Literature DB >> 18420863

Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis.

Matthieu Biais1, Karine Nouette-Gaulain, Vincent Cottenceau, Alain Vallet, Jean François Cochard, Philippe Revel, François Sztark.   

Abstract

BACKGROUND: Cardiac output (CO) and invasive hemodynamic measurements are useful during liver transplantation. The pulmonary artery catheter (PAC) is commonly used for these patients, despite the potential complications. Recently, a less invasive device (Vigileo/FloTrac) became available, which estimates CO using arterial pressure waveform analysis without external calibration. In this study, we compared CO obtained with a PAC using automatic thermodilution, instantaneous CO stat-mode (ICO(SM)), and CO obtained with the new device, arterial pressure waveform analysis (APCO) in patients undergoing liver transplantation.
METHODS: Twenty sets of simultaneous measurements of APCO and ICO(SM) were determined in sedated and mechanically ventilated patients undergoing liver transplantation. Time points were as follows: after PAC insertion (T1-3), after portal clamping (T4-6), during anhepathy (T7-9), after graft reperfusion (T10-15), and in the postoperative period in the intensive care unit (T15-20).
RESULTS: We enrolled 20 patients and 400 measurements were obtained. No data were rejected. Bias between ICO(SM) and APCO was 0.8 L/min, 95% limits of agreement were -1.8 to 3.5 L/min. The percentage error was 43%. Bias between ICO(SM) and APCO was correlated with systemic vascular resistance [r(2) = 0.55, P < 0.0001, y = 15.8-2.2 ln(x)] and subgroup analysis revealed an increase in the bias and in the percentage error in patients with low systemic vascular resistance (Child-Pugh grade B and C patients). There was no difference between the different surgical periods.
CONCLUSIONS: Our results suggest that Vigileo/FloTrac CO monitoring data do not agree well with those of automatic thermodilution in patients undergoing liver transplantation, especially in Child-Pugh grade B and C patients with low systemic vascular resistance.

Entities:  

Mesh:

Year:  2008        PMID: 18420863     DOI: 10.1213/ane.0b013e318168b309

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  41 in total

Review 1.  Pulse pressure variation: where are we today?

Authors:  Maxime Cannesson; Mateo Aboy; Christoph K Hofer; Mohamed Rehman
Journal:  J Clin Monit Comput       Date:  2011-02       Impact factor: 2.502

2.  Can the "FloTrac" really track flow in septic patients?

Authors:  Xavier Monnet; Daniel Lahner
Journal:  Intensive Care Med       Date:  2010-12-10       Impact factor: 17.440

3.  The effect of head up tilting on bioreactance cardiac output and stroke volume readings using suprasternal transcutaneous Doppler as a control in healthy young adults.

Authors:  Jie Zhang; Lester A H Critchley; Daniel C W Lee; Kim S Khaw; Shara W Y Lee
Journal:  J Clin Monit Comput       Date:  2016-02-12       Impact factor: 2.502

4.  Agreement in hemodynamic monitoring during orthotopic liver transplantation: a comparison of FloTrac/Vigileo at two monitoring sites with pulmonary artery catheter thermodilution.

Authors:  Matthew Lee; Laurence Weinberg; Brett Pearce; Nicholas Scurrah; David A Story; Param Pillai; Peter R McCall; Larry P McNicol; Philip J Peyton
Journal:  J Clin Monit Comput       Date:  2016-02-16       Impact factor: 2.502

5.  Cardiac index value measurement by invasive, semi-invasive and non invasive techniques: a prospective study in postoperative off pump coronary artery bypass surgery patients.

Authors:  Murali Chakravarthy; Subramanyam Rajeev; Vivek Jawali
Journal:  J Clin Monit Comput       Date:  2009-05-02       Impact factor: 2.502

6.  The impact of systemic vascular resistance on the accuracy of the FloTrac/Vigileo™ system in the perioperative period of cardiac surgery: a prospective observational comparison study.

Authors:  Yohei Sotomi; Katsuomi Iwakura; Yoshiharu Higuchi; Kazuo Abe; Junko Yoshida; Takafumi Masai; Kenshi Fujii
Journal:  J Clin Monit Comput       Date:  2013-06-08       Impact factor: 2.502

Review 7.  Advanced hemodynamic monitoring: principles and practice in neurocritical care.

Authors:  Christos Lazaridis
Journal:  Neurocrit Care       Date:  2012-02       Impact factor: 3.210

8.  Importance of re-calibration time on pulse contour analysis agreement with thermodilution measurements of cardiac output: a retrospective analysis of intensive care unit patients.

Authors:  Christopher G Scully; Shanti Gomatam; Shawn Forrest; David G Strauss
Journal:  J Clin Monit Comput       Date:  2015-08-19       Impact factor: 2.502

9.  Pitfalls in haemodynamic monitoring based on the arterial pressure waveform.

Authors:  Luigi Camporota; Richard Beale
Journal:  Crit Care       Date:  2010-03-05       Impact factor: 9.097

10.  Perforation of the right ventricle induced by pulmonary artery catheter at induction of anesthesia for the surgery for liver transplantation: a case report and reviewed of literature.

Authors:  Maria Auxiliadora-Martins; Erick Apinagés Dos Santos; Daniel Adans Wenzinger; Gil Cezar Alkmim-Teixeira; Gerardo Cristino de M Neto; Ajith Kumar Sankarankutty; Orlando de Castro E Silva; Olindo Assis Martins-Filho; Anibal Basile-Filho
Journal:  Case Rep Med       Date:  2009-12-31