Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis.

S G Sakka1, J Kozieras, O Thuemer, N van Hout.   

Abstract

BACKGROUND: Recently, continuous monitoring of cardiac output (CO) based on pulse contour analysis (Vigileo) has been introduced into practice. In this clinical study, we evaluated the accuracy of this system by comparing it with the transpulmonary thermodilution technique (TPID) in septic patients.
METHODS: We studied 24 mechanically ventilated patients with septic shock (16 male, 8 female, age 26-77 yr) receiving treatment with norepinephrine who for clinical indication underwent haemodynamic monitoring by the transpulmonary thermodilution technique using a PiCCO plus system (Pulsion Medical Systems, Munich, Germany). In parallel, arterial pulse contour was applied using the femoral arterial pressure curve (FloTrac pressure sensor, Vigileo monitor, Edwards Lifesciences, Irvine, USA). After baseline measurement, mean arterial pressure was elevated by increasing norepinephrine dosage, and CO was measured again before mean arterial pressure was reduced back to baseline levels. Fluid status and ventilator settings remained unchanged throughout. At each time point, CO by transpulmonary thermodilution was calculated from three central venous bolus injections of 15 ml of saline (<8 degrees C). Linear regression and the Bland-Altman method were used for statistical analysis.
RESULTS: Overall, CO was 6.7 (sd 1.8) (3.2-10.1) litre min(-1) for CO(TPID) and 6.2 (2.4) (3.0-17.6) litre min(-1) for CO(Vigileo((R))). Linear regression revealed: CO(Vigileo) = 1.54 + 0.72 x CO(TPID) litre min(-1), r(2) = 0.26 (P < 0.0001). Mean bias between techniques [CO(TPID)-CO(Vigileo)] was 0.5 litre min(-1) (SD 2.3 litre min(-1)). Correlation coefficients at the three time points were not significantly different from each other.
CONCLUSIONS: Pulse contour analysis-derived CO (Vigileo system) underestimates CO(TPID) and is not as reliable as transpulmonary thermodilution in septic patients.

Entities:  

Mesh:

Year:  2007        PMID: 17611251     DOI: 10.1093/bja/aem177

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  31 in total

Review 1.  Reproducibility of transpulmonary thermodilution cardiac output measurements in clinical practice: a systematic review.

Authors:  Raphaël Giraud; Nils Siegenthaler; Paolo Merlani; Karim Bendjelid
Journal:  J Clin Monit Comput       Date:  2016-01-11       Impact factor: 2.502

2.  Non-invasive automated measurement of cardiac output during stable cardiac surgery using a fully integrated differential CO(2) Fick method.

Authors:  Philip J Peyton; Daniel Thompson; Paul Junor
Journal:  J Clin Monit Comput       Date:  2008-07-12       Impact factor: 2.502

3.  The impact of systemic vascular resistance on the accuracy of the FloTrac/Vigileo™ system in the perioperative period of cardiac surgery: a prospective observational comparison study.

Authors:  Yohei Sotomi; Katsuomi Iwakura; Yoshiharu Higuchi; Kazuo Abe; Junko Yoshida; Takafumi Masai; Kenshi Fujii
Journal:  J Clin Monit Comput       Date:  2013-06-08       Impact factor: 2.502

Review 4.  [Haemodynamic monitoring in the perioperative phase. Available systems, practical application and clinical data].

Authors:  U Wittkowski; C Spies; M Sander; J Erb; A Feldheiser; C von Heymann
Journal:  Anaesthesist       Date:  2009-08       Impact factor: 1.041

Review 5.  Hemodynamic monitoring during liver transplantation: A state of the art review.

Authors:  Mona Rezai Rudnick; Lorenzo De Marchi; Jeffrey S Plotkin
Journal:  World J Hepatol       Date:  2015-06-08

6.  Cardiac output monitoring in septic shock: evaluation of the third-generation Flotrac-Vigileo.

Authors:  Sophie Marqué; Antoine Gros; Loic Chimot; Arnaud Gacouin; Sylvain Lavoué; Christophe Camus; Yves Le Tulzo
Journal:  J Clin Monit Comput       Date:  2013-01-30       Impact factor: 2.502

Review 7.  Alternatives to the Swan-Ganz catheter.

Authors:  Daniel De Backer; Jan Bakker; Maurizio Cecconi; Ludhmila Hajjar; Da Wei Liu; Suzanna Lobo; Xavier Monnet; Andrea Morelli; Sheila Neinan Myatra; Azriel Perel; Michael R Pinsky; Bernd Saugel; Jean-Louis Teboul; Antoine Vieillard-Baron; Jean-Louis Vincent
Journal:  Intensive Care Med       Date:  2018-05-03       Impact factor: 17.440

8.  Pitfalls in haemodynamic monitoring based on the arterial pressure waveform.

Authors:  Luigi Camporota; Richard Beale
Journal:  Crit Care       Date:  2010-03-05       Impact factor: 9.097

9.  Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.

Authors:  Xavier Monnet; Nadia Anguel; Brice Naudin; Julien Jabot; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2010-06-10       Impact factor: 9.097

10.  Changes in stroke volume induced by passive leg raising in spontaneously breathing patients: comparison between echocardiography and Vigileo/FloTrac device.

Authors:  Matthieu Biais; Lionel Vidil; Philippe Sarrabay; Vincent Cottenceau; Philippe Revel; François Sztark
Journal:  Crit Care       Date:  2009-12-07       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.