| Literature DB >> 19183431 |
Maurizio Cecconi1, Andrew Rhodes, Jan Poloniecki, Giorgio Della Rocca, R Michael Grounds.
Abstract
Bland-Altman analysis is used for assessing agreement between two measurements of the same clinical variable. In the field of cardiac output monitoring, its results, in terms of bias and limits of agreement, are often difficult to interpret, leading clinicians to use a cutoff of 30% in the percentage error in order to decide whether a new technique may be considered a good alternative. This percentage error of +/- 30% arises from the assumption that the commonly used reference technique, intermittent thermodilution, has a precision of +/- 20% or less. The combination of two precisions of +/- 20% equates to a total error of +/- 28.3%, which is commonly rounded up to +/- 30%. Thus, finding a percentage error of less than +/- 30% should equate to the new tested technique having an error similar to the reference, which therefore should be acceptable. In a worked example in this paper, we discuss the limitations of this approach, in particular in regard to the situation in which the reference technique may be either more or less precise than would normally be expected. This can lead to inappropriate conclusions being drawn from data acquired in validation studies of new monitoring technologies. We conclude that it is not acceptable to present comparison studies quoting percentage error as an acceptability criteria without reporting the precision of the reference technique.Entities:
Mesh:
Year: 2009 PMID: 19183431 PMCID: PMC2688094 DOI: 10.1186/cc7129
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Figure 1Bull's-eye representation of accuracy and precision. With respect to the Bland-Altman plot, accurate measurements mean small bias and precise measurements mean narrow limits of agreement.
Figure 2Different combinations of precision for a reference and a new method that can lead to a percentage error (PE) of 30%. A 30% PE can derive from several combinations of precisions for the two methods compared.
Cardiac output in 20 patients: repeated measurements with the reference technique and single test measurements
| Patient | CO1, L/min | CO2, L/min | CO3, L/min | CO4, L/min | Mean CO, L/min | CV, ± % | CE, ± % | Studied CO, L/min |
| 1 | 6.5 | 8.8 | 7.0 | 7.7 | 7.5 | 13 | 7 | 6.9 |
| 2 | 11.8 | 15.1 | 14.3 | 12.8 | 13.5 | 11 | 5 | 10.9 |
| 3 | 7.8 | 6.7 | 6.5 | 6.6 | 6.9 | 9 | 4 | 6.7 |
| 4 | 7.0 | 7.3 | 6.5 | 7.1 | 7.0 | 5 | 2 | 8.2 |
| 5 | 6.1 | 6.7 | 7.6 | 6.5 | 6.7 | 9 | 5 | 5.9 |
| 6 | 13.2 | 14.4 | 12.8 | 13.6 | 13.5 | 5 | 3 | 14.2 |
| 7 | 13.1 | 11.7 | 14.7 | 13.3 | 13.2 | 9 | 5 | 11.9 |
| 8 | 6.1 | 6.3 | 6.6 | 7.4 | 6.6 | 9 | 4 | 7.3 |
| 9 | 16.2 | 12.7 | 13.3 | 14.4 | 14.2 | 11 | 5 | 13.8 |
| 10 | 5.2 | 5.2 | 3.9 | 5.6 | 5.0 | 15 | 7 | 6.0 |
| 11 | 6.8 | 7.4 | 6.1 | 6.4 | 6.7 | 8 | 4 | 7.1 |
| 12 | 8.2 | 7.7 | 8.2 | 7.6 | 7.9 | 4 | 2 | 7.4 |
| 13 | 6.7 | 5.5 | 6.9 | 5.8 | 6.2 | 11 | 5 | 6.8 |
| 14 | 3.9 | 4.3 | 4.7 | 5.0 | 4.5 | 11 | 5 | 3.9 |
| 15 | 6.1 | 6.6 | 6.7 | 4.9 | 6.1 | 14 | 7 | 7.8 |
| 16 | 8.0 | 8.5 | 8.0 | 8.2 | 8.2 | 3 | 1 | 10.2 |
| 17 | 8.0 | 6.9 | 7.5 | 8.1 | 7.6 | 7 | 4 | 9.4 |
| 18 | 7.0 | 6.3 | 7.5 | 6.5 | 6.8 | 8 | 4 | 5.6 |
| 19 | 8.2 | 7.5 | 8.8 | 8.3 | 8.2 | 7 | 3 | 10.1 |
| 20 | 4.5 | 3.9 | 4.1 | 4.6 | 4.3 | 8 | 4 | 4.5 |
| Average | 8.0 | 8.0 | 8.1 | 8.0 | 8.0 | 9 | 4 | 8.2 |
Mean cardiac output (CO) is the mean of the four measurements. CV is the coefficient of variation for a single measurement, and CE is the coefficient of error when four measurements are averaged. The studied CO is the single measurement for the studied technique.
Figure 3Bland-Altman plot for new technique versus reference technique. Dotted lines represent bias and limits of agreement. Data from Table 1 are used.
Effect of the number of measurements of the reference technique on the percentage error
| Measurements for the reference technique, number | Ref precision, ± percentage | Study precision, ± percentage | PE, ± percentage | ± 30% fulfilled |
| 1 | 18 | 29 | 34 | No |
| 2 | 12 | 29 | 32 | No |
| 3 | 10 | 29 | 31 | No |
| 4 | 9 | 29 | 30 | Yes |
| 5 | 8 | 29 | 30 | Yes |
| 6 | 7 | 29 | 30 | Yes |
| 7 | 7 | 29 | 30 | Yes |
| 8 | 6 | 29 | 30 | Yes |
'Measurements for the reference technique' means the number of measurements averaged for the reference technique. 'Ref precision' is the precision for the reference technique according to the number of averaged measurements, 'study precision' is the precision of the studied technique as measured by the worked example, and PE is the percentage error for the Bland-Altman plot for the reference technique minus the studied technique. The '± 30% fulfilled' column shows whether the PE would be accepted according to a cutoff of 30%.
Figure 4Precision of the reference technique for n averaged measurements and the corresponding percentage error (PE) from the Bland-Altman plot for a fixed level of precision of the studied technique (29%). The PE can change simply by using a more or less precise reference technique, even when the precision of the studied technique is not changed. This may lead to the acceptance of a studied technique even though its performance in terms of precision stays the same. CE, coefficient of error.