 h after surgery (T8). CO was indexed to the body surface area (cardiac index, CI).
RESULTS: A total of 244 pairs of CI measurements were analysed. Bias and precision (1.96 sd of the bias) were 0.46 litre min(-1) m(-2) and +/- 1.15 litre min(-1) m(-2) (r = 0.53) resulting in an overall percentage error of 46%. Subgroup analysis revealed a percentage error of 51% for data pairs obtained intraoperatively (T1-T4), 42% in ICU (T5-T8), and 56% for values obtained during low CI (T1-T8).
CONCLUSIONS: In cardiac surgery patients, CO measured by a new semi-invasive arterial pressure waveform analysis device showed only moderate agreement with intermittent pulmonary artery thermodilution measurement.

Entities:  

Mesh:

Year:  2007        PMID: 17218375     DOI: 10.1093/bja/ael341

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  34 in total

1.  Evaluation of an improved algorithm for arterial pressure-based cardiac output assessment without external calibration.

Authors:  Christopher Prasser; Benedikt Trabold; Alexander Schwab; Cornelius Keyl; Susanne Ziegler; Christoph Wiesenack
Journal:  Intensive Care Med       Date:  2007-09-26       Impact factor: 17.440

2.  Cardiac index value measurement by invasive, semi-invasive and non invasive techniques: a prospective study in postoperative off pump coronary artery bypass surgery patients.

Authors:  Murali Chakravarthy; Subramanyam Rajeev; Vivek Jawali
Journal:  J Clin Monit Comput       Date:  2009-05-02       Impact factor: 2.502

Review 3.  Pressure pulse contour-derived stroke volume and cardiac output in the morbidly obese patient.

Authors:  Donald P Bernstein
Journal:  Obes Surg       Date:  2008-04-29       Impact factor: 4.129

4.  Cross-comparisons of trending accuracies of continuous cardiac-output measurements: pulse contour analysis, bioreactance, and pulmonary-artery catheter.

Authors:  Bouchra Lamia; Hyung Kook Kim; Donald A Severyn; Michael R Pinsky
Journal:  J Clin Monit Comput       Date:  2017-02-10       Impact factor: 2.502

5.  Validation of continuous cardiac output technologies: consensus still awaited.

Authors:  Maurizio Cecconi; Andrew Rhodes
Journal:  Crit Care       Date:  2009-06-26       Impact factor: 9.097

6.  Pitfalls in haemodynamic monitoring based on the arterial pressure waveform.

Authors:  Luigi Camporota; Richard Beale
Journal:  Crit Care       Date:  2010-03-05       Impact factor: 9.097

7.  Intraoperative fluid optimization using stroke volume variation in high risk surgical patients: results of prospective randomized study.

Authors:  Jan Benes; Ivan Chytra; Pavel Altmann; Marek Hluchy; Eduard Kasal; Roman Svitak; Richard Pradl; Martin Stepan
Journal:  Crit Care       Date:  2010-06-16       Impact factor: 9.097

8.  Clinical evaluation of the flotrac/Vigileo system for continuous cardiac output monitoring in patients undergoing regional anesthesia for elective cesarean section: a pilot study.

Authors:  José Otavio Costa Auler; Marcelo L A Torres; Mônica M Cardoso; Thais C Tebaldi; André P Schmidt; Mario M Kondo; Marcelo Zugaib
Journal:  Clinics (Sao Paulo)       Date:  2010-06       Impact factor: 2.365

9.  Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.

Authors:  Xavier Monnet; Nadia Anguel; Brice Naudin; Julien Jabot; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2010-06-10       Impact factor: 9.097

10.  Brachial artery peak velocity variation to predict fluid responsiveness in mechanically ventilated patients.

Authors:  Manuel Ignacio Monge García; Anselmo Gil Cano; Juan Carlos Díaz Monrové
Journal:  Crit Care       Date:  2009-09-03       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.